Small molecules Process map

Chemically synthesized APIs

This reference process map organizes the chemical synthesis and analysis of small-molecule APIs in process order: reaction → workup and crystallization → purification → characterization. The best method depends on conditions such as the synthetic route and crystal polymorph, so each step also lists alternative and complementary methods.

Editor's note. Small-molecule analytical methods look mature not because they are dull, but because ICH gave the world a common yardstick. HPLC is the backbone of every step because one principle can tie purity, content and reaction endpoint together, and the receiving lab can reproduce it anywhere with confidence. Beginners tend to stumble on the solid-state side instead: a batch is within specification by HPLC but fails on solubility or stability, the typical case of overlooking a polymorph difference that XRPD would have shown. The main battleground in recent years is nitrosamines. Even when the assay is comfortably in range, whether you can chase ppb levels by LC-MS/MS can make or break an API. General-purpose instruments cannot reach that level, so this is the one step where you should be prepared to commit to capital investment.

Used to makeSmall-molecule API that becomes the cytotoxic payload of an ADC (MMAE, DM1, etc.), API that becomes the small-molecule payload of a cell-free ADC

At a glance

Select a step to jump to its card. ◇ marks a step whose standard method changes with the condition.

  1. Synthesis and reaction3 steps
    1. Incoming testing, raw material identification (changes with the condition)
    2. In-process control
    3. Intermediate isolation and IPC (changes with the condition)
  2. Workup and crystallization4 steps
    1. Extraction, phase separation, water washing
    2. Decolorization and clarifying filtration
    3. Crystallization and polymorph control (changes with the condition)
    4. Filtration and drying
  3. Formulation and fill-finish2 steps
    1. Micronization / jet milling
    2. Final blending and homogenization
  4. Quality attributes and analytics11 steps
    1. Structure confirmation (NMR, MS)
    2. Assay and related substances
    3. Ultra-trace nitrosamine analysis
    4. Residual solvents / headspace GC
    5. Elemental impurities (ICP-MS) (changes with the condition)
    6. Chiral purity / chiral HPLC (changes with the condition)
    7. Crystal polymorph
    8. Thermal analysis and water (DSC, TGA, KF)
    9. Water determination (Karl Fischer)
    10. Particle size distribution
    11. Microbial limits and endotoxin

Synthesis and reaction(3 steps)

Step 1: Incoming testing, raw material identification◇ Branches

Starting-material purity determines the quality of the whole synthesis, and impurities from the first step are carried through to later steps.Resolving agent added as a critical raw material

Standard methodIdentification by FT-IR (infrared spectroscopy) + purity confirmation by HPLCIdentification by FT-IR (infrared spectroscopy) + purity confirmation by HPLC
ReagentsHPLC mobile phase solventsAPI reference standards and impurity standards (CRS)In addition to the raw materials for the API backbone, the chiral resolving agent (L-(+)-tartaric acid, di-p-toluoyl/dibenzoyl-D-tartaric acid, L-(+)-mandelic acid, (1S)-camphorsulfonic acid, etc.) is received as a critical raw material. In addition to identifying the resolving agent itself (FT-IR), confirmation of optical rotation/chiral purity is essential
With Racemic synthesis + optical resolution (diastereomeric salt crystallization)Resolving agent added as a critical raw materialWith Asymmetric synthesis (direct construction of a single enantiomer)Identification by FT-IR (infrared spectroscopy) + purity confirmation by HPLC
Details and 3 alternatives

This step coversIncoming identification and quality check of raw materials and reagents (incoming testing, raw material identification)

Think of it as inspecting the ingredients in cooking. Before charging to the reactor, this step checks at the door that the raw material or reagent is really the target compound and that its purity is within specification.

The procedure is simple. Infrared spectroscopy (FT-IR) identifies the molecular skeleton like a fingerprint, and high-performance liquid chromatography (HPLC) measures purity. Together, the two pin down both identity and quality.

The outcome of a synthesis is determined by the purity of the starting materials. If poor-quality material gets in at the first step, impurities are carried through to the later steps. The key is therefore to stop quality problems at the entrance.

Synthesis is determined by the purity of the starting materials. If poor-quality material enters the first step, impurities are carried through to the later steps. So before charging to the reaction, confirm 'is this really the target compound?' and 'is the purity within specification?'

MakersAgilent, Thermo Fisher, Bruker, Merck, Shimadzu

Why it is the standard. ICH Q7 (GMP for APIs) requires identification testing of starting materials and reagents. IR gives rapid identification from the functional-group fingerprint and is a standard identification method in the pharmacopoeias (USP/JP/EP). Purity is complemented by HPLC.

Check also. FT-IR (infrared spectroscopy) can rapidly identify 'what compound it is' from the fingerprint of its functional groups, but it cannot distinguish enantiomers, crystal polymorphs or trace structural analogues. The purity and stereochemistry of the starting material are covered by tests based on other principles, such as HPLC and chiral analysis.

With Racemic synthesis + optical resolution (diastereomeric salt crystallization). Lot control is tightened because the optical purity of the resolving agent directly determines the final ee

Alternatives in practice

  • Rapid identification by near-infrared/Raman spectroscopy · Bruker MPA II (NIR)/Thermo Scientific TruScan RM (handheld Raman, raw material identification through the container) — Effective for nondestructive identification through the container at receipt. Discriminating variants requires building a library first.
  • Gas chromatography (volatile raw materials and solvents) · Agilent 8890 GC (FID)/Shimadzu Nexis GC-2030 — For confirming the purity/impurities of volatile liquid raw materials and solvents.
  • Confirmation of physical constants such as melting point and optical rotation · Mettler Toledo MP90 melting point apparatus/Anton Paar MCP 5100 polarimeter — A simple supporting identification. Low molecular specificity, so unsuitable as a primary method.

Step guide

What you actually do

  1. Sample the incoming raw material containers, using the number of containers and the method set in the procedure
  2. Record a spectrum by FT-IR (infrared spectroscopy) in ATR mode and identify the material by overlaying it with the reference standard or a registered spectrum
  3. Measure purity and the main impurities by HPLC (high-performance liquid chromatography) and compare them with the specification
  4. Compare the supplier's certificate of analysis (CoA) values with your own results and, once the quality unit has made its decision, move the lot to released status

What changes with the conditionWhether the critical raw materials are only those that build the skeleton, or a chiral resolving agent as well (asymmetric synthesis uses no resolving agent; with racemic synthesis and optical resolution the optical purity of the resolving agent feeds through to the final ee, so incoming testing adds optical rotation and chiral purity to FT-IR and lot control is tighter)

Equipment and materials

EquipmentHandheld spectrometer (Raman, near-infrared)FT-IR spectrometer (ATR)Analytical HPLC (UV/DAD detection)Polarimeter
Materials and reagentsSolvents for HPLC mobile phasesReference standard (CRS)Chiral resolving agent (optical resolution route)

Decision points

Identify every container, or only representative containers?
EU GMP Annex 8 asks that, as a rule, samples be taken from every container of a starting material for medicinal products and identified. Raw materials and reagents used to make an API fall under ICH Q7, where at least 1 identity test per lot is the baseline. Even so, a mix-up is not a matter of degree, so more sites now verify all containers without opening them, using handheld Raman or near-infrared. How far you can cut back follows your approved procedures and the decision of the quality unit.
How far can you rely on the certificate of analysis (CoA)?
Where you have qualified the supplier and have a record of repeatedly comparing your own test results with the CoA values and finding them consistent, you can build a control strategy that relies on the CoA for items such as purity. When the supplier changes its manufacturing site, its testing laboratory or the method, repeat that comparison; identity, at least, is confirmed at the receiving site as a rule.
Can the specification stay as the supplier's standard specification?
You decide for yourself whether the impurities that matter in your own synthesis (those the later steps cannot remove, those with a mutagenicity concern) are covered by the specification. Even when the sheet says 99.9% purity, it does not tell you what the remaining 0.1% is, so take an impurity profile at least once.
Do you treat the chiral resolving agent as a critical raw material as well?
On the racemic synthesis and optical resolution route, the optical purity of the resolving agent (the optically active reagent used to separate the racemate) feeds straight through to the final enantiomeric excess (ee), so optical rotation and chiral purity are checked on top of the FT-IR identity. The asymmetric synthesis route uses no resolving agent, and incoming testing centres on the raw materials that build the API skeleton.

Common mistakes

  • Skipping incoming testing because a CoA is available, and missing that the supplier has changed its testing laboratory or manufacturing site
  • Taking comfort from a passing FT-IR identity test and overlooking that it cannot distinguish enantiomers, polymorphs or structurally similar impurities

Check at your own site

  • For which of your raw materials do you identify every container?
  • When did you last compare the CoA values with your own test results for your main raw materials?

Step 2: In-process control

Misjudging the stop point leaves unreacted starting material or increases by-products. Endpoint determination governs yield.Misjudging the stop point leaves unreacted starting material or increases by-products. Endpoint determination governs yield.

Standard methodMonitoring of conversion and impurity profile by HPLC (sampled in-process analysis)Monitoring of conversion and impurity profile by HPLC (sampled in-process analysis)
Details and 3 alternatives

This step coversReaction progress monitoring and endpoint determination (in-process control) (in-process control/reaction monitoring)

A mid-process check to judge when to 'stop' the reaction. Like tasting a simmering dish, you check partway how far the running reaction has progressed.

The method is to draw a small amount of liquid from the reaction vessel and follow the conversion and the appearance of impurities by high-performance liquid chromatography (HPLC). This is called in-process control.

Misjudging the stop point is costly. Too early and unreacted starting material remains; too late and over-reaction increases by-products. Determine the optimal endpoint before moving on to the next operation.

If you do not check how far the reaction has progressed along the way, you will misjudge when to stop. Stop too early and unreacted starting material remains; too late and over-reaction increases by-products. Track the conversion in real time or by sampling, and move on at the optimal endpoint.

MakersMerck, Agilent, Waters, Bruker, Honeywell

Why it is the standard. ICH Q7 requires in-process control (IPC). HPLC is the industry standard for confirming reaction completion and can quantitatively track residual starting material, intermediates and impurities at the same time (consistent with practical descriptions such as Pharmaguideline).

Alternatives in practice

  • In situ ATR-FTIR (real-time reaction tracking) · Mettler Toledo ReactIR 702L (DiComp/SiComp probes) — Monitors reaction start and endpoint in real time without sampling. Better suited to understanding behavior than to quantitation.
  • TLC (simple reaction tracking) · Merck TLC silica gel 60 F254 plates + UV cabinet (CAMAG UV Cabinet 4) — Quick at the bench. Low quantitation, so it complements the primary method.
  • In-line/benchtop NMR, NIR · Bruker Fourier 80 (benchtop NMR)/Bruker Matrix-F (process NIR) — Depends on the solvent and reaction system, but allows nondestructive continuous monitoring.

Step guide

What you actually do

  1. Run the reaction in a jacketed reactor at the set temperature and agitation
  2. Withdraw reaction mixture at the set times, stop the reaction immediately (quench) and dilute it
  3. Separate starting material, intermediate and product by HPLC and calculate the conversion and the amount of the main impurities
  4. Compare with the approved acceptance criteria; if they are met, move on to the next operation, and if not, extend the reaction within the limits of the procedure

Decision points

Withdraw samples and measure by HPLC, or follow the reaction in line?
Where you want figures that resolve the related substances, a withdrawn sample measured by HPLC is often the choice. Where you want to see the onset and the shape of the endpoint as they happen, in-line FT-IR with an ATR probe inserted in the vessel suits better, and a configuration that combines both is also used. In-line values are estimates returned by a calibration model, so compare them with HPLC at the key points.
What do you set the endpoint criterion on?
Not only on the amount of starting material left, but on how much by-product grows when the reaction is extended and how far the later extraction and crystallization can each remove them. Draw the line from the reaction time-course data taken in development, and revise it against manufacturing experience.
Does the reaction keep going between sampling and measurement?
The reaction continues in the withdrawn sample as well, so set a procedure that stops it right after sampling by cooling or by adding a reagent, or that dilutes it immediately. If the way you stop it varies, the same reaction mixture gives a different value every time you measure.
When an IPC value falls outside the range, who may act, and how far?
ICH Q7 allows trained production staff to adjust the process as long as they stay within a range pre-approved by the quality unit. Outside that range, record it as a deviation; how it is handled from there follows the approved procedure and the decision of the quality unit.

Common mistakes

  • Reading a conversion higher than the real one because the reaction continued inside the withdrawn sample
  • Drawing only one criterion line, so that normal variation is raised as a deviation every time, or an early sign of a problem is missed

Check at your own site

  • Was your reaction endpoint criterion set with a look at the by-products that grow when the reaction is extended?
  • Does your procedure say who may act, and how far, when an IPC value falls outside the range?

Step 3: Intermediate isolation and IPC◇ Branches

In multistep synthesis, each intermediate is isolated and brought into specification so that impurities are not carried forwardIntermediate isolation doubles as optical resolution

Standard methodIsolation of intermediates by crystallization and filtration, with in-process control (purity, content and impurity confirmation by HPLC/GC)Add the resolving agent to the racemic intermediate to form diastereomeric salts, then crystallize the target salt preferentially, or by fractional recrystallization (repeated as needed), based on the solubility difference to enrich it optically. Racemization and recycling of the opposite enantiomer in the mother liquor are considered
ReagentsBulk organic solventsAPI reference standards and impurity standards (CRS)Resolving agent (tartaric acid derivatives, mandelic acid, CSA, etc.) + recrystallization solvent system
With Racemic synthesis + optical resolution (diastereomeric salt crystallization)Add the resolving agent to the racemic intermediate to form diastereomeric salts, then crystallize the target salt preferentially, or by fractional recrystallization (repeated as needed), based on the solubility difference to enrich it optically. Racemization and recycling of the opposite enantiomer in the mother liquor are consideredWith Asymmetric synthesis (direct construction of a single enantiomer)Isolation of intermediates by crystallization and filtration, with in-process control (purity, content and impurity confirmation by HPLC/GC)
Details and 2 alternatives

This step coversIsolation and purity confirmation of intermediates (key intermediates and registered starting materials) — in multistep synthesis, the intermediate is isolated after each reaction by crystallization, filtration, etc., and brought into conformance with the IPC specification before moving to the next step (intermediate isolation and IPC)

A checkpoint: each time a reaction in a multistep synthesis finishes, the key intermediate or registered starting material (RSM/KSM) is taken out of the liquid, its purity is confirmed, and only then is it passed on.

It is isolated as a solid by crystallization and filtration, dried, and then judged for purity, content and impurities against the specification. If it does not pass, it does not go on to the next reaction.

Why check at every stage? Because the more impurities are removed at the intermediate stage, the more stable the quality of the final API. What is built up along the way determines the final quality.

In the standard multistep synthesis of a small-molecule API, it is usual to isolate and dry the key intermediates and registered starting materials (RSM/KSM) after the reaction ends, and to judge purity and impurity profile against the specification before moving to the next reaction. Removing impurities at the intermediate stage directly affects the quality of the final API.

MakersMerck, Agilent, Teledyne ISCO, Honeywell, Pfaudler

Why it is the standard. Intermediate isolation is a step that standardly must be passed in multistep API synthesis and is the most important control point for cutting off impurity carry-over.

Check also. Which intermediate is designated as the registered starting material (RSM) in the filing is a matter to be agreed with the regulatory authority; all steps and impurities from that point on become subject to control. Where the line is drawn determines the later filing burden, so it is not decided by internal judgment alone.

With Racemic synthesis + optical resolution (diastereomeric salt crystallization). The theoretical yield is at most 50% per stage. Preferential crystallization is possible only for conglomerates, and most cases are salt-formation-type resolutions

Alternatives in practice

  • Telescoping (carrying the intermediate into the next reaction without isolation) · Mettler Toledo OptiMax 1001 (with ReactIR for in-line concentration monitoring) — Shortens the process and improves yield. The risk is that the quality of each intermediate cannot be controlled directly.
  • Intermediate purification by column chromatography · Biotage Selekt (Sfär Silica cartridges)/Teledyne ISCO CombiFlash NextGen — Effective for hard-to-separate intermediates. At large scale, cost and solvent load are high.

Step guide

What you actually do

  1. Cool the post-reaction mixture, or add an anti-solvent (a solvent the product dissolves in poorly), to precipitate the intermediate
  2. Collect the crystals on a Nutsche filter (a pressure or vacuum filter that receives the crystals on a filter plate) or on a filtering centrifuge, then wash and dry them
  3. Measure purity, assay and impurities by HPLC or GC and compare them with the IPC specification
  4. Charge the released intermediate to the next reaction (in a telescoped process it is carried forward as a solution, without isolation)

What changes with the conditionWhether the isolation only purifies, or also carries the optical resolution (asymmetric synthesis isolates the intermediate by an ordinary crystallization; racemic synthesis with optical resolution forms a diastereomeric salt with a resolving agent, crystallizes the wanted salt preferentially and repeats the recrystallization where needed. The theoretical yield is at most 50% per pass, and whether to racemize the opposite enantiomer in the mother liquor and recycle it is a question as well)

Decision points

Isolate the intermediate, or carry it forward as a solution (telescoping)?
Isolating cuts the impurities off at that stage and lets you check the quality of the intermediate directly, but it adds steps and time. Telescoping shortens the process and tends to raise the yield, while the quality of each intermediate is harder to control directly, so you need data showing that the carried-over impurities are removed in the later stages.
Which intermediate do you draw as the registered starting material (RSM)?
Everything downstream of the RSM comes under GMP and the filing, so drawing the line earlier adds control burden, and drawing it later puts impurity control outside the regulator's view. If an impurity formed before the candidate carries through to the API, move the line earlier; in the end it is agreed with the regulatory authority, not decided in-house alone.
What do you base the intermediate's IPC specification on?
Work back from how far the next reaction and the final crystallization can remove each impurity, and set how much may be carried over. For impurities with a mutagenicity concern, confirm early whether they can be controlled at this stage.
On the optical resolution route, what does this isolation look at?
A resolving agent is added to the racemic intermediate to form a diastereomeric salt (turning the difference in properties between the enantiomers into a difference in solubility by making a salt), and the less soluble one is crystallized preferentially, so this isolation is itself the optical resolution. The theoretical yield is at most 50% per pass, so whether to racemize the opposite enantiomer left in the mother liquor and recycle it is considered as well.

Common mistakes

  • Carrying a telescoped intermediate into the next reaction without looking at its quality, and carrying over an impurity the later stages do not remove
  • Setting the RSM line on the cheaper side, and being challenged on impurity control later

Check at your own site

  • Can you show that impurities formed before your RSM do not carry through to the API?
  • For the stages you telescope, can you show with data that the carried-over impurities are removed later?

Workup and crystallization(4 steps)

Step 4: Extraction, phase separation, water washing

Primary purification that removes salts, catalyst and water-soluble impurities coarsely by phase separation, reducing the load on the later crystallization.Primary purification that removes salts, catalyst and water-soluble impurities coarsely by phase separation, reducing the load on the later crystallization.

Standard methodLiquid-liquid extraction, phase separation and washing (water wash, acid/base wash)Liquid-liquid extraction, phase separation and washing (water wash, acid/base wash)
Details and 3 alternatives

This step coversPrimary purification of the crude product by extraction, phase separation and washing (extraction, phase separation, water washing)

The liquid after the reaction is a 'hodgepodge' in which salts, catalyst, unreacted material and solvent are mixed in with the target compound. This is the primary purification that first washes out the bulk of them.

The liquid is washed after being separated into a layer compatible with water (aqueous layer) and a layer compatible with oil (organic layer). Water-soluble impurities and inorganic salts are moved to the aqueous side and discarded, and the target stays in the organic layer.

The more is removed coarsely here, the cleaner the later crystallization proceeds. It is, so to speak, the prep work that makes the finishing crystallization easier.

In addition to the target compound, the reaction liquid contains salts, catalyst, unreacted material and solvent. Separating it into aqueous and organic layers and washing first removes water-soluble impurities and inorganic salts to a large extent. The more is removed coarsely here, the easier the later crystallization becomes.

MakersMerck, Sartorius, Pall, Honeywell, Sulzer

Why it is the standard. Standard workup for chemically synthesized APIs. Phase separation removes water-soluble impurities and inorganic salts efficiently, and in ICH Q11 process design it is also positioned as a basic primary purification operation.

Alternatives in practice

  • Activated carbon treatment + filtration (decolorization and removal of trace impurities) · Cabot NORIT powdered activated carbon + Pall SUPRApak depth filter/Sartorius Sartoclear depth filter — Effective for pigments and polar impurities. Watch for API loss and the handling of carbon.
  • Quench + filtration (removal of solid by-products) · Agitated Nutsche Filter (Rosenmund RVNF/De Dietrich Guedu) — A simple way to remove salts and inorganic solids.
  • Continuous countercurrent extraction (continuous liquid-liquid extraction) · Sulzer centrifugal extractor/Koch-Glitsch countercurrent extraction column — Suited to large-volume, continuous production. Capital investment and emulsion management are the challenges.

Step guide

What you actually do

  1. Treat the reaction mixture with a quench solution or with water to deactivate the remaining reagents
  2. Add a water-immiscible organic solvent (ethyl acetate, toluene and the like), agitate, and move the product into the organic layer
  3. Stop the agitation, wait for the two layers to separate and draw the aqueous layer off from the bottom; the interface is judged visually or by conductivity
  4. Wash the organic layer with dilute acid, dilute alkali and brine to remove the residual salts and the acidic and basic impurities

Decision points

What do you select the extraction solvent on?
On the product dissolving well in it while staying immiscible with water, on the two layers separating quickly, and on the solvent sitting in the lower-toxicity classes of the residual solvent guideline (ICH Q3C). Matching it to the solvent used in the following crystallization saves a solvent swap.
How do you set the wash pH and the number of washes?
Choose the pH of the wash solution from whether the impurity you want to remove is acidic or basic, and confirm experimentally the pH range in which the product does not escape into the aqueous layer. Set the number of washes by measuring how far the impurity falls with each one.
What if an emulsion (the two layers no longer separating) forms?
Adding salt to raise the density of the aqueous layer, changing the temperature, or filtering off the solids collected at the interface are the moves usually tried. If it happens on every batch in manufacturing, revisit the solvent and concentration conditions themselves in development.
Could the quench and wash reagents themselves become a source of new impurities?
There are cases where work-up that decomposes excess azide with nitrite became the source of a nitrosamine. Work through the process flow sheet to find whether an amine and nitrite are ever present in the same step, and whether recovered solvent is bringing amines in.

Common mistakes

  • Rushing the phase split while an emulsion is still present, and discarding the product together with the interface layer
  • Amines carried in with recovered solvent become the source of an impurity in the next lot

Check at your own site

  • Is there any work-up operation in your process where an amine and nitrite are present at the same time?
  • Was your number of washes set by measuring how far the impurity falls with each one?

Step 5: Decolorization and clarifying filtration

To remove colored impurities, trace metals and foreign matter by activated carbon and clarifying filtration before crystallization and obtain high-purity crystalsTo remove colored impurities, trace metals and foreign matter by activated carbon and clarifying filtration before crystallization and obtain high-purity crystals

Standard methodActivated carbon treatmentActivated carbon treatment
Details and 2 alternatives

This step coversDecolorization (activated carbon treatment) and clarifying (polish) filtration — activated carbon is added to the crude product solution and stirred to adsorb and remove colored impurities, trace metals and fine particles, and the carbon and foreign matter are removed by clarifying filtration before the solution goes to crystallization (decolorization and clarifying filtration/carbon treatment)

A 'polishing' step done just before crystallization. Activated carbon is added to the crude product solution and stirred, so that colored substances, residual metals and trace impurities are soaked up onto the carbon surface.

After adsorption, clarifying (polish) filtration filters off carbon grains and insoluble foreign matter. This gives a clear, clean solution, ready for crystallization.

Activated carbon treatment before crystallization is the norm in API purification. If color and trace impurities are removed first, the target crystal form can be precipitated cleanly in the crystallization that follows.

In API purification, it is standard to treat with activated carbon immediately before crystallization to adsorb and remove colored substances, residual metals and trace impurities. The clarifying (polish) filtration that follows removes carbon particles and insoluble foreign matter, and the target crystal form is precipitated from a clean solution.

MakersSartorius, Pall, 3M

Why it is the standard. Activated carbon decolorization + clarifying filtration before crystallization is a step that standardly must be passed in small-molecule API purification.

Check also. Activated carbon adsorbs colored substances and trace metals well, but it can also adsorb the API itself and lower the yield. The amount of carbon and the contact time are set by balancing decolorizing effect against recovery, and attention is also paid to adsorption properties that vary from lot to lot.

Alternatives in practice

  • Removal of residual catalyst (Pd, etc.) with metal scavenger resins · Johnson Matthey QuadraPure TU/Smopex fibers, SiliCycle SiliaMetS Thiol/Thiourea — Selective removal of residual Pd/metals. Decolorization is limited, so choose between it and carbon treatment as appropriate.
  • In-line passage through cartridge-type activated carbon · 3M Zeta Plus activated carbon cartridges (CUNO)/Pall Seitz AKS carbon modules — Decolorization in a closed system, avoiding the addition of powdered carbon. Capacity is smaller than with depth-type carbon.

Step guide

What you actually do

  1. Add powdered activated carbon to the crude product solution and agitate at the set temperature and time
  2. Remove the carbon on a filter precoated with a filter aid (diatomaceous earth and the like), or on a depth filter (a filter that traps particles within a thick filter medium)
  3. Pass the solution through a polish filter (a final cartridge that removes foreign particles) and send it to the crystallizer
  4. Where residual palladium and the like are the target, use a metal scavenger (silica or resin carrying functional groups that bind metals)

Decision points

What do you set the carbon loading and the contact time on?
Carbon can adsorb the API itself and cut the yield, so vary the conditions at small scale and decide on both the decolorization and impurity removal and the recovery. The adsorption behaviour of carbon varies by brand and by lot, so repeat the same trial when you change brand.
Add powdered carbon, or pass the solution through a carbon-impregnated cartridge?
Powdered carbon is easy to adjust in amount and has the larger capacity, but it means handling powder and filtering it off. A carbon cartridge decolorizes in a closed system with no powder to handle, while its capacity is smaller, so choose on the volume and on how strong the colour is.
Remove the residual metal with carbon, or use a metal scavenger?
Where you want to remove a specific metal such as palladium selectively, scavengers such as silica carrying thiol or thiourea groups are candidates. Their decolorizing effect is limited, so use them alongside carbon. How far you have to go comes from the permitted limits in the elemental impurity guideline (ICH Q3D).

Common mistakes

  • Raising the carbon loading to prioritize decolorization, and losing yield because the API is adsorbed as well
  • Changing the brand or lot of carbon without repeating the adsorption trial, so that decolorization and recovery change

Check at your own site

  • Were your carbon loading and contact time set with the recovery in view as well?
  • Where does the residual metal after your decolorization sit against the ICH Q3D permitted limits?

Step 6: Crystallization and polymorph control◇ Branches

The most important step, handling purification and solid form determination at once; polymorph and salt determine solubility, stability and manufacturability.Ordinary crystallization → fractional crystallization of diastereomeric salts

Standard methodControlled crystallization (cooling/antisolvent/reactive crystallization) + seeding and PAT monitoringIn addition to controlled crystallization of the polymorph and solid form, fractional recrystallization is performed using the solubility difference of the diastereomeric salts. After resolution, the salt is decomposed with acid/base to regenerate the free-form enantiomer, and the final crystallization of the API is then performed
ReagentsBulk organic solventsChiral resolving agent salt; acid/base for salt decomposition (HCl, NaOH/NaHCO3, etc.)
With Racemic synthesis + optical resolution (diastereomeric salt crystallization)In addition to controlled crystallization of the polymorph and solid form, fractional recrystallization is performed using the solubility difference of the diastereomeric salts. After resolution, the salt is decomposed with acid/base to regenerate the free-form enantiomer, and the final crystallization of the API is then performedWith Asymmetric synthesis (direct construction of a single enantiomer)Controlled crystallization (cooling/antisolvent/reactive crystallization) + seeding and PAT monitoring
Details and 3 alternatives

This step coversPurification by crystallization and building the target solid form (polymorph, salt form) (crystallization and polymorph control)

The peak of the synthesis, handling purification and solid-form building at the same time. The solution is cooled or an antisolvent is added so that only the target precipitates as crystals.

With conditions set properly, impurities stay in the mother liquor (the remaining liquid) and only the target comes out as crystals. Seed crystals are added, and precipitation is controlled while being monitored by process analytical technology (PAT).

In addition, which crystal polymorph or salt form is produced is also decided here. It is the most important stage, which also determines the solid form and thereby affects ease of dissolution, stability and ease of manufacture.

Crystallization is the most important step, handling purification and solid form determination at the same time. With conditions set properly, impurities stay in the mother liquor and only the target comes out as crystals. Moreover, which crystal polymorph or salt is produced determines solubility, stability and ease of manufacture.

MakersMerck, Honeywell, Sigma-Aldrich

Why it is the standard. ICH Q6A requires control of polymorphs and solid forms. Supersaturation control by seeding and PAT (FBRM/ATR-FTIR) is the standard approach for obtaining the target polymorph, particle size and purity reproducibly (consistent with the practice of Crystal Pharmatech and others).

Check also. Crystallization is the most important step, handling purification and solid form determination at once, but there is no guarantee that the targeted crystal polymorph or salt form will reproduce on scale-up. Behavior changes with stirring, cooling rate and seed crystals, and an undesired polymorph can occasionally appear suddenly, so PAT (process analytical technology) monitoring and polymorph control are carried out in parallel.

With Racemic synthesis + optical resolution (diastereomeric salt crystallization). The endpoint of fractional recrystallization is determined by chiral purity (optical rotation/chiral HPLC), and residual resolving agent and the opposite diastereomer are controlled as impurities

Alternatives in practice

  • Simple cooling recrystallization · Mettler Toledo EasyMax 102 (programmed cooling) — Simple, but control of polymorph and particle size is limited.
  • Salt and cocrystal screening · Technobis Crystal16/Crystalline (multi-reactor crystallization screener) — Comprehensive exploration of solubility and polymorphs. Small scale.
  • Continuous crystallization (MSMPR/continuous flow) · Alconbury Weston/Cambridge Reactor Design continuous crystallizers, monitored with Mettler Toledo PAT — Uniform particle size and higher productivity. Startup and control are complex.

Step guide

What you actually do

  1. Dissolve the crude product in the crystallization solvent and adjust it to the set temperature and concentration
  2. Within the supersaturation range (more dissolved than the solubility allows), add seed crystals (seed of the crystal form you are aiming at)
  3. Control the cooling rate or the anti-solvent addition rate to grow the crystals, and follow the number and the size of the particles with a probe such as FBRM (which counts particle chord lengths)
  4. Age the batch at the set endpoint temperature and send the crystal slurry to filtration

What changes with the conditionWhether it is only a controlled crystallization, or adds the fractional recrystallization of a diastereomeric salt (asymmetric synthesis runs only the crystallization that builds the polymorph and the solid form; optical resolution raises the optical purity by fractional recrystallization, then breaks the salt back to the free form before the final crystallization, and the remaining diastereomer of the opposite configuration is controlled as an impurity as well)

Equipment and materials

Decision points

Cooling crystallization, anti-solvent crystallization or reactive crystallization?
Cooling is the candidate for compounds whose solubility changes strongly with temperature, anti-solvent addition for compounds whose solubility does not, and reactive crystallization where a salt is formed and precipitated. Whichever you choose, pin down the conditions under which the target polymorph forms stably by polymorph screening first.
Do you seed, and if so, when and how much?
Seeding is the way to steer the batch to the crystal form you are aiming at and to even out the particle size. Set the point of addition from the solubility curve and the metastable zone (the supersaturation range in which nuclei do not readily form on their own), and set the amount and the seed size by varying them at small scale and looking at the form and the size distribution that come out.
Do you need PAT (process analytical technology) on the crystallization?
On scale-up the agitation and cooling conditions change, and the target polymorph or particle size may not reproduce. Following the supersaturation and the change in the particles with FBRM, PVM (in-situ particle imaging) and in-situ FT-IR makes it easier to isolate the cause when the behaviour shifts with scale.
On the optical resolution route, how does the role of the crystallization change?
A fractional recrystallization that uses the difference in solubility between the diastereomeric salts is run, and the endpoint is judged on chiral purity (optical rotation or chiral HPLC). The salt is then broken with acid or base back to the free form and the final API crystallization is run again, so there is one crystallization more.

Common mistakes

  • Transferring the laboratory conditions straight to the plant, and getting a different polymorph, or crystals far too fine, from the difference in cooling and agitation
  • A more stable polymorph appears later, and the dissolution behaviour of the formulations made until then changes

Check at your own site

  • Over what range of conditions did you run polymorph screening for your API?
  • Have you confirmed that your crystallization conditions can be reproduced with the cooling and agitation of the plant equipment?

Step 7: Filtration and drying

Meet residual solvent and moisture specifications and avoid polymorph transition and decomposition by carefully setting the drying temperature and time.Meet residual solvent and moisture specifications and avoid polymorph transition and decomposition by carefully setting the drying temperature and time.

Standard methodVacuum filtration/centrifugal filtration + vacuum drying (conical/tray)Vacuum filtration/centrifugal filtration + vacuum drying (conical/tray)
Details and 3 alternatives

This step coversRemoval of solvent from wet crystals by filtration and drying (filtration and drying/filter dryer, ANFD)

The crystals produced by crystallization are still in a 'wet' state, containing mother liquor and solvent. They are separated from the liquid and the remaining solvent is driven off to finish a dry API.

After the liquid is removed by vacuum filtration or centrifugal filtration, vacuum drying removes moisture and solvent to within specification. Temperature and time are carefully set in a conical dryer or by tray drying.

Judgment is key here. Inadequate drying puts residual solvent out of specification, and overdoing it can cause the crystal polymorph to transition or decompose.

Crystals obtained by crystallization contain mother liquor and solvent. Filtration removes the liquid, and drying drives off the remaining solvent. Inadequate drying leaves residual solvent and moisture out of specification, and excessive drying can cause polymorph transition or decomposition, so temperature and time are carefully set.

MakersMerck, GEA, Honeywell, Glatt

Why it is the standard. ANFD (agitated Nutsche filter dryer) performs filtration, washing and drying in a single vessel and, being closed, limits cross-contamination and exposure; it is the standard equipment for chemically synthesized APIs. Meeting the ICH Q3C residual solvent limits is built up in the drying step.

Check also. Drying is not mere solvent removal; heating and desolvation can also transition the crystal polymorph or break down hydrates. If temperature and time are pushed too hard, the targeted solid form is lost, so the polymorph is checked after drying in another step, such as XRPD (powder X-ray diffraction).

Alternatives in practice

  • Fluid bed drying · GEA/Glatt fluid bed dryers (GPCG series) — Fast and uniform drying. Watch for fine-powder scattering and static charge.
  • Freeze drying (heat-sensitive/amorphous APIs) · SP VirTis/Millrock freeze dryers (GMP-compliant) — Effective for heat-labile products. High cost and long time.
  • Rotary vacuum drying (double cone) · GEA/Italvacuum double-cone vacuum dryers — Gentle mixing-drying. More product-discharge steps than with ANFD.

Step guide

What you actually do

  1. Send the crystal slurry to an agitated Nutsche filter dryer (ANFD) or a filtering centrifuge and separate the mother liquor
  2. Wash the cake (the layer of crystals built up on the filter medium) with the crystallization solvent or the anti-solvent to push out the impurities left in the mother liquor
  3. Blow the liquid through with nitrogen, then drive the solvent off under vacuum and heat; in an ANFD the agitator blade breaks the cake up while it dries
  4. Confirm the endpoint by loss on drying (LOD) or by GC, and discharge into the container with the system kept closed

Decision points

Do it all in an ANFD, or split it between a filtering centrifuge and a dryer?
An ANFD filters, washes and dries in one closed vessel, which cuts the number of discharges and the exposure of the operators. Where you want stronger deliquoring, or the throughput is large, separating the liquid on a filtering centrifuge and transferring the solids to a vacuum dryer is also chosen.
Displacement wash, or reslurry wash?
Passing solvent down through the cake is fast, but cracks or unevenness in the cake give uneven washing. Where that does not remove enough of the impurity, a reslurry wash that stirs the cake back up in solvent is a candidate, but confirm as well how much crystal dissolves and is lost.
How do you set the drying temperature and time?
On both bringing the residual solvent down to specification and leaving the crystal form unchanged. Heat and desolvation can convert a polymorph or break a hydrate, so confirm the solid form by XRPD (powder X-ray diffraction) whenever you change the conditions.
What do you judge the drying endpoint on?
Following the progress in process by loss on drying (the weight lost on heating), and confirming the individual solvents by headspace GC, is the common combination. ICH Q3C(R9) states that where only Class 3 solvents have been used, a suitably validated loss on drying may be used for the decision.

Common mistakes

  • Estimating the plant from the laboratory filtration time, and filtration taking longer than expected because of the resistance of the cake (the heel) left in the machine
  • Raising the temperature to dry faster, and breaking a hydrate or converting a polymorph

Check at your own site

  • Which solvents does your drying endpoint judge, and with which test?
  • Does your filtration time estimate include the resistance of the cake left in the plant filter?

Formulation and fill-finish(2 steps)

Step 8: Micronization / jet milling

Particle size is directly tied to solubility and formulability. For low-solubility drugs, micronization gains surface area and brings the particles to the target distribution.Particle size is directly tied to solubility and formulability. For low-solubility drugs, micronization gains surface area and brings the particles to the target distribution.

Standard methodMicronization by jet mill (air-jet fine grinding)Micronization by jet mill (air-jet fine grinding)
ReagentsNo grinding media needed. Grinding gas: compressed nitrogen or dry compressed air (pharmaceutical grade, oil-free).
Details and 3 alternatives

This step coversParticle size adjustment by milling and micronization (micronization/jet milling)

A step that finely crushes the particles of dry API to the target size. In a jet mill, the API rides an air stream and the particles collide at high speed to form a fine powder.

Particle size is directly tied to how the drug works. For poorly soluble drugs, making the particles smaller increases surface area and makes the drug dissolve more readily (= be absorbed more readily).

But too small is also a problem. Particles pick up static charge and flowability drops, so bringing them to just the right size distribution is where on-site skill shows.

The particle size of an API is directly tied to how readily it dissolves (that is, how it works) and to how easily it can be handled in the formulation. For poorly soluble drugs, the particles are made smaller to increase surface area. On the other hand, particles that are too small cause static electricity and flowability problems, so they are brought to the target size distribution.

MakersHosokawa Micron, Sturtevant, Fluid Energy

Why it is the standard. The jet mill uses no media, so contamination is low, and it can micronize down to a few µm, making it the standard micronization method for poorly soluble APIs. Being low-temperature and media-free, it also suits heat-sensitive APIs and is widely adopted as pretreatment for the formulation process.

Check also. Micronization such as jet milling can partly amorphize the crystal surface or change the polymorph through mechanical and thermal stress. In addition to bringing the particles to the target size, confirm that the solid form is retained before and after micronization.

Alternatives in practice

  • Pin mill/hammer mill (coarse to medium milling) · Hosokawa Alpine UPZ Fine Impact Mill/Quadro Fitzmill (hammer mill) — Down to tens of µm. Insufficient for sub-10 µm micronization.
  • Wet bead mill (nanonization) · NETZSCH LabStar/Bühler wet bead mill — Effective for preparing nanosuspensions. A drying step and solvent management are added.
  • Particle design by spheronization/crystallization · Controlled crystallization with Mettler Toledo OptiMax (spherical agglomeration crystallization) — No milling needed. The process development burden is heavy.

Step guide

What you actually do

  1. Feed the dried API to the jet mill at a constant rate with a metering feeder
  2. Break the particles against each other in a swirling stream of compressed nitrogen or dry air
  3. Collect the fine particles in a cyclone or on a filter, and return the coarse particles to the milling chamber
  4. Measure the particle size after milling by laser diffraction, and confirm as well that the crystal form has been preserved

Equipment and materials

EquipmentJet millContainment isolatorLaser diffraction particle size analyzerCone mill (delumping and sizing)
PartsContainment transfer parts (split butterfly valves and the like)

Decision points

Is micronization needed at all?
Where poor solubility comes from the dissolution rate, making the particles smaller to increase the surface area works. Smaller particles do not move the equilibrium solubility itself, so for compounds whose poor solubility comes from the strength of the crystal lattice, micronization alone may not be enough.
Jet mill, or pin mill and hammer mill?
Where you want to go down to a few µm, a jet mill is the candidate; it uses no grinding media, so there is little foreign matter and little heat input. Where a few tens of µm is enough, the coarser milling of a pin mill or hammer mill may do.
How do you confirm that milling has not changed the solid form?
The mechanical stress of milling can make part of the crystal surface amorphous, or convert the polymorph. Compare XRPD and DSC before and after milling, and where amorphous material remains, look as well at recrystallization and moisture-driven change during storage.
What containment do you need when milling a highly potent API?
Milling, discharge and container changes are the operations where powder is most easily released. According to the band set from the occupational exposure limit (OEL, the ceiling on the airborne concentration allowed in the workplace), choose anywhere from local exhaust to a containment isolator, and cut the moments in the process at which powder is open.

Common mistakes

  • Looking only at the particle size, missing the amorphous material milling has created, and having the solid form change during storage
  • Milling too fine, so that static makes the powder stick and the recovery and the uniformity of the later blend fall

Check at your own site

  • Was your target particle size for micronization set from dissolution testing?
  • Have you ever measured the amount of amorphous material before and after your milling?

Step 9: Final blending and homogenization

To mix multiple lots after micronization into a final API batch of uniform content and particle sizeTo mix multiple lots after micronization into a final API batch of uniform content and particle size

Standard methodHomogenizing blending of the final API in a bin (container) blender/tumble blenderHomogenizing blending of the final API in a bin (container) blender/tumble blender
ReagentsNone (API only). Carried out under an inert nitrogen atmosphere as needed.
Details and 2 alternatives

This step coversFinal blending and homogenization (batch homogenization) — the micronized API sublots or multiple charges are mixed in a blender to homogenize them into a single final API batch of uniform content, particle size and impurities (final blending and homogenization)

A finishing step that combines the multiple small portions (sublots) or charges after micronization into one and mixes them uniformly.

They are mixed in a bin (container) blender or tumble blender and homogenized into a single final batch in which content, particle size and impurities are the same wherever you sample.

GMP (ICH Q7, etc.) requires uniformity of the whole batch. Mix and even it out, then proceed to release specification testing. This holds down variability within the lot.

GMP (21 CFR 211 and ICH Q7) requires uniformity of the whole API batch, and it is standard for multiple micronized and dried sublots to be mixed and homogenized by final blending before being submitted to release specification testing. This assures content uniformity and limits within-lot variability.

MakersGEA, Quadro

Why it is the standard. Final blending and homogenization is a standard GMP step that assures batch uniformity.

Alternatives in practice

  • Blending after cone mill deagglomeration and sizing · Quadro Comil U20/U30 (conical mill)/Frewitt sizing mill — Breaks up agglomerates and evens out particle size. Improves blend homogeneity.
  • Tumble blending (double cone/V-type) · GEA/Servolift double-cone blender, Patterson-Kelley V-Blender — Gentle mixing. Take care with powders prone to segregation.

Step guide

What you actually do

  1. Delump any agglomerates on a cone mill beforehand
  2. Charge the milled sub-lots or batches into the blender bin in the set order and amounts
  3. Tumble the bin at the set rotation speed and time
  4. Take samples from several positions in the bin, confirm the variation in assay and particle size, and pass the batch on to release testing

Equipment and materials

EquipmentBlender (bin or container type, tumble type)Cone mill (delumping and sizing)Analytical HPLCContainment isolator

Decision points

How do you set the blending time?
Too short leaves non-uniformity, and some powders separate out with longer blending because of the difference in particle size (segregation), so vary the time, measure the variation in assay and particle size at several positions, and set it from the range in which the blend is uniform.
What do you confirm the uniformity with?
Take samples from set positions such as the top, middle and bottom of the bin and compare assay, impurities and particle size. In a warning letter issued by FDA in 2017, a company that blended and shipped powder for homeopathic products was cited for having no in-process test to confirm the uniformity of the blend.
Bin blender, or tumble type (double-cone, V-shape)?
A bin (container) blender lets you blend in the container you move, which cuts powder transfers and exposure. Tumble types blend gently, while powders that segregate easily need their conditions confirmed.

Common mistakes

  • Assuming that the longer you blend the more uniform it gets, and making a powder that segregates easily less uniform instead
  • Judging the blend uniform from a sample taken at a single position

Check at your own site

  • Was your final blending time set by measuring the variation at several positions?

Quality attributes and analytics(11 steps)

Step 10: Structure confirmation (NMR, MS)

Retention time cannot guarantee structure, so spectroscopy and mass spectrometry finally prove that it is the target molecule.Retention time cannot guarantee structure, so spectroscopy and mass spectrometry finally prove that it is the target molecule.

Standard methodStructure confirmation by NMR (1H/13C, plus 19F/31P as needed) + mass spectrometryStructure confirmation by NMR (1H/13C, plus 19F/31P as needed) + mass spectrometry
Details and 3 alternatives

This step coversConfirmation of chemical structure (identification test) (structure confirmation/NMR, MS)

An identification test that finally proves 'whether what was made is really the target molecule.' Chromatographic retention time alone cannot guarantee structure.

Nuclear magnetic resonance (NMR) reads the skeleton and the arrangement of functional groups, and mass spectrometry corroborates the molecular weight. Layering several spectroscopic methods makes the structure certain.

A step that is the linchpin of consistent 'identity confirmation' from the starting materials to the final API. If this wavers, the foundation for discussing purity collapses.

Finally proves that 'what was made is really the target molecule.' HPLC retention time alone cannot guarantee structure, so the skeleton, functional groups and molecular weight are corroborated by several spectroscopic methods. It is the linchpin of identification from the starting materials through to the final API.

MakersAgilent, Waters, Bruker, Honeywell, Sigma-Aldrich

Why it is the standard. NMR is the reference method for primary structure analysis, able to establish structure at the atomic level. MS confirms the molecular weight and provides identification on a different principle. The two are routinely used together for structure confirmation of a new API (ICH Q6A identification test).

Alternatives in practice

  • FT-IR (functional group identification, comparison with reference standard) · Thermo Scientific Nicolet iS20/Bruker ALPHA II — Effective for confirming functional groups and comparing against a reference. Cannot determine the full structure.
  • UV-Vis spectroscopy · Agilent Cary 60/Thermo Evolution UV-Vis — A supplementary check for chromophores.
  • Elemental analysis (CHN) · Elementar vario MICRO cube/Thermo Flash 2000 CHNS — Supports confirmation of the empirical formula. Cannot determine the structure itself.

Step guide

What you actually do

  1. Dissolve the API in a deuterated solvent (a solvent whose hydrogen is replaced by deuterium) and record 1H and 13C spectra by NMR (nuclear magnetic resonance)
  2. Where needed, confirm the connectivity between the atoms by two-dimensional NMR
  3. Measure the accurate molecular mass by high-resolution mass spectrometry (HRMS) and check it against the molecular formula
  4. Compare the FT-IR spectrum with the reference standard, and bring together whether methods based on different principles all point to the same structure

Equipment and materials

Decision points

Which methods do you combine?
A single method cannot rule out compounds of similar structure, so stack methods based on different principles: NMR for the skeleton, mass spectrometry for the molecular mass, infrared absorption for the functional groups. ICH Q6A likewise asks for a combination of highly specific methods for identification.
Do you determine the structure again for every batch?
The structure is determined once in development, and release testing commonly confirms by infrared absorption and the like that the material is the same as the reference standard. When the synthetic route changes, revisit the structure confirmation.
Is a benchtop NMR enough?
Benchtop NMR instruments with permanent magnets can be used to confirm known compounds and to follow a process in line, but their resolution and sensitivity are below those of superconducting-magnet instruments. At the stage of determining a new structure, measurement on a high-field instrument is usual.

Common mistakes

  • Concluding that it is the same compound only because the HPLC retention time matches the reference standard
  • Mistaking the signals of water or residual solvent in the deuterated solvent for an impurity or product signal

Check at your own site

  • How many methods based on different principles does your identification testing stack?

Step 11: Assay and related substances

Quantify the content responsible for efficacy and the related substances from synthesis and degradation to secure the core of quality.Quantify the content responsible for efficacy and the related substances from synthesis and degradation to secure the core of quality.

Standard methodQuantitation by reverse-phase HPLC/UHPLC (UV detection, stability-indicating method)Quantitation by reverse-phase HPLC/UHPLC (UV detection, stability-indicating method)
Details and 3 alternatives

This step coversAssay (content) and quantitation of organic impurities and related substances (assay and related substances/HPLC quantitation)

The main event of quality testing, measuring the core of API purity. It looks at the 'content (assay)' responsible for efficacy and the related substances that get mixed in, together.

Content is quantified by reverse-phase HPLC/UHPLC, and related substances from synthesis and degradation are separated and tracked one by one. This reveals which impurities are present and how much.

Stability studies also follow how degradation products increase, so a 'stability-indicating method' that can properly separate degradation products is indispensable. The story of purity is told here.

How pure the API is, and which impurities are present and how much, is the core of quality. The content responsible for efficacy is measured, and related substances derived from synthesis and degradation are followed one by one. Stability studies also look at how degradation products increase, so a method that can separate the degradation products is essential.

MakersMerck, Agilent, Waters, Metrohm, Honeywell

Why it is the standard. The industry standard based on ICH Q3A (impurities in drug substances), Q6A and Q2 (analytical method validation). HPLC/UHPLC-UV can quantify content and related substances at the same time, and stability-indicating capability is assured by validation (many method development examples such as PMC11371568).

Check also. UV detection in reverse-phase HPLC works well for impurities with a chromophore, but it misses impurities that barely absorb UV, as well as inorganic salts and counterions. When the mass balance does not close, complement it with other detection principles such as CAD/ELSD (evaporative light scattering) or mass spectrometry.

Alternatives in practice

  • LC-MS/MS (assignment and quantitation of trace impurities) · Sciex Triple Quad 7500/Waters Xevo TQ-XS — Effective for impurities with weak UV absorption and for structural assignment.
  • Ion chromatography · Thermo Scientific Dionex Integrion/Metrohm 940 Professional IC — For quantifying counterions and inorganic ions.
  • CAD/ELSD detection · Thermo Scientific Vanquish Charged Aerosol Detector (CAD)/Agilent 1290 Infinity II ELSD — General-purpose quantitation of compounds without UV absorption.

Step guide

What you actually do

  1. Dissolve the API and the reference standard and prepare them at the set concentrations
  2. Inject onto a reverse-phase column (C18 and the like, which separates the components on the difference in hydrophobicity) and elute with a gradient
  3. Detect each peak by UV (DAD) and confirm the system suitability (resolution, repeatability)
  4. Calculate the assay against the reference standard and the amount of the related substances from the peak areas, and compare them with the specification

Equipment and materials

Decision points

Where do the thresholds for related substances come from?
ICH Q3A draws the lines on the maximum dose per 1 day. At 2 g or less, the reporting threshold is 0.05%, the identification threshold 0.10% (or 1.0 mg/day, whichever is lower) and the qualification threshold 0.15% (or 1.0 mg/day, whichever is lower). Without a dose you cannot draw the lines, and if the dose rises during development you redraw them.
Can you report amounts as area percent (area%) as they are?
Area% assumes that the UV sensitivity of the main compound and of the impurity are the same. The European Pharmacopoeia approach treats a sensitivity ratio (response factor) of 0.8 to 1.2 as needing no correction and applies a correction factor to impurities outside that range, so confirm the response factors of the main impurities against standards.
How do you show that the method does not miss degradation products?
Run forced degradation that deliberately degrades the material with heat, light, acid, base and oxidation, and show that the degradation products formed separate cleanly from the main compound. Degrading the main compound by around 5 to 20% is the usual target; push it harder and you generate secondary degradation products that never occur in practice.
What do you do about impurities UV cannot see?
Impurities and counter-ions that barely absorb UV are missed, so where the mass balance does not close, supplement with CAD or ELSD (methods that nebulize and dry the liquid and detect the resulting particles by charge or by light scattering), or with mass spectrometry.

Common mistakes

  • Reading area% values as weight ratios without having confirmed the response factors
  • Taking development forward with a method whose quantitation limit does not reach the reporting threshold, and having to rebuild it later

Check at your own site

  • Have the response factors of your main related substances been confirmed against standards?
  • Where does the quantitation limit of your method sit relative to the reporting threshold?

Step 12: Ultra-trace nitrosamine analysis

Reliably control impurities that are carcinogenic even in trace amounts at ppb levels, orders of magnitude lower than ordinary tests.Reliably control impurities that are carcinogenic even in trace amounts at ppb levels, orders of magnitude lower than ordinary tests.

Standard methodUltra-trace quantitation by LC-MS/MS (or GC-MS/MS)Ultra-trace quantitation by LC-MS/MS (or GC-MS/MS)
Details and 3 alternatives

This step coversUltra-trace control of mutagenic impurities and nitrosamines (ultra-trace nitrosamine analysis)

A special control step that watches for impurities that are carcinogenic even in trace amounts at concentrations orders of magnitude lower. It deals with a range that ordinary purity testing cannot reach.

Alkylating agents and nitrosamines are the leading examples. They are quantified at the parts-per-billion (ppb) level using methods with extremely low detection limits, such as LC-MS/MS.

Since nitrosamine contamination was discovered in the 2018 sartan drug product (antihypertensive) problem, regulation has tightened all at once. It is a test at the front line of safety.

Impurities that carry a carcinogenic risk even at very low levels need to be controlled at ppb levels, orders of magnitude lower than in ordinary purity testing. Alkylating agents and nitrosamines are representative, and regulation has tightened since the 2018 sartan drug product problem. They are reliably controlled with methods that have extremely low detection limits.

MakersAgilent, Waters, Honeywell, Sigma-Aldrich

Why it is the standard. Control based on ICH M7 (mutagenic impurities). Nitrosamines and genotoxic impurities require ppb-level control, and LC-MS/MS is the standard for its high selectivity and sensitivity (consistent with the methods of LCGC/Shimadzu/PMC12655250). HS-GC-MS/ECD is also used for volatile alkyl halides.

Check also. For mutagenic impurities and nitrosamines, the practice is first to assess from the synthetic route and reagents whether they could form, as a structural alert, and then to measure at the ppb level only the compounds that are needed. Since the 2018 sartan drug product problem, regulation (ICH M7, etc.) and the control targets continue to be updated, so operate in line with the latest notifications.

Alternatives in practice

  • HS-GC-MS (volatile nitrosamines) · Agilent 8890 GC/7000E MS+7697A headspace — Effective for volatile low-molecular-weight nitrosamines.
  • GC-ECD (volatile alkyl halides, etc.) · Agilent 8890 GC+ECD detector — High-sensitivity detection of halogenated compounds. Limited versatility.
  • HRMS screening/process purge assessment · Thermo Orbitrap Exploris/Waters Xevo G3 QTof (HRMS) — Supports the search for unknown nitroso compounds and measured purge factors.

Step guide

What you actually do

  1. List the starting materials, reagents, intermediates and by-products of the synthetic route and assess them with (Q)SAR (prediction of mutagenicity from the structure)
  2. Work out, from the steps where an amine and nitrite coexist, which nitrosamines could form
  3. Set an acceptable intake for each compound of concern, and work back from it to the quantitation limit you need
  4. Add a stable-isotope-labelled internal standard, prepare the sample and quantify by LC-MS/MS or GC-MS/MS

Decision points

How do you narrow down what to measure?
ICH M7 asks you to run 2 kinds of (Q)SAR, one rule-based and one statistics-based, and where both are negative the impurity can in principle be treated as of no concern. Where the results disagree, an expert review follows, and where that still does not settle it, measure in a bacterial reverse mutation assay (Ames test). A negative prediction is not positive proof that the substance is not mutagenic.
How is the acceptable limit set?
For mutagenic impurities the baseline is the TTC (threshold of toxicological concern) of 1.5 µg per day, relaxed for shorter treatment durations: within 1 month it is 120 µg/day. Nitrosamines are stricter still; NDMA, for example, is 96 ng/day. For nitrosamines derived from the structure of the API (NDSRI), the limit is set with approaches such as the carcinogenic potency categorization approach (CPCA), which assigns a potency category from the structure.
LC-MS/MS, or GC-MS/MS?
NDSRI, which are not volatile, are usually measured by LC-MS/MS or high-resolution LC-MS, and small volatile nitrosamines and alkyl halides by GC-based methods. You are measuring traces in the presence of a large amount of API, so building the sample preparation and the separation conditions is the core of the method development.
If the process removes it, can you leave the final API untested?
ICH M7 also accepts a control strategy without testing of the final product, provided you can show removal in the process with measured data. The removal data is the premise; how far you rely on it is for the quality unit to judge and is settled in agreement with the regulatory authority.

Common mistakes

  • Treating a negative (Q)SAR result as proof of no mutagenicity, and settling by prediction what should be measured
  • An impurity nobody anticipated appears in neither the specification nor the test methods, and manufacturing continues with nobody measuring it

Check at your own site

  • Have you worked out every step in your synthetic route where an amine and nitrite could be present at the same time?
  • Was your NDSRI limit set from compound-specific data, from read-across or from CPCA?

Step 13: Residual solvents / headspace GC

Quantify residual solvents, whose limits differ with toxicity, as a final check of drying adequacy and safety.Quantify residual solvents, whose limits differ with toxicity, as a final check of drying adequacy and safety.

Standard methodQuantitation of residual solvents by headspace GC-FID (HS-GC)Quantitation of residual solvents by headspace GC-FID (HS-GC)
Details and 3 alternatives

This step coversQuantitation of residual organic solvents (residual solvents/headspace GC)

A safety test that confirms no trace of the solvents used in synthesis and crystallization remains in the API. It also serves as a final check that drying was sufficient.

Quantified by headspace gas chromatography (HS-GC), which warms the sample and captures the vaporized solvent. It efficiently picks up readily volatile solvents.

Toxicity differs by solvent type, so stricter limits are set for the more toxic ones. Pass/fail is judged by comparing the residual amount with the limit.

Solvents used in synthesis and crystallization remain in the API in trace amounts. Toxicity differs by type, so the more toxic the solvent, the stricter the limit set. It also serves as a final check that drying was sufficient, making it a test directly tied to safety.

MakersMerck, Agilent, Shimadzu, PerkinElmer, Metrohm

Why it is the standard. Standard method based on ICH Q3C (residual solvents) and USP<467>. HS-GC-FID separates volatile solvents from a nonvolatile matrix and quantifies them with high sensitivity, and is the method specified in the pharmacopoeias (consistent with methods such as PMC12614197).

Alternatives in practice

  • HS-GC-MS (when identification is needed) · Agilent 8890 GC/5977B MSD+7697A headspace — Effective for identifying peaks of unknown solvents. Quantitative reproducibility is somewhat inferior to FID.
  • Direct injection GC · Agilent 8890 GC-FID (direct liquid injection) — For high-boiling, hard-to-volatilize solvents. Watch for matrix contamination.
  • Karl Fischer (water measured separately) · Metrohm 851 Titrando/Mettler Toledo C20S — Residual water cannot be measured by GC, so a separate method is needed.

Step guide

What you actually do

  1. Dissolve the API in a diluent (water, DMSO, DMF and the like) and place it in a sealed vial
  2. Heat the vial and inject onto the GC (gas chromatograph) only the solvent that has come out into the gas phase
  3. Detect each solvent with a flame ionization detector (FID) and calculate the amount against the standard solutions
  4. Compare with the limit for each solvent (ICH Q3C), and identify unknown peaks by GC-MS

Equipment and materials

Decision points

Can the limits stay as the ppm values in the guideline table?
The ppm values in the table assume a dose of 10 g per day (Option 1). Once the actual maximum daily dose is known, Option 2, which converts directly from the PDE (permitted daily exposure), can be used; where that value exceeds Option 1, you have to show that the solvent has been reduced to the practical minimum.
Do you look for every solvent, or measure only the ones you used?
USP <467> splits the work into Procedure A, which looks for the presence of solvents on a G43 phase column, Procedure B, which confirms on a phase of different polarity, and Procedure C, which quantifies; where the solvents used are known, Procedure C alone is taken to be enough. Class 1 solvents are as a rule not used, and where one has had to be, you quantify to show that it has come below the limit.
Can loss on drying (LOD) take its place?
ICH Q3C(R9) states explicitly that where only Class 3 solvents have been used, a non-specific method such as a suitably validated loss on drying may be used for the decision. That validation has to allow for the effect of the volatility of the solvent on the result.
What do you choose as the diluent?
Water is common for APIs that dissolve in water, and DMF or DMSO for those that do not. DMSO can elute at the same position as solvents such as DMF in headspace work, so take care in the choice where DMF is also an analyte.

Common mistakes

  • The peak of the diluent overlaps the solvent of interest, and the solvents are mixed up
  • Believing that ICH Q3C has a "Class 4" category, and not providing the safety justification for a solvent with insufficient data

Check at your own site

  • Are your residual solvent limits set on Option 1 or on Option 2?

Step 14: Elemental impurities (ICP-MS)◇ Branches

Control metal catalyst residues, toxic even in trace amounts, within permitted limits and make a final check that removal was sufficient.Precious metal catalyst residues drop out of the control scope

Standard methodElemental quantitation by ICP-MS (microwave acid digestion pretreatment)Elemental quantitation by ICP-MS (microwave acid digestion pretreatment)
EquipmentICP-MS and ICP-OESICP-MS (microwave acid digestion pretreatment)
ReagentsAPI reference standards and impurity standards (CRS)Because the resolution route does not use asymmetric metal catalysts (Rh/Ru, etc.), monitoring of precious metal residues of the ICH Q3D classes such as Ru/Rh is normally unnecessary. Instead, the focus shifts to elements that can be introduced in the salt formation and decomposition steps (such as Na) and to those derived from the equipment used
With Racemic synthesis + optical resolution (diastereomeric salt crystallization)Precious metal catalyst residues drop out of the control scopeWith Asymmetric synthesis (direct construction of a single enantiomer)Elemental quantitation by ICP-MS (microwave acid digestion pretreatment)
Details and 3 alternatives

This step coversQuantitation of elemental impurities and metal catalyst residues (elemental impurities/ICP-MS)

A step that measures the 'leftovers' of metal catalysts used in synthesis. If palladium, nickel and the like remain in the API even in trace amounts, toxicity becomes a concern.

After pretreatment in which the sample is dissolved in acid, elements are quantified by inductively coupled plasma mass spectrometry (ICP-MS). Even trace metals are picked up accurately.

Permitted amounts are set for each route of administration. The elements specified by regulation are reliably measured, providing final support that catalyst removal was sufficient.

If metal catalysts used in synthesis, such as palladium and nickel, remain in the API even in trace amounts, toxicity becomes a concern. Permitted amounts are set for each route of administration, so the elements specified by regulation are reliably measured. It is also the final check that catalyst removal was sufficient.

MakersMerck, Agilent, PerkinElmer, Honeywell

Why it is the standard. Standard method based on ICH Q3D (elemental impurities) and USP<232>/<233>. ICP-MS quantifies many elements simultaneously from ppb to ppt and covers catalyst residues (Pd/Pt/Ni, etc.) and environment-derived metals (As/Cd/Hg/Pb) (consistent with the methods of PMC9433241/PMC10987250).

Check also. ICP-MS measures the total amount of an element but does not distinguish the chemical form (valence, species) that determines toxicity. Permitted amounts are set by ICH Q3D risk assessment for each route of administration, and the measurement is used only as a final check that catalyst residue removal was sufficient.

With Racemic synthesis + optical resolution (diastereomeric salt crystallization). In contrast, on the asymmetric synthesis side, Ru/Rh/Pd residues derived from catalysts such as BINAP-Ru/Rh are the main targets of ICP-MS monitoring

Alternatives in practice

  • ICP-OES (for high-concentration elements) · Agilent 5800/5900 ICP-OES/Thermo iCAP PRO — Simple and robust for high-concentration elements. Trace-level (ppb) sensitivity is inferior to ICP-MS.
  • AAS (atomic absorption) · Agilent 240/280 AA/PerkinElmer PinAAcle 900 — For single-element targets. Simultaneous multi-element measurement is not possible.
  • Colorimetric methods/simple kits (limit tests) · Merck heavy metal limit test kits/Agilent Cary 60 UV-Vis — Simple judgment in the manner of a limit test. Low quantitative accuracy and selectivity.

Step guide

What you actually do

  1. Work out in a risk assessment which elements could enter, and from where: the catalyst, the raw materials, the equipment, the containers
  2. Put the API into high-purity acid and digest the organic matter in a microwave digestion system to give a solution
  3. Quantify the elements of interest by ICP-MS (inductively coupled plasma mass spectrometry), correcting the drift in sensitivity with an internal standard
  4. Confirm the recovery of the digestion and the measurement with samples spiked with known amounts of the elements

What changes with the conditionWhether monitoring centres on the precious metals of an asymmetric catalyst, or on elements from salt formation and breaking and from the equipment (in asymmetric synthesis, catalyst residues such as Ru, Rh and Pd are the main targets; the optical resolution route uses no chiral metal catalyst, so the focus moves to elements that enter with salt formation and breaking, such as Na, and to elements from the equipment)

Decision points

How do you decide which elements to measure?
ICH Q3D asks you to start from a risk assessment rather than from the analysis. Metals from catalysts used intentionally are measured from the outset; for the rest, narrow the list by working through the process flow and the formulation to find entry from the raw materials, the equipment and the containers.
Measure every lot, or settle it with a one-off assessment?
30% of the permitted daily exposure (PDE) is taken as the control threshold, and where you stay below it and the element is under control, no additional control is held to be needed. Where the threshold is exceeded, decide between removal in the process, a change of raw material and control in the specification.
What changes the permitted limit?
The route of administration moves it by orders of magnitude: for mercury, for example, the oral limit is 30 µg/day against 3 µg/day parenteral and 1 µg/day inhalation. When you take an oral product on to an injectable, redo the assessment.
How do the elements you measure change with the synthetic route?
In asymmetric synthesis, the Ru, Rh and Pd from chiral catalysts such as BINAP-Ru and Rh complexes are the main monitoring targets. The racemic synthesis and optical resolution route uses no chiral metal catalyst, so precious metals that are not used can be taken off the list where the risk assessment can show that they are not introduced. If other steps use Pd couplings and the like, those catalysts are assessed separately. On that basis, the emphasis moves to elements that can enter with salt formation and breaking (Na and the like) and to elements from the equipment.

Common mistakes

  • Incomplete digestion lowers the recovery, and the value reported is lower than the real one
  • Not working out the entry from equipment and containers, which do not appear in the formulation, so that it is never found

Check at your own site

  • Does your elemental impurity risk assessment include entry from the equipment and the containers?
  • Where do your catalyst metals sit against 30% of the permitted limit?

Step 15: Chiral purity / chiral HPLC◇ Branches

An unwanted enantiomer changes efficacy and safety, so it is separated and quantified to assure the success of asymmetric synthesis.Chiral HPLC is promoted to the rate-determining IPC

Standard methodEnantiomer separation and quantitation by chiral HPLC (chiral stationary phase, CSP)Chiral HPLC (enantiomer separation and ee quantitation with a chiral stationary phase, CSP). On the resolution route, it is used as a rate-determining process control item for building up ee (endpoint determination of fractional recrystallization, in-process checks after each crystallization)
EquipmentChromatography systemPolysaccharide chiral columnChiral HPLC (polysaccharide CSP: amylose/cellulose derivative columns, etc.)
With Racemic synthesis + optical resolution (diastereomeric salt crystallization)Chiral HPLC (enantiomer separation and ee quantitation with a chiral stationary phase, CSP). On the resolution route, it is used as a rate-determining process control item for building up ee (endpoint determination of fractional recrystallization, in-process checks after each crystallization)With Asymmetric synthesis (direct construction of a single enantiomer)Enantiomer separation and quantitation by chiral HPLC (chiral stationary phase, CSP)
Details and 3 alternatives

This step coversConfirmation of chiral purity and enantiomeric excess (ee) (chiral purity/chiral HPLC)

A test to confirm that the molecule that is the mirror image (enantiomer), with left and right reversed as if reflected in a mirror, has not been mixed in on the unwanted side. It is essential for optically active APIs.

Enantiomers, related like left and right hands, can differ in efficacy and safety. The two are separated by HPLC with a chiral stationary phase and quantified as an impurity.

Did asymmetric synthesis or chiral resolution work as intended? A stereochemical quality check that assures the result in the form of enantiomeric excess (ee).

In an optically active API, if the enantiomer (a molecule related like left and right hands) is mixed in, efficacy and safety change. The unwanted enantiomer is separated and quantified as one impurity. This test assures that asymmetric synthesis or chiral resolution worked as intended.

MakersDaicel, Agilent, Waters, Merck, Honeywell

Why it is the standard. Chiral HPLC (polysaccharide CSP) is the most established technique for enantiomer separation and corresponds to the chiral API specification (enantiomeric purity) of ICH Q6A. The Daicel CHIRALPAK series is used as the industry standard.

With Racemic synthesis + optical resolution (diastereomeric salt crystallization). On the asymmetric synthesis side, ee is set by the catalyst, and chiral HPLC is mainly a confirmatory test at release. On the resolution side, it becomes a key analysis that governs the process

Alternatives in practice

  • Chiral SFC (supercritical fluid chromatography) · Waters ACQUITY UPC2/Agilent 1260 Infinity II SFC+Daicel CHIRALPAK columns — Fast and low in solvent, and also advantageous for preparative work. Requires equipment investment.
  • Optical rotation measurement · Anton Paar MCP 5100/JASCO P-2000 polarimeters — Simple, but unsuited to high-precision quantitation of enantiomeric excess.
  • Chiral GC · Agilent 8890 GC + cyclodextrin-based chiral column (Agilent CP-Chirasil-Dex CB) — Effective for volatile chiral compounds. Limited range of application.

Step guide

What you actually do

  1. Confirm the separation and the assignment of the peaks with a racemate and with standards of both enantiomers
  2. Inject onto a column with a polysaccharide-based chiral stationary phase (CSP) and separate the enantiomers under normal-phase or reverse-phase conditions
  3. Calculate the proportion from the peak area of the unwanted enantiomer, and from it the ee
  4. Compare with the specification. On the optical resolution route, measure after each crystallization as well, to decide whether the process may go on

What changes with the conditionWhether chiral HPLC is a confirmation test at release, or an in-process test that governs the process (in asymmetric synthesis the ee is set by the catalyst and confirmation at release is the centre of it; in optical resolution it judges the endpoint of the fractional recrystallization and serves as the IPC after each crystallization, deciding whether the process moves on)

Equipment and materials

EquipmentAnalytical HPLCPolarimeter

Decision points

Coated column, or immobilized?
On a coated column the selector (the polysaccharide derivative that discriminates the enantiomers) dissolves out in strong solvents such as THF and dichloromethane, so even traces carried in with the sample diluent or the autosampler wash will degrade the performance over time. Where there is any chance of using strong solvents, narrow the candidates to immobilized columns.
When it does not separate, will a smaller particle size separate it?
What decides whether the separation is possible is not the particle size but the backbone of the selector (cellulose-based or amylose-based) and the solvent conditions. Cellulose-based and amylose-based phases are complementary in selectivity, so screen with a set that includes both before settling the particle size and the column dimensions.
Is this test a confirmation at release, or a test that governs the process?
In asymmetric synthesis the ee is set by the catalyst and chiral HPLC is mainly a confirmation test at release. On the optical resolution route it is used to judge the endpoint of the fractional recrystallization and as an in-process test after each crystallization, so it is treated as a critical analysis that decides whether the process goes on.

Common mistakes

  • Letting small amounts of strong solvent from the sample diluent or the wash keep flowing onto a coated column, and losing its performance without noticing
  • Not confirming the separation from the related isomers, and overestimating the enantiomeric purity

Check at your own site

  • Is there any THF or dichloromethane in the sample diluent or the wash used in your chiral analysis?
  • Was your chiral column chosen after trying both cellulose-based and amylose-based phases?

Step 16: Crystal polymorph

Polymorph differences change solubility, stability and absorption, so consistent manufacture of the target polymorph is confirmed lot by lot.Polymorph differences change solubility, stability and absorption, so consistent manufacture of the target polymorph is confirmed lot by lot.

Standard methodPolymorph identification by powder X-ray diffraction (XRPD)Polymorph identification by powder X-ray diffraction (XRPD)
ReagentsNo reagents required (Cu Kα source, λ=1.5406 Å). Zero-background sample holder (single-crystal Si) used.
Details and 3 alternatives

This step coversConfirmation of crystal polymorph and crystallinity (crystal polymorph/powder X-ray diffraction, XRPD)

Even for the same molecule, a different arrangement within the crystal (polymorph) makes it behave like a different substance. Ease of dissolution, stability and absorption change.

Powder X-ray diffraction (XRPD) reads the arrangement of the crystal like a fingerprint and confirms that the target polymorph is being made consistently.

Transition to an unwanted polymorph or to the amorphous form leads directly to problems with efficacy and stability. That is why the identity of the crystal is watched lot by lot.

Even for the same molecule, a different arrangement of the crystal (polymorph) changes solubility, stability and absorption. Whether the target polymorph is being produced consistently is confirmed like a fingerprint. Transition to an unwanted polymorph or to the amorphous form leads directly to problems with efficacy and stability, so it is checked lot by lot.

MakersMalvern Panalytical, Bruker, Rigaku

Why it is the standard. XRPD is the reference method for polymorph identification, giving a fingerprint of the crystal structure. Crystals show sharp Bragg peaks and amorphous material a broad halo, and it is standard both for polymorph control under ICH Q6A and for patent protection (consistent with Malvern Panalytical's explanations).

Alternatives in practice

  • Solid-state NMR (ss-NMR) · Bruker AVANCE NEO 400 WB+CP-MAS probe — Strong at discriminating local structure of polymorphs and amorphous forms. Instrumentation and analysis are advanced.
  • Raman/FT-IR spectroscopy · Thermo DXR3 Raman/Bruker ALPHA II FT-IR — Effective for polymorph screening and mapping. Quantitative polymorph ratios require calibration.
  • Polarized light microscopy (PLM) · Olympus BX53P polarizing microscope/Leica DM2700 P — Confirmation of crystallinity and crystal habit. Supports polymorph identification.

Step guide

What you actually do

  1. Pack the API into the sample holder and level the surface
  2. Irradiate with X-rays and measure the intensity of the diffraction while varying the angle, to obtain the diffraction pattern
  3. Compare the positions and the intensities of the peaks with the reference pattern of the target polymorph
  4. Check that no peaks of another polymorph, and no amorphous halo (a broad, gentle rise), are present

Decision points

Can you judge the polymorph on XRPD alone?
XRPD is the basic method for identifying a crystal form, but preferred orientation of the crystals changes the intensity ratios, and very small amounts of another polymorph or of amorphous material can be missed. Where it bears on quality, supplement with DSC, Raman or solid-state NMR.
Do you put the crystal form in the specification?
ICH Q6A gives a decision tree for judging whether the difference between polymorphs bears on the performance of the product. Where it does, control the crystal form in the specification; where it does not, you may decide not to, but keep the justification on record either way.
Do you quantify small amounts of another polymorph?
Where the admixture bears on the product, calibrate with samples blended at known ratios and confirm how far you can detect and quantify it. The lower limit of detection varies with the pair of polymorphs and with the instrument, so establish it on your own samples.

Common mistakes

  • Reading intensity ratios changed by the orientation of the sample as the admixture of another polymorph
  • Declaring small amounts of amorphous material or of another polymorph "absent" without having confirmed the limit of detection

Check at your own site

  • Have you confirmed at what level another polymorph becomes detectable by your XRPD?

Step 17: Thermal analysis and water (DSC, TGA, KF)

Grasp polymorph stability and hydration/solvation from thermal behavior, and the effect of water content on stability and content calculation.Grasp polymorph stability and hydration/solvation from thermal behavior, and the effect of water content on stability and content calculation.

Standard methodDSC+TGA (thermal analysis) + Karl Fischer titration (water)DSC+TGA (thermal analysis) + Karl Fischer titration (water)
ReagentsDSC/TGA: no reagents required (high-purity nitrogen purge, Al/Pt crucibles). KF: Honeywell Hydranal reagents (Composite 5/2 for volumetric, Coulomat AG for coulometric).
Details and 3 alternatives

This step coversConfirmation of melting point, thermal behavior and water/solvation (thermal analysis and water/DSC, TGA, KF)

A thermal analysis step that reads the 'behavior' of the API when heated. From the melting temperature, transition temperature and weight changes, it probes polymorph stability and whether the material is a hydrate or solvate.

Differential scanning calorimetry (DSC) follows the heat flow in and out, thermogravimetry (TGA) follows the loss of weight, and Karl Fischer titration measures water at the same time.

Water is controlled separately. Because it affects stability and content calculation, it is separated from thermal behavior and measured accurately.

The temperatures of melting and transitions on heating, and changes in weight, tell whether a polymorph is stable and whether it is a hydrate or solvate. Water content is also measured accurately. Water affects stability and content calculation, so it is controlled separately.

MakersTA Instruments, Mettler Toledo, Metrohm, Honeywell

Why it is the standard. The standard combination: DSC evaluates melting point, transitions and polymorph conversion, and TGA evaluates solvation/hydration and thermal stability (consistent with Agno/Malvern explanations). For water, KF titration is the reference method specified by the pharmacopoeia (USP<921>).

Alternatives in practice

  • Dynamic vapor sorption (DVS) · Surface Measurement Systems DVS Intrinsic/TA Instruments Q5000 SA — Effective for understanding hygroscopicity and hydrate formation behavior.
  • Hot-stage microscopy (thermal microscopy) · Linkam THMS600 hot stage + Olympus polarizing microscope/Mettler Toledo FP90 — Melting and transitions are confirmed visually. Quantitative heat measurement depends on DSC.
  • Loss on drying (LOD) · Mettler Toledo HC103 halogen moisture analyzer/Sartorius MA37 infrared moisture analyzer — Simple measurement of volatiles. Cannot distinguish water from solvent.

Step guide

What you actually do

  1. Weigh out a small amount of the API, place it in a DSC pan and seal it
  2. Heat at a constant rate in DSC (differential scanning calorimetry) and record the heat flow of melting and of the transitions
  3. Follow the weight loss while heating in TGA (thermogravimetric analysis), and see at what temperature and how much water or solvent leaves
  4. Measure the water separately by Karl Fischer titration and judge, against the TGA weight loss, whether the material is a hydrate or a solvate

Equipment and materials

Decision points

Is the TGA weight loss water, or solvent?
TGA does not tell you what left, so compare it with the water content from Karl Fischer titration, and where a difference remains, consider solvent. Confirm the solvent by GC where needed.
Is the DSC pan sealed, pinhole or open?
How the pan is closed decides whether the water or solvent that leaves can escape, and that moves the shape and the temperature of the peaks. Measure samples you mean to compare with the same pan and the same heating rate.
How far do you investigate hygroscopicity?
For APIs that form a hydrate with humidity, or whose amorphous material returns to the crystalline state, look at the moisture uptake behaviour by dynamic vapour sorption (DVS, measuring the change in weight while varying the humidity).

Common mistakes

  • Reading the whole TGA weight loss as water, and missing a solvate
  • Comparing results obtained with different pans or heating rates, and reading the difference in transition temperature as a difference in polymorph

Check at your own site

  • Do you read your TGA weight loss against the Karl Fischer water content?

Step 18: Water determination (Karl Fischer)

To control the API water content by specification and assure hydration state, hygroscopicity and stabilityTo control the API water content by specification and assure hydration state, hygroscopicity and stability

Standard methodCoulometric Karl Fischer method (absolute quantitation of trace water, with a heating evaporation oven as needed)Coulometric Karl Fischer method (absolute quantitation of trace water, with a heating evaporation oven as needed)
ReagentsHoneywell Hydranal Coulomat AG (coulometric anolyte, product number 34836), also usable in cells without a diaphragm. When used with an oven, Hydranal Coulomat AG-Oven (product number 34739).
Details and 2 alternatives

This step coversQuantitation of water (water content) — water in the API is quantified by the Karl Fischer method to confirm hydrate state, hygroscopicity and compliance with the loss-on-drying specification (water determination/Karl Fischer)

A step that measures the water contained in the API itself, as an absolute amount. The water content required in the release specification is established here.

Trace water is directly quantified by coulometric Karl Fischer titration. If needed, an oven that heats the sample to drive out the water is used with it, picking up even tightly held water.

Used to judge hydrate stoichiometry, hygroscopicity and compliance with the loss-on-drying specification. Separately from thermal analysis (DSC/TGA), it is a test that treats water as the main subject.

Water (water content) is a mandatory item in API release specifications, and trace water is quantified directly by the Karl Fischer method. Separately from the thermal behavior seen by DSC/TGA, it is used to judge hydrate stoichiometry, hygroscopicity and compliance with the loss-on-drying specification.

MakersMetrohm, Honeywell

Why it is the standard. Karl Fischer water determination is mandatory in the standard API release tests.

Alternatives in practice

  • Volumetric Karl Fischer method (for items with high water content) · Metrohm 901 Titrando/Mettler Toledo V20S+Hydranal Composite 5 — Suitable for items with water of 1% or more. Higher upper limit than the coulometric method.
  • Loss on drying (LOD) · Mettler Toledo HC103 halogen moisture analyzer/vacuum dryer + electronic balance — Simple, but includes volatile solvents and is not water-specific.

Step guide

What you actually do

  1. Bring the titration cell to a dry state and wait for the drift (the amount of water finding its way into the cell) to settle
  2. Weigh out the API and either add it to the cell directly, or heat it in a moisture vaporizer (oven) so that a dry carrier gas brings only the water across
  3. Titrate until the water no longer consumes iodine, and calculate the water content from the consumption
  4. Confirm the state of the instrument and the reagents with a water standard, and compare the result with the specification

Equipment and materials

Materials and reagentsKarl Fischer reagents and water standards

Decision points

Volumetric, or coulometric?
The volumetric method adds reagent of known concentration and measures water quickly, from the mg level to the percent level, but the titer of the reagent has to be checked periodically. The coulometric method generates iodine by electrolysis, so no titer determination is needed and it suits trace water, while samples with a high water content take time.
Direct titration, or the vaporizer method?
For APIs that dissolve poorly in the solvent, or that contain components such as ketones that side-react with the reagent, the vaporizer method, which drives only the water out in an oven and carries it across, is a candidate. Too high a heating temperature degrades the sample and reads the water high, too low and not all of the water comes out, so set the temperature for each sample by measurement.
Which method does "Karl Fischer" in the specification mean?
USP <921> splits it into the volumetric method (Method Ia), back titration (Ib) and the coulometric method (Ic), and the European Pharmacopoeia likewise treats the volumetric and the coulometric method in separate sections. Decide from the specification side which method you test by, before you choose the instrument.

Common mistakes

  • Setting the vaporizer temperature too high, so that the sample degrades and the water reads high
  • Choosing the instrument on the width of its measuring range alone, without confirming the recovery and the variation on your own samples

Check at your own site

  • Which Method of USP <921> does the Karl Fischer test in your specification correspond to?

Step 19: Particle size distribution

Particle size variation affects solubility, flowability and drug product uniformity, so it is controlled as a critical quality attribute.Particle size variation affects solubility, flowability and drug product uniformity, so it is controlled as a critical quality attribute.

Standard methodParticle size distribution measurement by laser diffraction (Dv10/Dv50/Dv90)Particle size distribution measurement by laser diffraction (Dv10/Dv50/Dv90)
EquipmentLaser diffraction particle size analyzer: Malvern Panalytical Mastersizer 3000 (Aero dry/Hydro wet units)/Sympatec HELOS+RODOS/Microtrac SYNC.
ReagentsDry: oil-free compressed air/nitrogen. Wet: dispersion medium (water or isoparaffin-based, with surfactant added, e.g., Tween 20/lecithin).
Details and 3 alternatives

This step coversConfirmation of physical properties related to formulability, such as particle size distribution and specific surface area (particle size distribution/laser diffraction)

A physical property test that measures variation in particle size. Particle size affects dissolution, flowability and the uniformity of the final drug product, so it is an important quality attribute related to formulability.

The particle size distribution is measured by laser diffraction, and Dv10, Dv50 and Dv90 are read from the cumulative distribution from the small side. It confirms whether micronization gave the target distribution.

Especially for poorly soluble drugs, particle size is directly tied to variation in effect. That is why it is controlled as a critical quality attribute, down to the shape of the distribution.

Variation in particle size affects dissolution, flowability and the uniformity of the final drug product. Measure whether micronization gave the target distribution and, if needed, look at surface area as well. For poorly soluble drugs, particle size is directly tied to variation in effect, so it is controlled as a critical quality attribute.

MakersMalvern Panalytical, Sympatec, Microtrac

Why it is the standard. Laser diffraction is the industry standard that gives the particle size distribution as Dv10/50/90 quickly and with high reproducibility, and is widely adopted for particle size control of micronized APIs (the relevant ICH Q6A specification) (consistent with the practice of SK pharmteco/Malvern).

Alternatives in practice

  • BET specific surface area measurement · Micromeritics TriStar II Plus/3Flex (N2 adsorption) — An indicator of fine powder and adsorption surface area. Does not give the particle size distribution itself.
  • Dynamic image analysis · Malvern Morphologi 4/Sympatec QICPIC — Particle size analysis accompanied by shape information. Watch sample amount and statistics.
  • Sieving/microscopy · Retsch AS 200 sieve shaker/Olympus optical microscope + image analysis — A simple method for coarse particles. Low resolution in the fine-powder range.

Step guide

What you actually do

  1. Introduce the sample to the instrument dry (dispersed in air) or wet (dispersed in a liquid it does not dissolve in)
  2. For the wet route, choose a dispersion medium the API does not dissolve in, with a surfactant, and break up the agglomerates with ultrasound or the like
  3. Calculate the particle size distribution from the pattern of the scattered laser light
  4. Read Dv10, Dv50 and Dv90 from the distribution cumulated from the fine end, and compare them with the specification

Equipment and materials

EquipmentLaser diffraction particle size analyzer
Materials and reagentsDispersant (surfactant)

Decision points

Dry, or wet?
The dry route needs little preparation and is fast, while agglomerates can survive and fine particles can be broken. The wet route disperses more easily, but if the API dissolves even slightly in the dispersion medium the values change, so choose a medium it does not dissolve in.
How do you set the dispersion conditions?
Vary the dispersion pressure or the ultrasound power and look for the range in which the agglomerates come apart and the values settle, while the particles themselves are not broken. Once set, use the conditions unchanged.
How do you set the width of the specification?
For poorly soluble APIs the particle size feeds straight into the variability of the dissolution, so confirm how it acts in the dissolution testing of the formulation, and then decide whether to draw lines not only on Dv50 but on the spread of the distribution (Dv10 and Dv90) as well.
Do you look at the shape of the particles and the surface area as well?
Laser diffraction calculates on the assumption that the particles are spheres, so for needle-shaped crystals the information on shape is supplemented by image analysis. Where the surface area of the fines bears on the dissolution, add the BET method (measuring surface area from gas adsorption).

Common mistakes

  • Dispersion is not enough, and agglomerated lumps are measured as large particles
  • Setting the dispersion conditions too hard, and breaking the particles themselves

Check at your own site

  • Was your particle size specification set in connection with dissolution testing?

Step 20: Microbial limits and endotoxin

Check for microbial contamination before formulation. For raw materials of sterile drug products, pyrogens are also controlled according to the route of administration.Check for microbial contamination before formulation. For raw materials of sterile drug products, pyrogens are also controlled according to the route of administration.

Standard methodMicrobial limits test (TAMC/TYMC) + BET (LAL method) when neededMicrobial limits test (TAMC/TYMC) + BET (LAL method) when needed
Details and 3 alternatives

This step coversConfirmation of microbial limits and endotoxin (when needed) (microbial limits and endotoxin)

An API is a chemical, but this step confirms there is no microbial contamination before it goes on to formulation. The required tests are chosen according to the route of administration.

The microbial limits test measures the viable count (total aerobic microbial count, total yeast and mold count) to determine how many microbes are present. This is the basic check.

When the API becomes a raw material for a sterile drug product such as an injectable, endotoxin, a pyrogen, also becomes a concern. It is confirmed by BET (LAL method), putting a double safety net in place.

An API is a chemical, but before it goes on to formulation, it is checked for microbial contamination. Especially when it becomes a raw material for a sterile drug product such as an injectable, not only the microbial count but also pyrogens (endotoxin) are a concern. The required tests are chosen according to the route of administration of the product.

MakersMerck, Charles River, Lonza, bioMérieux, Sartorius

Why it is the standard. Standard based on USP<61>/<62> and <85>, and ICH Q6A. Microbial limits for oral/non-sterile APIs, and endotoxin as well for raw materials of sterile drug products. rFC is increasingly adopted as an alternative to animal-derived reagents.

Check also. Whether microbial limits or endotoxin testing is needed depends on which drug product and route of administration the API finally goes to. The level required differs greatly between oral solid dosage forms and injectables, so tests are chosen by working backward from downstream requirements rather than imposed uniformly at the API stage.

Alternatives in practice

  • Recombinant factor C (rFC) method · Lonza PyroGene rFC Assay/Charles River Endosafe Trillium rCR cartridges — Independent of horseshoe crabs and sustainable. Requires validation against each pharmacopoeia and method.
  • Rapid microbiological methods (RMM) · bioMérieux BacT/ALERT/Rapid Micro Biosystems Growth Direct — Speeds up results. An equivalence study is required at introduction.
  • Sterility test (for sterile APIs) · Sartorius Sterisart/Merck Steritest Symbio (membrane sterility test) — For sterile APIs. Not applicable to limit testing of non-sterile products.

Step guide

What you actually do

  1. Dissolve or suspend the API and either filter it through a membrane filter or mix it into agar medium
  2. Incubate on media for the total aerobic microbial count (TAMC) and for the total yeast and mould count (TYMC) at the set temperature and for the set number of days, and count the colonies
  3. Where the API goes into an injectable, measure endotoxin with LAL reagent (from horseshoe crab) or with recombinant factor C (rFC)
  4. Confirm with samples spiked with microorganisms or endotoxin that the API is not interfering with the test

Decision points

Which tests do you need?
It depends on which formulation and route of administration the API goes on to. For a raw material for oral solid dosage forms, microbial limit testing is the centre of it; for a raw material for injectables, endotoxin, a pyrogen, also matters; and for a sterile API, sterility testing is added, so work back from the requirements of the later steps.
Does the API itself interfere with the growth of the microorganisms or with the LAL reaction?
With antimicrobial or poorly soluble APIs the spiked microorganisms may not be recovered, so confirm that dilution, neutralization or filtration removes the interference. For endotoxin as well, confirm the recovery with spiked samples.
LAL reagent, or rFC?
Where you want to avoid reagents of animal origin, or put weight on security of supply, rFC is a candidate. When you switch, measure the same samples by both and compare them, and confirm the standing of the method in the pharmacopoeias.

Common mistakes

  • Testing an antimicrobial API without removing the interference, so that it looks free of microorganisms
  • Assuming that killing the microorganisms removes the endotoxin as well, and not controlling the two separately

Check at your own site

  • Are the microbiological test items for your API decided from which formulation and route of administration it goes on to?

Guides that go deeper

The reference (“gold standard”) method is the one broadly recognized in published practice and guidelines; your process, product and regulator may justify another. Makers and products are examples, not endorsements. This map is translated from the Japanese edition, where it is updated first (Japanese version).