10Step 10: Structure confirmation (NMR, MS)
Retention time cannot guarantee structure, so spectroscopy and mass spectrometry finally prove that it is the target molecule.Retention time cannot guarantee structure, so spectroscopy and mass spectrometry finally prove that it is the target molecule.
Standard methodStructure confirmation by NMR (1H/13C, plus 19F/31P as needed) + mass spectrometryStructure confirmation by NMR (1H/13C, plus 19F/31P as needed) + mass spectrometry
Details and 3 alternatives
This step coversConfirmation of chemical structure (identification test) (structure confirmation/NMR, MS)
An identification test that finally proves 'whether what was made is really the target molecule.' Chromatographic retention time alone cannot guarantee structure.
Nuclear magnetic resonance (NMR) reads the skeleton and the arrangement of functional groups, and mass spectrometry corroborates the molecular weight. Layering several spectroscopic methods makes the structure certain.
A step that is the linchpin of consistent 'identity confirmation' from the starting materials to the final API. If this wavers, the foundation for discussing purity collapses.
Finally proves that 'what was made is really the target molecule.' HPLC retention time alone cannot guarantee structure, so the skeleton, functional groups and molecular weight are corroborated by several spectroscopic methods. It is the linchpin of identification from the starting materials through to the final API.
MakersAgilent, Waters, Bruker, Honeywell, Sigma-Aldrich
Why it is the standard. NMR is the reference method for primary structure analysis, able to establish structure at the atomic level. MS confirms the molecular weight and provides identification on a different principle. The two are routinely used together for structure confirmation of a new API (ICH Q6A identification test).
Alternatives in practice
- FT-IR (functional group identification, comparison with reference standard) · Thermo Scientific Nicolet iS20/Bruker ALPHA II — Effective for confirming functional groups and comparing against a reference. Cannot determine the full structure.
- UV-Vis spectroscopy · Agilent Cary 60/Thermo Evolution UV-Vis — A supplementary check for chromophores.
- Elemental analysis (CHN) · Elementar vario MICRO cube/Thermo Flash 2000 CHNS — Supports confirmation of the empirical formula. Cannot determine the structure itself.
Step guide
What you actually do
- Dissolve the API in a deuterated solvent (a solvent whose hydrogen is replaced by deuterium) and record 1H and 13C spectra by NMR (nuclear magnetic resonance)
- Where needed, confirm the connectivity between the atoms by two-dimensional NMR
- Measure the accurate molecular mass by high-resolution mass spectrometry (HRMS) and check it against the molecular formula
- Compare the FT-IR spectrum with the reference standard, and bring together whether methods based on different principles all point to the same structure
Decision points
- Which methods do you combine?
- A single method cannot rule out compounds of similar structure, so stack methods based on different principles: NMR for the skeleton, mass spectrometry for the molecular mass, infrared absorption for the functional groups. ICH Q6A likewise asks for a combination of highly specific methods for identification.
- Do you determine the structure again for every batch?
- The structure is determined once in development, and release testing commonly confirms by infrared absorption and the like that the material is the same as the reference standard. When the synthetic route changes, revisit the structure confirmation.
- Is a benchtop NMR enough?
- Benchtop NMR instruments with permanent magnets can be used to confirm known compounds and to follow a process in line, but their resolution and sensitivity are below those of superconducting-magnet instruments. At the stage of determining a new structure, measurement on a high-field instrument is usual.
Common mistakes
- Concluding that it is the same compound only because the HPLC retention time matches the reference standard
- Mistaking the signals of water or residual solvent in the deuterated solvent for an impurity or product signal
Check at your own site
- How many methods based on different principles does your identification testing stack?
11Step 11: Assay and related substances
Quantify the content responsible for efficacy and the related substances from synthesis and degradation to secure the core of quality.Quantify the content responsible for efficacy and the related substances from synthesis and degradation to secure the core of quality.
Standard methodQuantitation by reverse-phase HPLC/UHPLC (UV detection, stability-indicating method)Quantitation by reverse-phase HPLC/UHPLC (UV detection, stability-indicating method)
Details and 3 alternatives
This step coversAssay (content) and quantitation of organic impurities and related substances (assay and related substances/HPLC quantitation)
The main event of quality testing, measuring the core of API purity. It looks at the 'content (assay)' responsible for efficacy and the related substances that get mixed in, together.
Content is quantified by reverse-phase HPLC/UHPLC, and related substances from synthesis and degradation are separated and tracked one by one. This reveals which impurities are present and how much.
Stability studies also follow how degradation products increase, so a 'stability-indicating method' that can properly separate degradation products is indispensable. The story of purity is told here.
How pure the API is, and which impurities are present and how much, is the core of quality. The content responsible for efficacy is measured, and related substances derived from synthesis and degradation are followed one by one. Stability studies also look at how degradation products increase, so a method that can separate the degradation products is essential.
MakersMerck, Agilent, Waters, Metrohm, Honeywell
Why it is the standard. The industry standard based on ICH Q3A (impurities in drug substances), Q6A and Q2 (analytical method validation). HPLC/UHPLC-UV can quantify content and related substances at the same time, and stability-indicating capability is assured by validation (many method development examples such as PMC11371568).
Check also. UV detection in reverse-phase HPLC works well for impurities with a chromophore, but it misses impurities that barely absorb UV, as well as inorganic salts and counterions. When the mass balance does not close, complement it with other detection principles such as CAD/ELSD (evaporative light scattering) or mass spectrometry.
Alternatives in practice
- LC-MS/MS (assignment and quantitation of trace impurities) · Sciex Triple Quad 7500/Waters Xevo TQ-XS — Effective for impurities with weak UV absorption and for structural assignment.
- Ion chromatography · Thermo Scientific Dionex Integrion/Metrohm 940 Professional IC — For quantifying counterions and inorganic ions.
- CAD/ELSD detection · Thermo Scientific Vanquish Charged Aerosol Detector (CAD)/Agilent 1290 Infinity II ELSD — General-purpose quantitation of compounds without UV absorption.
Step guide
What you actually do
- Dissolve the API and the reference standard and prepare them at the set concentrations
- Inject onto a reverse-phase column (C18 and the like, which separates the components on the difference in hydrophobicity) and elute with a gradient
- Detect each peak by UV (DAD) and confirm the system suitability (resolution, repeatability)
- Calculate the assay against the reference standard and the amount of the related substances from the peak areas, and compare them with the specification
Decision points
- Where do the thresholds for related substances come from?
- ICH Q3A draws the lines on the maximum dose per 1 day. At 2 g or less, the reporting threshold is 0.05%, the identification threshold 0.10% (or 1.0 mg/day, whichever is lower) and the qualification threshold 0.15% (or 1.0 mg/day, whichever is lower). Without a dose you cannot draw the lines, and if the dose rises during development you redraw them.
- Can you report amounts as area percent (area%) as they are?
- Area% assumes that the UV sensitivity of the main compound and of the impurity are the same. The European Pharmacopoeia approach treats a sensitivity ratio (response factor) of 0.8 to 1.2 as needing no correction and applies a correction factor to impurities outside that range, so confirm the response factors of the main impurities against standards.
- How do you show that the method does not miss degradation products?
- Run forced degradation that deliberately degrades the material with heat, light, acid, base and oxidation, and show that the degradation products formed separate cleanly from the main compound. Degrading the main compound by around 5 to 20% is the usual target; push it harder and you generate secondary degradation products that never occur in practice.
- What do you do about impurities UV cannot see?
- Impurities and counter-ions that barely absorb UV are missed, so where the mass balance does not close, supplement with CAD or ELSD (methods that nebulize and dry the liquid and detect the resulting particles by charge or by light scattering), or with mass spectrometry.
Common mistakes
- Reading area% values as weight ratios without having confirmed the response factors
- Taking development forward with a method whose quantitation limit does not reach the reporting threshold, and having to rebuild it later
Check at your own site
- Have the response factors of your main related substances been confirmed against standards?
- Where does the quantitation limit of your method sit relative to the reporting threshold?
12Step 12: Ultra-trace nitrosamine analysis
Reliably control impurities that are carcinogenic even in trace amounts at ppb levels, orders of magnitude lower than ordinary tests.Reliably control impurities that are carcinogenic even in trace amounts at ppb levels, orders of magnitude lower than ordinary tests.
Standard methodUltra-trace quantitation by LC-MS/MS (or GC-MS/MS)Ultra-trace quantitation by LC-MS/MS (or GC-MS/MS)
Details and 3 alternatives
This step coversUltra-trace control of mutagenic impurities and nitrosamines (ultra-trace nitrosamine analysis)
A special control step that watches for impurities that are carcinogenic even in trace amounts at concentrations orders of magnitude lower. It deals with a range that ordinary purity testing cannot reach.
Alkylating agents and nitrosamines are the leading examples. They are quantified at the parts-per-billion (ppb) level using methods with extremely low detection limits, such as LC-MS/MS.
Since nitrosamine contamination was discovered in the 2018 sartan drug product (antihypertensive) problem, regulation has tightened all at once. It is a test at the front line of safety.
Impurities that carry a carcinogenic risk even at very low levels need to be controlled at ppb levels, orders of magnitude lower than in ordinary purity testing. Alkylating agents and nitrosamines are representative, and regulation has tightened since the 2018 sartan drug product problem. They are reliably controlled with methods that have extremely low detection limits.
MakersAgilent, Waters, Honeywell, Sigma-Aldrich
Why it is the standard. Control based on ICH M7 (mutagenic impurities). Nitrosamines and genotoxic impurities require ppb-level control, and LC-MS/MS is the standard for its high selectivity and sensitivity (consistent with the methods of LCGC/Shimadzu/PMC12655250). HS-GC-MS/ECD is also used for volatile alkyl halides.
Check also. For mutagenic impurities and nitrosamines, the practice is first to assess from the synthetic route and reagents whether they could form, as a structural alert, and then to measure at the ppb level only the compounds that are needed. Since the 2018 sartan drug product problem, regulation (ICH M7, etc.) and the control targets continue to be updated, so operate in line with the latest notifications.
Alternatives in practice
- HS-GC-MS (volatile nitrosamines) · Agilent 8890 GC/7000E MS+7697A headspace — Effective for volatile low-molecular-weight nitrosamines.
- GC-ECD (volatile alkyl halides, etc.) · Agilent 8890 GC+ECD detector — High-sensitivity detection of halogenated compounds. Limited versatility.
- HRMS screening/process purge assessment · Thermo Orbitrap Exploris/Waters Xevo G3 QTof (HRMS) — Supports the search for unknown nitroso compounds and measured purge factors.
Step guide
What you actually do
- List the starting materials, reagents, intermediates and by-products of the synthetic route and assess them with (Q)SAR (prediction of mutagenicity from the structure)
- Work out, from the steps where an amine and nitrite coexist, which nitrosamines could form
- Set an acceptable intake for each compound of concern, and work back from it to the quantitation limit you need
- Add a stable-isotope-labelled internal standard, prepare the sample and quantify by LC-MS/MS or GC-MS/MS
Decision points
- How do you narrow down what to measure?
- ICH M7 asks you to run 2 kinds of (Q)SAR, one rule-based and one statistics-based, and where both are negative the impurity can in principle be treated as of no concern. Where the results disagree, an expert review follows, and where that still does not settle it, measure in a bacterial reverse mutation assay (Ames test). A negative prediction is not positive proof that the substance is not mutagenic.
- How is the acceptable limit set?
- For mutagenic impurities the baseline is the TTC (threshold of toxicological concern) of 1.5 µg per day, relaxed for shorter treatment durations: within 1 month it is 120 µg/day. Nitrosamines are stricter still; NDMA, for example, is 96 ng/day. For nitrosamines derived from the structure of the API (NDSRI), the limit is set with approaches such as the carcinogenic potency categorization approach (CPCA), which assigns a potency category from the structure.
- LC-MS/MS, or GC-MS/MS?
- NDSRI, which are not volatile, are usually measured by LC-MS/MS or high-resolution LC-MS, and small volatile nitrosamines and alkyl halides by GC-based methods. You are measuring traces in the presence of a large amount of API, so building the sample preparation and the separation conditions is the core of the method development.
- If the process removes it, can you leave the final API untested?
- ICH M7 also accepts a control strategy without testing of the final product, provided you can show removal in the process with measured data. The removal data is the premise; how far you rely on it is for the quality unit to judge and is settled in agreement with the regulatory authority.
Common mistakes
- Treating a negative (Q)SAR result as proof of no mutagenicity, and settling by prediction what should be measured
- An impurity nobody anticipated appears in neither the specification nor the test methods, and manufacturing continues with nobody measuring it
Check at your own site
- Have you worked out every step in your synthetic route where an amine and nitrite could be present at the same time?
- Was your NDSRI limit set from compound-specific data, from read-across or from CPCA?
13Step 13: Residual solvents / headspace GC
Quantify residual solvents, whose limits differ with toxicity, as a final check of drying adequacy and safety.Quantify residual solvents, whose limits differ with toxicity, as a final check of drying adequacy and safety.
Standard methodQuantitation of residual solvents by headspace GC-FID (HS-GC)Quantitation of residual solvents by headspace GC-FID (HS-GC)
Details and 3 alternatives
This step coversQuantitation of residual organic solvents (residual solvents/headspace GC)
A safety test that confirms no trace of the solvents used in synthesis and crystallization remains in the API. It also serves as a final check that drying was sufficient.
Quantified by headspace gas chromatography (HS-GC), which warms the sample and captures the vaporized solvent. It efficiently picks up readily volatile solvents.
Toxicity differs by solvent type, so stricter limits are set for the more toxic ones. Pass/fail is judged by comparing the residual amount with the limit.
Solvents used in synthesis and crystallization remain in the API in trace amounts. Toxicity differs by type, so the more toxic the solvent, the stricter the limit set. It also serves as a final check that drying was sufficient, making it a test directly tied to safety.
MakersMerck, Agilent, Shimadzu, PerkinElmer, Metrohm
Why it is the standard. Standard method based on ICH Q3C (residual solvents) and USP<467>. HS-GC-FID separates volatile solvents from a nonvolatile matrix and quantifies them with high sensitivity, and is the method specified in the pharmacopoeias (consistent with methods such as PMC12614197).
Alternatives in practice
- HS-GC-MS (when identification is needed) · Agilent 8890 GC/5977B MSD+7697A headspace — Effective for identifying peaks of unknown solvents. Quantitative reproducibility is somewhat inferior to FID.
- Direct injection GC · Agilent 8890 GC-FID (direct liquid injection) — For high-boiling, hard-to-volatilize solvents. Watch for matrix contamination.
- Karl Fischer (water measured separately) · Metrohm 851 Titrando/Mettler Toledo C20S — Residual water cannot be measured by GC, so a separate method is needed.
Step guide
What you actually do
- Dissolve the API in a diluent (water, DMSO, DMF and the like) and place it in a sealed vial
- Heat the vial and inject onto the GC (gas chromatograph) only the solvent that has come out into the gas phase
- Detect each solvent with a flame ionization detector (FID) and calculate the amount against the standard solutions
- Compare with the limit for each solvent (ICH Q3C), and identify unknown peaks by GC-MS
Decision points
- Can the limits stay as the ppm values in the guideline table?
- The ppm values in the table assume a dose of 10 g per day (Option 1). Once the actual maximum daily dose is known, Option 2, which converts directly from the PDE (permitted daily exposure), can be used; where that value exceeds Option 1, you have to show that the solvent has been reduced to the practical minimum.
- Do you look for every solvent, or measure only the ones you used?
- USP <467> splits the work into Procedure A, which looks for the presence of solvents on a G43 phase column, Procedure B, which confirms on a phase of different polarity, and Procedure C, which quantifies; where the solvents used are known, Procedure C alone is taken to be enough. Class 1 solvents are as a rule not used, and where one has had to be, you quantify to show that it has come below the limit.
- Can loss on drying (LOD) take its place?
- ICH Q3C(R9) states explicitly that where only Class 3 solvents have been used, a non-specific method such as a suitably validated loss on drying may be used for the decision. That validation has to allow for the effect of the volatility of the solvent on the result.
- What do you choose as the diluent?
- Water is common for APIs that dissolve in water, and DMF or DMSO for those that do not. DMSO can elute at the same position as solvents such as DMF in headspace work, so take care in the choice where DMF is also an analyte.
Common mistakes
- The peak of the diluent overlaps the solvent of interest, and the solvents are mixed up
- Believing that ICH Q3C has a "Class 4" category, and not providing the safety justification for a solvent with insufficient data
Check at your own site
- Are your residual solvent limits set on Option 1 or on Option 2?
14Step 14: Elemental impurities (ICP-MS)◇ Branches
Control metal catalyst residues, toxic even in trace amounts, within permitted limits and make a final check that removal was sufficient.Precious metal catalyst residues drop out of the control scope
Standard methodElemental quantitation by ICP-MS (microwave acid digestion pretreatment)Elemental quantitation by ICP-MS (microwave acid digestion pretreatment)
ReagentsAPI reference standards and impurity standards (CRS)Because the resolution route does not use asymmetric metal catalysts (Rh/Ru, etc.), monitoring of precious metal residues of the ICH Q3D classes such as Ru/Rh is normally unnecessary. Instead, the focus shifts to elements that can be introduced in the salt formation and decomposition steps (such as Na) and to those derived from the equipment used ◇ With Racemic synthesis + optical resolution (diastereomeric salt crystallization)Precious metal catalyst residues drop out of the control scope◇ With Asymmetric synthesis (direct construction of a single enantiomer)Elemental quantitation by ICP-MS (microwave acid digestion pretreatment)
Details and 3 alternatives
This step coversQuantitation of elemental impurities and metal catalyst residues (elemental impurities/ICP-MS)
A step that measures the 'leftovers' of metal catalysts used in synthesis. If palladium, nickel and the like remain in the API even in trace amounts, toxicity becomes a concern.
After pretreatment in which the sample is dissolved in acid, elements are quantified by inductively coupled plasma mass spectrometry (ICP-MS). Even trace metals are picked up accurately.
Permitted amounts are set for each route of administration. The elements specified by regulation are reliably measured, providing final support that catalyst removal was sufficient.
If metal catalysts used in synthesis, such as palladium and nickel, remain in the API even in trace amounts, toxicity becomes a concern. Permitted amounts are set for each route of administration, so the elements specified by regulation are reliably measured. It is also the final check that catalyst removal was sufficient.
MakersMerck, Agilent, PerkinElmer, Honeywell
Why it is the standard. Standard method based on ICH Q3D (elemental impurities) and USP<232>/<233>. ICP-MS quantifies many elements simultaneously from ppb to ppt and covers catalyst residues (Pd/Pt/Ni, etc.) and environment-derived metals (As/Cd/Hg/Pb) (consistent with the methods of PMC9433241/PMC10987250).
Check also. ICP-MS measures the total amount of an element but does not distinguish the chemical form (valence, species) that determines toxicity. Permitted amounts are set by ICH Q3D risk assessment for each route of administration, and the measurement is used only as a final check that catalyst residue removal was sufficient.
◇ With Racemic synthesis + optical resolution (diastereomeric salt crystallization). In contrast, on the asymmetric synthesis side, Ru/Rh/Pd residues derived from catalysts such as BINAP-Ru/Rh are the main targets of ICP-MS monitoring
Alternatives in practice
- ICP-OES (for high-concentration elements) · Agilent 5800/5900 ICP-OES/Thermo iCAP PRO — Simple and robust for high-concentration elements. Trace-level (ppb) sensitivity is inferior to ICP-MS.
- AAS (atomic absorption) · Agilent 240/280 AA/PerkinElmer PinAAcle 900 — For single-element targets. Simultaneous multi-element measurement is not possible.
- Colorimetric methods/simple kits (limit tests) · Merck heavy metal limit test kits/Agilent Cary 60 UV-Vis — Simple judgment in the manner of a limit test. Low quantitative accuracy and selectivity.
Step guide
What you actually do
- Work out in a risk assessment which elements could enter, and from where: the catalyst, the raw materials, the equipment, the containers
- Put the API into high-purity acid and digest the organic matter in a microwave digestion system to give a solution
- Quantify the elements of interest by ICP-MS (inductively coupled plasma mass spectrometry), correcting the drift in sensitivity with an internal standard
- Confirm the recovery of the digestion and the measurement with samples spiked with known amounts of the elements
What changes with the conditionWhether monitoring centres on the precious metals of an asymmetric catalyst, or on elements from salt formation and breaking and from the equipment (in asymmetric synthesis, catalyst residues such as Ru, Rh and Pd are the main targets; the optical resolution route uses no chiral metal catalyst, so the focus moves to elements that enter with salt formation and breaking, such as Na, and to elements from the equipment)
Decision points
- How do you decide which elements to measure?
- ICH Q3D asks you to start from a risk assessment rather than from the analysis. Metals from catalysts used intentionally are measured from the outset; for the rest, narrow the list by working through the process flow and the formulation to find entry from the raw materials, the equipment and the containers.
- Measure every lot, or settle it with a one-off assessment?
- 30% of the permitted daily exposure (PDE) is taken as the control threshold, and where you stay below it and the element is under control, no additional control is held to be needed. Where the threshold is exceeded, decide between removal in the process, a change of raw material and control in the specification.
- What changes the permitted limit?
- The route of administration moves it by orders of magnitude: for mercury, for example, the oral limit is 30 µg/day against 3 µg/day parenteral and 1 µg/day inhalation. When you take an oral product on to an injectable, redo the assessment.
- How do the elements you measure change with the synthetic route?
- In asymmetric synthesis, the Ru, Rh and Pd from chiral catalysts such as BINAP-Ru and Rh complexes are the main monitoring targets. The racemic synthesis and optical resolution route uses no chiral metal catalyst, so precious metals that are not used can be taken off the list where the risk assessment can show that they are not introduced. If other steps use Pd couplings and the like, those catalysts are assessed separately. On that basis, the emphasis moves to elements that can enter with salt formation and breaking (Na and the like) and to elements from the equipment.
Common mistakes
- Incomplete digestion lowers the recovery, and the value reported is lower than the real one
- Not working out the entry from equipment and containers, which do not appear in the formulation, so that it is never found
Check at your own site
- Does your elemental impurity risk assessment include entry from the equipment and the containers?
- Where do your catalyst metals sit against 30% of the permitted limit?
15Step 15: Chiral purity / chiral HPLC◇ Branches
An unwanted enantiomer changes efficacy and safety, so it is separated and quantified to assure the success of asymmetric synthesis.Chiral HPLC is promoted to the rate-determining IPC
Standard methodEnantiomer separation and quantitation by chiral HPLC (chiral stationary phase, CSP)Chiral HPLC (enantiomer separation and ee quantitation with a chiral stationary phase, CSP). On the resolution route, it is used as a rate-determining process control item for building up ee (endpoint determination of fractional recrystallization, in-process checks after each crystallization)
◇ With Racemic synthesis + optical resolution (diastereomeric salt crystallization)Chiral HPLC (enantiomer separation and ee quantitation with a chiral stationary phase, CSP). On the resolution route, it is used as a rate-determining process control item for building up ee (endpoint determination of fractional recrystallization, in-process checks after each crystallization)◇ With Asymmetric synthesis (direct construction of a single enantiomer)Enantiomer separation and quantitation by chiral HPLC (chiral stationary phase, CSP)
Details and 3 alternatives
This step coversConfirmation of chiral purity and enantiomeric excess (ee) (chiral purity/chiral HPLC)
A test to confirm that the molecule that is the mirror image (enantiomer), with left and right reversed as if reflected in a mirror, has not been mixed in on the unwanted side. It is essential for optically active APIs.
Enantiomers, related like left and right hands, can differ in efficacy and safety. The two are separated by HPLC with a chiral stationary phase and quantified as an impurity.
Did asymmetric synthesis or chiral resolution work as intended? A stereochemical quality check that assures the result in the form of enantiomeric excess (ee).
In an optically active API, if the enantiomer (a molecule related like left and right hands) is mixed in, efficacy and safety change. The unwanted enantiomer is separated and quantified as one impurity. This test assures that asymmetric synthesis or chiral resolution worked as intended.
MakersDaicel, Agilent, Waters, Merck, Honeywell
Why it is the standard. Chiral HPLC (polysaccharide CSP) is the most established technique for enantiomer separation and corresponds to the chiral API specification (enantiomeric purity) of ICH Q6A. The Daicel CHIRALPAK series is used as the industry standard.
◇ With Racemic synthesis + optical resolution (diastereomeric salt crystallization). On the asymmetric synthesis side, ee is set by the catalyst, and chiral HPLC is mainly a confirmatory test at release. On the resolution side, it becomes a key analysis that governs the process
Alternatives in practice
- Chiral SFC (supercritical fluid chromatography) · Waters ACQUITY UPC2/Agilent 1260 Infinity II SFC+Daicel CHIRALPAK columns — Fast and low in solvent, and also advantageous for preparative work. Requires equipment investment.
- Optical rotation measurement · Anton Paar MCP 5100/JASCO P-2000 polarimeters — Simple, but unsuited to high-precision quantitation of enantiomeric excess.
- Chiral GC · Agilent 8890 GC + cyclodextrin-based chiral column (Agilent CP-Chirasil-Dex CB) — Effective for volatile chiral compounds. Limited range of application.
Step guide
What you actually do
- Confirm the separation and the assignment of the peaks with a racemate and with standards of both enantiomers
- Inject onto a column with a polysaccharide-based chiral stationary phase (CSP) and separate the enantiomers under normal-phase or reverse-phase conditions
- Calculate the proportion from the peak area of the unwanted enantiomer, and from it the ee
- Compare with the specification. On the optical resolution route, measure after each crystallization as well, to decide whether the process may go on
What changes with the conditionWhether chiral HPLC is a confirmation test at release, or an in-process test that governs the process (in asymmetric synthesis the ee is set by the catalyst and confirmation at release is the centre of it; in optical resolution it judges the endpoint of the fractional recrystallization and serves as the IPC after each crystallization, deciding whether the process moves on)
Equipment and materials
EquipmentAnalytical HPLCPolarimeter
Decision points
- Coated column, or immobilized?
- On a coated column the selector (the polysaccharide derivative that discriminates the enantiomers) dissolves out in strong solvents such as THF and dichloromethane, so even traces carried in with the sample diluent or the autosampler wash will degrade the performance over time. Where there is any chance of using strong solvents, narrow the candidates to immobilized columns.
- When it does not separate, will a smaller particle size separate it?
- What decides whether the separation is possible is not the particle size but the backbone of the selector (cellulose-based or amylose-based) and the solvent conditions. Cellulose-based and amylose-based phases are complementary in selectivity, so screen with a set that includes both before settling the particle size and the column dimensions.
- Is this test a confirmation at release, or a test that governs the process?
- In asymmetric synthesis the ee is set by the catalyst and chiral HPLC is mainly a confirmation test at release. On the optical resolution route it is used to judge the endpoint of the fractional recrystallization and as an in-process test after each crystallization, so it is treated as a critical analysis that decides whether the process goes on.
Common mistakes
- Letting small amounts of strong solvent from the sample diluent or the wash keep flowing onto a coated column, and losing its performance without noticing
- Not confirming the separation from the related isomers, and overestimating the enantiomeric purity
Check at your own site
- Is there any THF or dichloromethane in the sample diluent or the wash used in your chiral analysis?
- Was your chiral column chosen after trying both cellulose-based and amylose-based phases?
16Step 16: Crystal polymorph
Polymorph differences change solubility, stability and absorption, so consistent manufacture of the target polymorph is confirmed lot by lot.Polymorph differences change solubility, stability and absorption, so consistent manufacture of the target polymorph is confirmed lot by lot.
Standard methodPolymorph identification by powder X-ray diffraction (XRPD)Polymorph identification by powder X-ray diffraction (XRPD)
ReagentsNo reagents required (Cu Kα source, λ=1.5406 Å). Zero-background sample holder (single-crystal Si) used.
Details and 3 alternatives
This step coversConfirmation of crystal polymorph and crystallinity (crystal polymorph/powder X-ray diffraction, XRPD)
Even for the same molecule, a different arrangement within the crystal (polymorph) makes it behave like a different substance. Ease of dissolution, stability and absorption change.
Powder X-ray diffraction (XRPD) reads the arrangement of the crystal like a fingerprint and confirms that the target polymorph is being made consistently.
Transition to an unwanted polymorph or to the amorphous form leads directly to problems with efficacy and stability. That is why the identity of the crystal is watched lot by lot.
Even for the same molecule, a different arrangement of the crystal (polymorph) changes solubility, stability and absorption. Whether the target polymorph is being produced consistently is confirmed like a fingerprint. Transition to an unwanted polymorph or to the amorphous form leads directly to problems with efficacy and stability, so it is checked lot by lot.
MakersMalvern Panalytical, Bruker, Rigaku
Why it is the standard. XRPD is the reference method for polymorph identification, giving a fingerprint of the crystal structure. Crystals show sharp Bragg peaks and amorphous material a broad halo, and it is standard both for polymorph control under ICH Q6A and for patent protection (consistent with Malvern Panalytical's explanations).
Alternatives in practice
- Solid-state NMR (ss-NMR) · Bruker AVANCE NEO 400 WB+CP-MAS probe — Strong at discriminating local structure of polymorphs and amorphous forms. Instrumentation and analysis are advanced.
- Raman/FT-IR spectroscopy · Thermo DXR3 Raman/Bruker ALPHA II FT-IR — Effective for polymorph screening and mapping. Quantitative polymorph ratios require calibration.
- Polarized light microscopy (PLM) · Olympus BX53P polarizing microscope/Leica DM2700 P — Confirmation of crystallinity and crystal habit. Supports polymorph identification.
Step guide
What you actually do
- Pack the API into the sample holder and level the surface
- Irradiate with X-rays and measure the intensity of the diffraction while varying the angle, to obtain the diffraction pattern
- Compare the positions and the intensities of the peaks with the reference pattern of the target polymorph
- Check that no peaks of another polymorph, and no amorphous halo (a broad, gentle rise), are present
Decision points
- Can you judge the polymorph on XRPD alone?
- XRPD is the basic method for identifying a crystal form, but preferred orientation of the crystals changes the intensity ratios, and very small amounts of another polymorph or of amorphous material can be missed. Where it bears on quality, supplement with DSC, Raman or solid-state NMR.
- Do you put the crystal form in the specification?
- ICH Q6A gives a decision tree for judging whether the difference between polymorphs bears on the performance of the product. Where it does, control the crystal form in the specification; where it does not, you may decide not to, but keep the justification on record either way.
- Do you quantify small amounts of another polymorph?
- Where the admixture bears on the product, calibrate with samples blended at known ratios and confirm how far you can detect and quantify it. The lower limit of detection varies with the pair of polymorphs and with the instrument, so establish it on your own samples.
Common mistakes
- Reading intensity ratios changed by the orientation of the sample as the admixture of another polymorph
- Declaring small amounts of amorphous material or of another polymorph "absent" without having confirmed the limit of detection
Check at your own site
- Have you confirmed at what level another polymorph becomes detectable by your XRPD?
17Step 17: Thermal analysis and water (DSC, TGA, KF)
Grasp polymorph stability and hydration/solvation from thermal behavior, and the effect of water content on stability and content calculation.Grasp polymorph stability and hydration/solvation from thermal behavior, and the effect of water content on stability and content calculation.
Standard methodDSC+TGA (thermal analysis) + Karl Fischer titration (water)DSC+TGA (thermal analysis) + Karl Fischer titration (water)
ReagentsDSC/TGA: no reagents required (high-purity nitrogen purge, Al/Pt crucibles). KF: Honeywell Hydranal reagents (Composite 5/2 for volumetric, Coulomat AG for coulometric).
Details and 3 alternatives
This step coversConfirmation of melting point, thermal behavior and water/solvation (thermal analysis and water/DSC, TGA, KF)
A thermal analysis step that reads the 'behavior' of the API when heated. From the melting temperature, transition temperature and weight changes, it probes polymorph stability and whether the material is a hydrate or solvate.
Differential scanning calorimetry (DSC) follows the heat flow in and out, thermogravimetry (TGA) follows the loss of weight, and Karl Fischer titration measures water at the same time.
Water is controlled separately. Because it affects stability and content calculation, it is separated from thermal behavior and measured accurately.
The temperatures of melting and transitions on heating, and changes in weight, tell whether a polymorph is stable and whether it is a hydrate or solvate. Water content is also measured accurately. Water affects stability and content calculation, so it is controlled separately.
MakersTA Instruments, Mettler Toledo, Metrohm, Honeywell
Why it is the standard. The standard combination: DSC evaluates melting point, transitions and polymorph conversion, and TGA evaluates solvation/hydration and thermal stability (consistent with Agno/Malvern explanations). For water, KF titration is the reference method specified by the pharmacopoeia (USP<921>).
Alternatives in practice
- Dynamic vapor sorption (DVS) · Surface Measurement Systems DVS Intrinsic/TA Instruments Q5000 SA — Effective for understanding hygroscopicity and hydrate formation behavior.
- Hot-stage microscopy (thermal microscopy) · Linkam THMS600 hot stage + Olympus polarizing microscope/Mettler Toledo FP90 — Melting and transitions are confirmed visually. Quantitative heat measurement depends on DSC.
- Loss on drying (LOD) · Mettler Toledo HC103 halogen moisture analyzer/Sartorius MA37 infrared moisture analyzer — Simple measurement of volatiles. Cannot distinguish water from solvent.
Step guide
What you actually do
- Weigh out a small amount of the API, place it in a DSC pan and seal it
- Heat at a constant rate in DSC (differential scanning calorimetry) and record the heat flow of melting and of the transitions
- Follow the weight loss while heating in TGA (thermogravimetric analysis), and see at what temperature and how much water or solvent leaves
- Measure the water separately by Karl Fischer titration and judge, against the TGA weight loss, whether the material is a hydrate or a solvate
Equipment and materials
PartsSample pans for DSC and TGA
Decision points
- Is the TGA weight loss water, or solvent?
- TGA does not tell you what left, so compare it with the water content from Karl Fischer titration, and where a difference remains, consider solvent. Confirm the solvent by GC where needed.
- Is the DSC pan sealed, pinhole or open?
- How the pan is closed decides whether the water or solvent that leaves can escape, and that moves the shape and the temperature of the peaks. Measure samples you mean to compare with the same pan and the same heating rate.
- How far do you investigate hygroscopicity?
- For APIs that form a hydrate with humidity, or whose amorphous material returns to the crystalline state, look at the moisture uptake behaviour by dynamic vapour sorption (DVS, measuring the change in weight while varying the humidity).
Common mistakes
- Reading the whole TGA weight loss as water, and missing a solvate
- Comparing results obtained with different pans or heating rates, and reading the difference in transition temperature as a difference in polymorph
Check at your own site
- Do you read your TGA weight loss against the Karl Fischer water content?
18Step 18: Water determination (Karl Fischer)
To control the API water content by specification and assure hydration state, hygroscopicity and stabilityTo control the API water content by specification and assure hydration state, hygroscopicity and stability
Standard methodCoulometric Karl Fischer method (absolute quantitation of trace water, with a heating evaporation oven as needed)Coulometric Karl Fischer method (absolute quantitation of trace water, with a heating evaporation oven as needed)
ReagentsHoneywell Hydranal Coulomat AG (coulometric anolyte, product number 34836), also usable in cells without a diaphragm. When used with an oven, Hydranal Coulomat AG-Oven (product number 34739).
Details and 2 alternatives
This step coversQuantitation of water (water content) — water in the API is quantified by the Karl Fischer method to confirm hydrate state, hygroscopicity and compliance with the loss-on-drying specification (water determination/Karl Fischer)
A step that measures the water contained in the API itself, as an absolute amount. The water content required in the release specification is established here.
Trace water is directly quantified by coulometric Karl Fischer titration. If needed, an oven that heats the sample to drive out the water is used with it, picking up even tightly held water.
Used to judge hydrate stoichiometry, hygroscopicity and compliance with the loss-on-drying specification. Separately from thermal analysis (DSC/TGA), it is a test that treats water as the main subject.
Water (water content) is a mandatory item in API release specifications, and trace water is quantified directly by the Karl Fischer method. Separately from the thermal behavior seen by DSC/TGA, it is used to judge hydrate stoichiometry, hygroscopicity and compliance with the loss-on-drying specification.
MakersMetrohm, Honeywell
Why it is the standard. Karl Fischer water determination is mandatory in the standard API release tests.
Alternatives in practice
- Volumetric Karl Fischer method (for items with high water content) · Metrohm 901 Titrando/Mettler Toledo V20S+Hydranal Composite 5 — Suitable for items with water of 1% or more. Higher upper limit than the coulometric method.
- Loss on drying (LOD) · Mettler Toledo HC103 halogen moisture analyzer/vacuum dryer + electronic balance — Simple, but includes volatile solvents and is not water-specific.
Step guide
What you actually do
- Bring the titration cell to a dry state and wait for the drift (the amount of water finding its way into the cell) to settle
- Weigh out the API and either add it to the cell directly, or heat it in a moisture vaporizer (oven) so that a dry carrier gas brings only the water across
- Titrate until the water no longer consumes iodine, and calculate the water content from the consumption
- Confirm the state of the instrument and the reagents with a water standard, and compare the result with the specification
Equipment and materials
Materials and reagentsKarl Fischer reagents and water standards
Decision points
- Volumetric, or coulometric?
- The volumetric method adds reagent of known concentration and measures water quickly, from the mg level to the percent level, but the titer of the reagent has to be checked periodically. The coulometric method generates iodine by electrolysis, so no titer determination is needed and it suits trace water, while samples with a high water content take time.
- Direct titration, or the vaporizer method?
- For APIs that dissolve poorly in the solvent, or that contain components such as ketones that side-react with the reagent, the vaporizer method, which drives only the water out in an oven and carries it across, is a candidate. Too high a heating temperature degrades the sample and reads the water high, too low and not all of the water comes out, so set the temperature for each sample by measurement.
- Which method does "Karl Fischer" in the specification mean?
- USP <921> splits it into the volumetric method (Method Ia), back titration (Ib) and the coulometric method (Ic), and the European Pharmacopoeia likewise treats the volumetric and the coulometric method in separate sections. Decide from the specification side which method you test by, before you choose the instrument.
Common mistakes
- Setting the vaporizer temperature too high, so that the sample degrades and the water reads high
- Choosing the instrument on the width of its measuring range alone, without confirming the recovery and the variation on your own samples
Check at your own site
- Which Method of USP <921> does the Karl Fischer test in your specification correspond to?
19Step 19: Particle size distribution
Particle size variation affects solubility, flowability and drug product uniformity, so it is controlled as a critical quality attribute.Particle size variation affects solubility, flowability and drug product uniformity, so it is controlled as a critical quality attribute.
Standard methodParticle size distribution measurement by laser diffraction (Dv10/Dv50/Dv90)Particle size distribution measurement by laser diffraction (Dv10/Dv50/Dv90)
EquipmentLaser diffraction particle size analyzer: Malvern Panalytical Mastersizer 3000 (Aero dry/Hydro wet units)/Sympatec HELOS+RODOS/Microtrac SYNC.
ReagentsDry: oil-free compressed air/nitrogen. Wet: dispersion medium (water or isoparaffin-based, with surfactant added, e.g., Tween 20/lecithin).
Details and 3 alternatives
This step coversConfirmation of physical properties related to formulability, such as particle size distribution and specific surface area (particle size distribution/laser diffraction)
A physical property test that measures variation in particle size. Particle size affects dissolution, flowability and the uniformity of the final drug product, so it is an important quality attribute related to formulability.
The particle size distribution is measured by laser diffraction, and Dv10, Dv50 and Dv90 are read from the cumulative distribution from the small side. It confirms whether micronization gave the target distribution.
Especially for poorly soluble drugs, particle size is directly tied to variation in effect. That is why it is controlled as a critical quality attribute, down to the shape of the distribution.
Variation in particle size affects dissolution, flowability and the uniformity of the final drug product. Measure whether micronization gave the target distribution and, if needed, look at surface area as well. For poorly soluble drugs, particle size is directly tied to variation in effect, so it is controlled as a critical quality attribute.
MakersMalvern Panalytical, Sympatec, Microtrac
Why it is the standard. Laser diffraction is the industry standard that gives the particle size distribution as Dv10/50/90 quickly and with high reproducibility, and is widely adopted for particle size control of micronized APIs (the relevant ICH Q6A specification) (consistent with the practice of SK pharmteco/Malvern).
Alternatives in practice
- BET specific surface area measurement · Micromeritics TriStar II Plus/3Flex (N2 adsorption) — An indicator of fine powder and adsorption surface area. Does not give the particle size distribution itself.
- Dynamic image analysis · Malvern Morphologi 4/Sympatec QICPIC — Particle size analysis accompanied by shape information. Watch sample amount and statistics.
- Sieving/microscopy · Retsch AS 200 sieve shaker/Olympus optical microscope + image analysis — A simple method for coarse particles. Low resolution in the fine-powder range.
Step guide
What you actually do
- Introduce the sample to the instrument dry (dispersed in air) or wet (dispersed in a liquid it does not dissolve in)
- For the wet route, choose a dispersion medium the API does not dissolve in, with a surfactant, and break up the agglomerates with ultrasound or the like
- Calculate the particle size distribution from the pattern of the scattered laser light
- Read Dv10, Dv50 and Dv90 from the distribution cumulated from the fine end, and compare them with the specification
Equipment and materials
EquipmentLaser diffraction particle size analyzer
Decision points
- Dry, or wet?
- The dry route needs little preparation and is fast, while agglomerates can survive and fine particles can be broken. The wet route disperses more easily, but if the API dissolves even slightly in the dispersion medium the values change, so choose a medium it does not dissolve in.
- How do you set the dispersion conditions?
- Vary the dispersion pressure or the ultrasound power and look for the range in which the agglomerates come apart and the values settle, while the particles themselves are not broken. Once set, use the conditions unchanged.
- How do you set the width of the specification?
- For poorly soluble APIs the particle size feeds straight into the variability of the dissolution, so confirm how it acts in the dissolution testing of the formulation, and then decide whether to draw lines not only on Dv50 but on the spread of the distribution (Dv10 and Dv90) as well.
- Do you look at the shape of the particles and the surface area as well?
- Laser diffraction calculates on the assumption that the particles are spheres, so for needle-shaped crystals the information on shape is supplemented by image analysis. Where the surface area of the fines bears on the dissolution, add the BET method (measuring surface area from gas adsorption).
Common mistakes
- Dispersion is not enough, and agglomerated lumps are measured as large particles
- Setting the dispersion conditions too hard, and breaking the particles themselves
Check at your own site
- Was your particle size specification set in connection with dissolution testing?
20Step 20: Microbial limits and endotoxin
Check for microbial contamination before formulation. For raw materials of sterile drug products, pyrogens are also controlled according to the route of administration.Check for microbial contamination before formulation. For raw materials of sterile drug products, pyrogens are also controlled according to the route of administration.
Standard methodMicrobial limits test (TAMC/TYMC) + BET (LAL method) when neededMicrobial limits test (TAMC/TYMC) + BET (LAL method) when needed
Details and 3 alternatives
This step coversConfirmation of microbial limits and endotoxin (when needed) (microbial limits and endotoxin)
An API is a chemical, but this step confirms there is no microbial contamination before it goes on to formulation. The required tests are chosen according to the route of administration.
The microbial limits test measures the viable count (total aerobic microbial count, total yeast and mold count) to determine how many microbes are present. This is the basic check.
When the API becomes a raw material for a sterile drug product such as an injectable, endotoxin, a pyrogen, also becomes a concern. It is confirmed by BET (LAL method), putting a double safety net in place.
An API is a chemical, but before it goes on to formulation, it is checked for microbial contamination. Especially when it becomes a raw material for a sterile drug product such as an injectable, not only the microbial count but also pyrogens (endotoxin) are a concern. The required tests are chosen according to the route of administration of the product.
MakersMerck, Charles River, Lonza, bioMérieux, Sartorius
Why it is the standard. Standard based on USP<61>/<62> and <85>, and ICH Q6A. Microbial limits for oral/non-sterile APIs, and endotoxin as well for raw materials of sterile drug products. rFC is increasingly adopted as an alternative to animal-derived reagents.
Check also. Whether microbial limits or endotoxin testing is needed depends on which drug product and route of administration the API finally goes to. The level required differs greatly between oral solid dosage forms and injectables, so tests are chosen by working backward from downstream requirements rather than imposed uniformly at the API stage.
Alternatives in practice
- Recombinant factor C (rFC) method · Lonza PyroGene rFC Assay/Charles River Endosafe Trillium rCR cartridges — Independent of horseshoe crabs and sustainable. Requires validation against each pharmacopoeia and method.
- Rapid microbiological methods (RMM) · bioMérieux BacT/ALERT/Rapid Micro Biosystems Growth Direct — Speeds up results. An equivalence study is required at introduction.
- Sterility test (for sterile APIs) · Sartorius Sterisart/Merck Steritest Symbio (membrane sterility test) — For sterile APIs. Not applicable to limit testing of non-sterile products.
Step guide
What you actually do
- Dissolve or suspend the API and either filter it through a membrane filter or mix it into agar medium
- Incubate on media for the total aerobic microbial count (TAMC) and for the total yeast and mould count (TYMC) at the set temperature and for the set number of days, and count the colonies
- Where the API goes into an injectable, measure endotoxin with LAL reagent (from horseshoe crab) or with recombinant factor C (rFC)
- Confirm with samples spiked with microorganisms or endotoxin that the API is not interfering with the test
Equipment and materials
Materials and reagentsMedia for microbiological testing
Decision points
- Which tests do you need?
- It depends on which formulation and route of administration the API goes on to. For a raw material for oral solid dosage forms, microbial limit testing is the centre of it; for a raw material for injectables, endotoxin, a pyrogen, also matters; and for a sterile API, sterility testing is added, so work back from the requirements of the later steps.
- Does the API itself interfere with the growth of the microorganisms or with the LAL reaction?
- With antimicrobial or poorly soluble APIs the spiked microorganisms may not be recovered, so confirm that dilution, neutralization or filtration removes the interference. For endotoxin as well, confirm the recovery with spiked samples.
- LAL reagent, or rFC?
- Where you want to avoid reagents of animal origin, or put weight on security of supply, rFC is a candidate. When you switch, measure the same samples by both and compare them, and confirm the standing of the method in the pharmacopoeias.
Common mistakes
- Testing an antimicrobial API without removing the interference, so that it looks free of microorganisms
- Assuming that killing the microorganisms removes the endotoxin as well, and not controlling the two separately
Check at your own site
- Are the microbiological test items for your API decided from which formulation and route of administration it goes on to?