mRNA-LNP

This reference process map organizes the manufacturing and analysis of mRNA-LNP in the order of synthesis by IVT → capping → purification → LNP encapsulation → characterization. Both the quality of the mRNA and the lipid nanoparticle affect efficacy, and the best method depends on conditions, so each step also lists alternative and complementary methods.

Editor's note. The real decisive points in this process come down to two analyses: capping efficiency and residual dsRNA. The former determines the 'true yield' of IVT and the latter determines translational activity and the innate immune response, so on the shop floor these are nailed down before the purity of the mRNA itself. Where beginners stumble is in jumping to the conclusion that quality is assured if the full-length fraction by CGE is high; a full-length molecule can still be uncapped or have a shortened poly(A) tail and pass through. With LNP, too, it is easy to feel reassured by looking only at the DLS PDI, but encapsulation efficiency and particle size distribution are sensitive to the mixer flow rate and the ionizable lipid ratio, and these are usually the first things to break at transfer. At scale-up, how to overcome the throughput limit of microfluidics becomes a practical issue that stands in the way even before the specifications do.

Made fromPlasmid DNA (linearized template)

At a glance

Select a step to jump to its card. ◇ marks a step whose standard method changes with the condition.

  1. Synthesis and reaction3 steps
    1. Template linearization
    2. IVT, mRNA synthesis (changes with the condition)
    3. Capping (changes with the condition)
  2. Purification5 steps
    1. Template DNA removal (changes with the condition)
    2. Pre-capture conditioning
    3. Capture purification
    4. DsRNA removal and quantitation
    5. UF/DF, buffer exchange
  3. Formulation and fill-finish4 steps
    1. Lipid phase
    2. LNP encapsulation, mixing
    3. Ethanol removal, concentration
    4. Fill-finish
  4. Quality attributes and analytics8 steps
    1. Integrity evaluation
    2. Identity, sequence confirmation
    3. Poly(A) tail analysis
    4. Particle size, zeta potential
    5. Encapsulation efficiency, content
    6. Lipid quantitation, impurities
    7. Potency test
    8. General and safety tests

Synthesis and reaction(3 steps)

Step 1: Template linearization

Uncut template causes read-through, producing dsRNA and overlong chains, so the linearization rate and residual supercoiled fraction are confirmed before release.Uncut template causes read-through, producing dsRNA and overlong chains, so the linearization rate and residual supercoiled fraction are confirmed before release.

Standard methodConfirmation of linearization rate and residual supercoiled fraction by agarose/capillary gel electrophoresisConfirmation of linearization rate and residual supercoiled fraction by agarose/capillary gel electrophoresis
Details and 3 alternatives

This step coversLinearization of the DNA template and quality confirmation (plasmid linearization) (template linearization)

Preparation work that readies the 'template' for the transcription reaction known as IVT. The circular plasmid, the blueprint, is cut at just 1 site around the ring with a restriction enzyme and opened into a single linear DNA.

If any is left uncut, transcription goes once around the template (read-through), generating overlong products and double-stranded RNA. Gel electrophoresis confirms the linearization rate and the amount of circular plasmid that remains.

In other words, this step judges 'was it cut properly at only 1 site?' before release. If this is sloppy, impurities in later steps increase sharply, so it serves as the first quality gate.

The template for IVT is a circular plasmid cut at just 1 site with a restriction enzyme to make it linear. If linearization is incomplete, transcription goes around the plasmid (read-through), causing overlong transcripts and double-stranded RNA. So before release, confirm 'was it cut properly?' and 'how much supercoiled plasmid remains?'

MakersCytiva, Bio-Rad, Agilent, NEB, Implen

Why it is the standard. The linearization rate is an important process parameter directly tied to accurate termination of transcription and dsRNA reduction. The residual supercoiled fraction in the linear template has been reported as a key IPC for mRNA drug substance manufacturing (Mol Ther Nucleic Acids 2024, PMC11214521).

Check also. The linearization rate is normally checked by electrophoresis, but a small amount of remaining uncut plasmid can seed overlong transcripts and dsRNA. Restriction enzyme star activity (cutting at sites other than the intended one) conversely mixes in short fragments, so it is safer to review enzyme amount and reaction time for each template.

Alternatives in practice

  • Assessment of linearization rate/template integrity by qPCR · Applied Biosystems QuantStudio 5/Bio-Rad CFX Opus 96 + PowerUp SYBR Green Master Mix — Quantifies undigested (residual supercoiled) template with primers spanning the cut junction. Trace-level, high sensitivity.
  • Purification of linearized plasmid by preparative chromatography/precipitation · Cytiva ÄKTA pure 150 + HiScreen/HiTrap Q anion exchange, or ethanol/PEG precipitation — Removes enzyme, buffer and cut-off pieces, doubling as pre-IVT purification.
  • Concentration and purity confirmation by UV absorbance · Thermo NanoDrop One/Implen NanoPhotometer N60 — Screens for residual protein and organic matter by A260/A280 and A260/A230. Simple, but cannot determine the linearization rate.

Step guide

What you actually do

  1. Measure the concentration of the purified plasmid DNA by absorbance and set the amount to put into the reaction
  2. Add a restriction enzyme that cuts at only one site, outside the region to be transcribed, together with its dedicated buffer, and run the reaction at the set temperature and time
  3. Run the reaction mixture on electrophoresis and read the proportion that has become linear and the proportion of uncut circular form (supercoiled)
  4. Where needed, insert a purification (precipitation or a column) that removes enzyme and buffer components, and hand the material on as the IVT template with the concentration adjusted

Decision points

What decides the restriction enzyme?
The premise is an enzyme with no recognition sequence inside the region to be transcribed, which cuts at only one site. A blunt cut or a 5' overhang suits; a 3' overhang is a cause of long transcripts that run round onto the opposite strand. Where you want to cut immediately after the poly(A) tail and leave no extra sequence at the 3' end, a Type IIS enzyme, which cuts outside its recognition sequence, is a candidate. Supply at GMP grade and the presence or absence of animal-derived components are also compared, as for any raw material.
Do you look at the linearization rate on an agarose gel or by capillary electrophoresis?
An agarose gel is cheap and easy, but as the band intensity is read by eye it is limited in quantitation. Capillary electrophoresis returns numbers and traces and can handle a number of samples, so it suits in-process control and records. Where you want to examine a trace of uncut material, backing it up with qPCR using primers that span the cut site is also an option.
Do you purify the template after the reaction?
Enzyme and buffer components carried over can interfere with the transcription reaction or become impurities in later steps, so a configuration that inserts a purification is often seen. Purification adds recovery losses and a further step, so confirm at small scale whether the amount carried over is within what IVT can tolerate, and decide from that.
What do you do when a lot of material is left uncut?
The options include further digestion, re-purification and placing the lot on hold, but which one you take, and how many times retesting is allowed, are written into the approved procedure in advance. Decisions on individual lots are passed to that procedure and the quality unit.

Common mistakes

  • Not reviewing the enzyme amount and reaction time for each template, so that uncut material, or short fragments from star activity, come along
  • Recording only the score or the numbers from the electrophoresis, and keeping no trace or gel image
  • Choosing an enzyme that leaves a 3' overhang without a blunting step, and inviting long transcripts

Check at your own site

  • By which method is the linearization rate of your template measured, and how is the handling of uncut material written in your procedure?
  • Is there a record confirming by sequence that your restriction enzyme has no recognition sequence inside the transcribed region?

Step 2: IVT, mRNA synthesis◇ Branches

Mg2+ and the NTP ratio change yield, integrity and the amount of dsRNA, so reaction progress is followed by AEX-HPLC to decide when to stop.The alternative runs IVT with standard NTPs and no cap analog, so nucleotides do not compete with cap addition and transcription yield rises (the default side lowers the GTP ratio and adds CleanCap AG/AU to the reaction)

Standard methodIVT with T7 RNA polymerase + reaction monitoring by anion-exchange (AEX)-HPLCT7 IVT with standard NTP composition without cap analog (no co-transcriptional cap). Reaction monitoring by AEX-HPLC is common to both
ReagentsIVT enzymes reagents (T7, NTP, DNase)T7 RNA polymerase + ATP/GTP/CTP/UTP at the usual ratio (trinucleotide analogs such as CleanCap AG/AU are not used)
With Post-transcriptional enzymatic capping (Vaccinia capping enzyme + 2'-O-MTase)T7 IVT with standard NTP composition without cap analog (no co-transcriptional cap). Reaction monitoring by AEX-HPLC is common to bothWith Co-transcriptional capping (trinucleotide analogs such as CleanCap)IVT with T7 RNA polymerase + reaction monitoring by anion-exchange (AEX)-HPLC
Details and 3 alternatives

This step coversmRNA synthesis by in vitro transcription (IVT) and reaction monitoring (IVT, mRNA synthesis)

This is the heart of the process, where the mRNA itself is made. An enzyme called T7 (or SP6/T3) links the 4 raw materials (NTPs) into a single strand using linear DNA as the template, synthesizing mRNA.

Yield, integrity and the amount of double-stranded RNA by-product all change with Mg2+ concentration, NTP ratio and reaction time. Progress is monitored in nearly real time by AEX-HPLC.

The point is judging 'when to stop.' Quality varies whether the reaction is pushed too far or stopped too early, so the call to end the reaction at the optimal moment is made while watching the data.

T7 (or the related SP6/T3) RNA polymerase synthesizes mRNA from 4 NTPs using linear DNA as the template. Yield, integrity and the amount of dsRNA change with Mg2+ concentration, NTP ratio and reaction time, so the progress of the reaction is watched in close to real time to decide when to stop.

MakersSartorius, Agilent, Eppendorf, NEB, Aldevron

Why it is the standard. AEX-HPLC can individually quantify each NTP, cap analog, mRNA and residual DNA in under 6 minutes, and has been reported to be useful for understanding and controlling the IVT process (Front Mol Biosci 2024, 3389/fmolb.2024.1250833; PMC13013469). The T7 system is the de facto standard polymerase for commercial mRNA.

Check also. Reaction monitoring is effective for deciding when to stop, but it cannot tell the amount of dsRNA formed as a by-product. dsRNA is always confirmed in a later step (#6) by a different principle (such as an anti-dsRNA antibody).

With Post-transcriptional enzymatic capping (Vaccinia capping enzyme + 2'-O-MTase). Uncapped 5'-triphosphate RNA is produced, which presupposes enzymatic capping in a later step. Competition affecting full-length synthesis is reduced compared with the co-transcriptional method, but downstream steps increase

Alternatives in practice

  • Lateral flow/rapid IPC · Commercial simple RNA detection strips + handheld reader — An ultra-rapid in-process check at the level of presence/absence. Low quantitation.
  • Time-course UV (A260) measurement · Thermo NanoDrop One/Agilent Cary 60, or in-line UV-Vis probe — Tracks the increase in mRNA concentration over time. Simple, but purity and integrity require separate testing.
  • co-transcriptional capping (cap added simultaneously with IVT) · The T7 IVT system above + TriLink CleanCap Reagent AG [N-7113] — Omits the enzymatic capping step. AEX-HPLC monitoring is the same.

Step guide

What you actually do

  1. Add the linearized template, T7 RNA polymerase, the 4 NTPs (plus the modified UTP where one is used) and transcription buffer containing magnesium, and run the reaction at a held temperature
  2. For co-transcriptional capping, put the cap analogue into the same reaction mixture
  3. Draw samples during the reaction and follow the progress of synthesis by anion exchange HPLC
  4. Decide when to stop, end the reaction, and hand the reaction mixture on to the DNase treatment step

What changes with the conditionWhether the cap is added during the reaction (co-transcriptional) or afterwards in a separate enzymatic reaction (in co-transcriptional capping the cap analogue is added to the reaction mixture and it all proceeds in one reaction. In the post-transcriptional method, IVT is run with the ordinary NTP composition and no cap analogue, and a separate capping step follows. Choose by the number of steps, the yield and the start sequences you can use)

Decision points

What decides when the reaction is stopped?
The Mg2+ concentration, the NTP ratio, the temperature and the time change the yield, the proportion of full-length material and the amount of dsRNA. A common approach is to measure the reaction mixture by HPLC over time and look for the point just before full-length material starts to fall even as yield rises. The stopping point itself is set during condition screening and fixed in the production procedure.
Do you use a modified UTP (m1Ψ and the like)?
Designs that replace uridine entirely are widely used, to hold down immunogenicity and raise translation. Reaction conditions and how the cap is added change, so confirm yield and quality side by side with the unmodified case. Availability of the modified NTP at GMP grade, and the number of suppliers, also come into the decision.
What grade of enzyme and NTPs do you use?
Research grade and GMP grade differ in specifications and documentation. If you have the clinical or commercial stage in view, choose a supplier early that offers GMP grade or equivalent quality documentation. Compare on the specifications for endotoxin and residual nuclease, lot-to-lot reproducibility, and whether a second source exists.
What do you confirm when you raise the scale of the reaction?
When the uniformity of mixing and temperature changes, the yield and the proportion of full-length material can move. Confirm the conditions set at small scale in the actual reaction vessel, and also fix where and when samples are drawn.

Common mistakes

  • Extending the reaction looking at yield alone, lowering the proportion of full-length material and increasing the amount of dsRNA
  • Setting the conditions with research-grade enzyme and NTPs, so that the yield and the impurity profile change when they are switched to GMP grade
  • Not noticing unevenness in temperature or mixing when the scale is raised, and using the small-scale conditions as they are

Check at your own site

  • Is the stopping point of your IVT set by a fixed time, or by actual measurement of the progress?
  • How far have you confirmed the effect of changing the grade or the supplier of your enzyme and NTPs?

Step 3: Capping◇ Branches

Uncapped mRNA is not translated and becomes a source of inflammation, so this step quantifies the rate of correct Cap1 addition by LC-MS.In the default, the cap is formed during IVT and there is no separate step, whereas in the alternative, cap-0 is added (VCE) as a separate enzymatic reaction after IVT and then converted to cap-1 by 2'-O-methylation

Standard method5' cap structure identification and capping efficiency quantitation by enzymatic digestion + reverse-phase/ion-pair reverse-phase LC-MSPost-transcriptional enzymatic capping reaction (37 °C, about 1 hour): cap-0 with Vaccinia capping enzyme, followed by cap-1 with mRNA cap 2'-O-methyltransferase. Capping efficiency up to ~99.9%
EquipmentLC-MS (liquid chromatography mass spectrometry)Independent reaction vessel with a temperature-controlled incubator (a separate batch reaction in the two-pot method). Enzyme removal is integrated into the later oligo-dT/AEX or LiCl precipitation
ReagentsmRNA capping reagentsIon-pair reagents for IP-RP LC-MSVaccinia capping enzyme (GuaT), mRNA 2'-O-methyltransferase, GTP, S-adenosylmethionine (SAM). LC-MS/IP-RP-LC for efficiency quantitation is common to both
With Post-transcriptional enzymatic capping (Vaccinia capping enzyme + 2'-O-MTase)Post-transcriptional enzymatic capping reaction (37 °C, about 1 hour): cap-0 with Vaccinia capping enzyme, followed by cap-1 with mRNA cap 2'-O-methyltransferase. Capping efficiency up to ~99.9%With Co-transcriptional capping (trinucleotide analogs such as CleanCap)5' cap structure identification and capping efficiency quantitation by enzymatic digestion + reverse-phase/ion-pair reverse-phase LC-MS
Details and 3 alternatives

This step coversAddition of the 5' cap and confirmation of capping efficiency (capping)

A step that puts a 'cap', like a hat, on the front end (5' end) of the synthesized mRNA. With this hat, the mRNA is easily translated in the cell and is less likely to be regarded as foreign by the immune system.

A bare mRNA without the hat works poorly and also triggers inflammation. So it is not enough that the cap is attached; 'what percent was correctly attached' must be pinned down as a number.

LC-MS identifies the cap structure (mainly Cap1) and quantifies capping efficiency. It is an unglamorous but indispensable finishing move that affects efficacy and safety.

When a cap (the Cap1 structure is mainstream) is attached to the 5' end of mRNA, it is easily translated in the cell and is less likely to be treated as foreign by innate immunity. mRNA without a cap works poorly and causes inflammation, so 'what percent was correctly capped' is measured.

MakersAgilent, Waters, NEB, Honeywell, TriLink

Why it is the standard. LC-MS is established as the platform method for evaluating 5' cap efficiency and can analyze cap structure identification, sequence mapping and even poly(A) chain length (LCGC; Pharm Technol). >95% capping has been reported with the CleanCap co-transcriptional method.

Check also. Capping efficiency is normally quantified by LC-MS, but the achievable upper limit and the assumptions behind measurement differ between the co-transcriptional and enzymatic methods. Specifications and measurement conditions, including the cap structure (Cap0 or Cap1), need to be aligned for each product.

With Post-transcriptional enzymatic capping (Vaccinia capping enzyme + 2'-O-MTase). Reaction buffer, enzymes and SAM/GTP are added, and the burden of removing enzyme protein increases

Alternatives in practice

  • Capping efficiency assay by ribozyme/DNAzyme · The IP-RPLC-UV system above (Agilent 1290 Bio LC) — Calculates capping efficiency from the relative ratio of 5' fragments by UV detection without using MS.
  • Vaccinia-based enzymatic capping + 2'-O-methylation (conversion to Cap1) · NEB Vaccinia Capping System [M2080] + 2'-O-MTase [M0366], temperature-controlled reaction system — Enzymatic Cap1 addition when co-transcriptional capping is not used.
  • 5' fragment profiling by IP-RP-UHPLC/UV · Waters ACQUITY Premier + Oligonucleotide BEH C18 column, UV detection — For labs without MS. Separates and quantifies capped/uncapped fragments by retention time.

Step guide

What you actually do

  1. For co-transcriptional capping, the cap analogue put into the IVT reaction mixture adds the cap at the same time as transcription (with a trimer analogue, Cap 1 is added in one stage)
  2. For post-transcriptional capping, react capping enzyme and 2'-O-methyltransferase together with GTP and SAM (the methyl donor) to give Cap 0 and then Cap 1
  3. For the measurement, add a biotinylated DNA probe matching a position near the 5' end together with RNase H, and cut out a short fragment containing the cap
  4. Recover the fragment on streptavidin magnetic beads to remove the matrix, and derive the proportions of uncapped, Cap 0, Cap 1 and reverse cap by LC-MS

What changes with the conditionWhether the cap is added co-transcriptionally or enzymatically after transcription (in co-transcriptional capping there is no separate capping step; it is done within IVT. In the post-transcriptional method, IVT is run without the cap and a separate enzymatic reaction (an example is around 1 hour at 37 °C) gives Cap 0 and then Cap 1, so adding the enzymes and SAM and removing the enzymes come into the process. Choose by the number of steps, the yield and the start sequence)

Decision points

Co-transcriptional, or the post-transcriptional enzymatic method?
Co-transcriptional capping has fewer steps, and with a trimer analogue Cap 1 is added in one stage. In exchange, constraints come in on the transcription start sequence, and the suppliers of the analogue are limited. The enzymatic method has fewer sequence constraints and is also used for long chains, but adds the enzymatic reaction and a re-purification. Choose by which failure weighs more in your own process.
How do you read a supplier's "capping efficiency"?
Whether that figure is the proportion of molecules that reached Cap 1, whether it includes Cap 0, and whether reverse caps are excluded differ from product to product. The definition of 1 unit of enzyme is also sometimes set on a short standard transcript. Do not use a supplier's figures for comparison until you have measured again with the length and the sequence of your own mRNA.
What do you count as "capped"?
Counting Cap 1 only, including Cap 0, or putting reverse caps into the denominator gives different figures from the same spectrum. Set the definition first and write it into the test method.
How do you design the sample preparation (the probe) for the measurement?
Predict the secondary structure near the 5' end and set the position of the cut from it. If the fragment is too short the duplex becomes unstable, and if it is too long the mass difference of the cap is buried, so one approach is to start at around 20 bases and close in while watching the separation in the mass spectrum. The probe is redesigned when the product changes, and the cleavage conditions are fixed as a procedure.

Common mistakes

  • Reading the capping efficiency from the catalogue as the proportion of Cap 1
  • A reverse cap has the same mass as the correct orientation, so it cannot be told apart by mass alone
  • Not redesigning the probe when the product changes, so the value moves with the shift in the position of the cut

Check at your own site

  • Is your capping efficiency defined as the proportion of Cap 1?
  • When you change the capping approach, can your test method separate the reason the value moved into a difference in efficiency and a difference in measurement?

Purification(5 steps)

Step 4: Template DNA removal◇ Branches

A step that degrades the template to stop the reaction and keeps residual DNA, a safety-related impurity, within specification by qPCR.In the two-pot method of the alternative, an intermediate capture/purification is inserted before the capping reaction, so the flow after DNase treatment and reaction stop is reorganized to straddle the capping step

Standard methodTemplate degradation by DNase I digestion + residual DNA quantitation by qPCRDNase I digestion + qPCR quantitation of residual DNA are common to both, but in the two-pot method the steps are inserted in the order DNase → intermediate purification → enzymatic capping → final purification (in the improved one-pot method, enzymes can also be added in the same vessel after dilution)
With Post-transcriptional enzymatic capping (Vaccinia capping enzyme + 2'-O-MTase)DNase I digestion + qPCR quantitation of residual DNA are common to both, but in the two-pot method the steps are inserted in the order DNase → intermediate purification → enzymatic capping → final purification (in the improved one-pot method, enzymes can also be added in the same vessel after dilution)With Co-transcriptional capping (trinucleotide analogs such as CleanCap)Template degradation by DNase I digestion + residual DNA quantitation by qPCR
Details and 3 alternatives

This step coversTemplate DNA removal by DNase treatment and reaction stop (template DNA removal)

The reaction liquid after transcription still contains the DNA used as the template. This step breaks it down finely with an enzyme called DNase I and removes it.

This treatment serves 2 purposes: (1) to stop the transcription reaction properly, and (2) to reduce residual DNA, which would be a safety risk if left in the final product.

After degradation, qPCR measures the amount of residual DNA to confirm it is within specification. It is the step at the entrance of purification that 'erases the template and closes the reaction.'

When IVT is finished, the template DNA is degraded and removed with DNase. This stops the reaction and reduces residual DNA that would be mixed into the final product. Residual DNA is a safety-related impurity, so it is quantified later and kept within specification.

MakersCytiva, Bio-Rad, Eppendorf, Molecular Devices, NEB

Why it is the standard. A standard step in which DNase I digests the template and stops the reaction, yielding full-length mRNA drug substance with low residual dsDNA (advancingrna.com end-to-end). Residual DNA is quantified by qPCR targeting the origin of replication sequence, with a standard curve prepared from linear plasmid (PMC8641981).

Check also. DNase treatment is a step that lowers residual DNA and is not itself a quantitation method. The final amount of residual DNA is measured separately by qPCR or similar and judged against the specification. Note also that after LNP encapsulation, encapsulated nucleic acid that DNase cannot reach may remain.

With Post-transcriptional enzymatic capping (Vaccinia capping enzyme + 2'-O-MTase). Step order and batch management change. In the default co-transcriptional method there is no separate capping step, and the process from this step onward is linear

Alternatives in practice

  • Residual DNA quantitation with a fluorescent dye (PicoGreen) · Fluorescence plate reader (BMG CLARIOstar/Molecular Devices SpectraMax) + Thermo Quant-iT PicoGreen dsDNA [P11496] — Simple quantitation of total dsDNA. No sequence specificity, and pretreatment is needed in the presence of mRNA.
  • Immunoassay-based residual DNA measurement · ELISA/plate reader method succeeding the Thermo Threshold system — Specific detection of host/plasmid DNA. For orthogonal confirmation of qPCR.
  • Physical DNA removal by downstream chromatography/TFF · Cytiva ÄKTA process + oligo-dT/AEX, or hollow fiber TFF (ÄKTA flux) — Removes fragmented DNA by differences in size/affinity without relying on enzymatic degradation.

Step guide

What you actually do

  1. To the finished IVT reaction mixture, add DNase I and buffer containing the magnesium and calcium needed
  2. Run the reaction at the set temperature and time to break the template DNA into short fragments
  3. Stop the enzyme with EDTA or similar, or remove the enzyme in the purification that follows
  4. Draw a sample and measure the remaining DNA by qPCR or digital PCR, with primers and probes targeting the origin of replication or similar

What changes with the conditionWhether the flow from here runs straight on with co-transcriptional capping, or a step is inserted for post-transcriptional enzymatic capping (the two-pot method) (in the two-pot method the order after DNase treatment is intermediate purification, enzymatic reaction, final purification. In improved one-pot methods there are also configurations where the enzymes are added in the same vessel after dilution)

Equipment and materials

EquipmentTemperature-controlled reaction vessel (including where the same vessel as IVT is used straight on)Real-time PCR and digital PCR instruments

Decision points

Is DNase treatment alone enough?
DNase treatment is an operation that cuts the DNA short; the fragments stay in the liquid. Configurations that drop the fragments in a later oligo-dT capture or TFF step are often seen, so write down in advance which step is given the role of removing which impurity.
How do you set the amount of enzyme and the time?
More enzyme is faster but costs more and leaves more residual enzyme. Enzyme concentration, time, temperature and the presence of magnesium and calcium all matter, so confirm the rate of digestion and the length of the fragments with the actual reaction mixture and decide from that. If the upstream buffer contains EDTA the enzyme will not work, so check the composition as well.
How do you control the enzyme you added?
DNase itself can remain as a process-related impurity. Put confirmation that it is removed downstream, and control of the amount remaining, into the plan. Grade, origin, the presence or absence of animal-derived components, and lot-to-lot differences in specific activity are subject to supplier control.
Do you measure residual DNA by qPCR or by digital PCR?
qPCR quantifies against a standard curve, while digital PCR gives an absolute amount with no calibration curve and is said to be less affected by inhibitors. Decide in advance which sequence you target and whether the primer design can pick up fragments derived from the template.

Common mistakes

  • Taking the mere fact that DNase treatment was done as having shown the amount of residual DNA
  • Adding the enzyme to a liquid carrying over a chelator such as EDTA, so that digestion does not proceed
  • Not measuring the residue of the enzyme added, and leaving it off the impurity control table

Check at your own site

  • At which step are the fragments cut by your DNase treatment planned to be removed?
  • Does your residual DNA measurement use a target sequence that can pick up fragments derived from the template?

Step 5: Pre-capture conditioning

Because IVT reaction liquid containing high concentrations of salt, NTPs and enzymes cannot be loaded directly onto affinity captureBecause IVT reaction liquid containing high concentrations of salt, NTPs and enzymes cannot be loaded directly onto affinity capture

Standard methodConcentration and diafiltration by hollow fiber TFF (100 kDa MWCO, for ssRNA) for pre-capture buffer exchange and desaltingConcentration and diafiltration by hollow fiber TFF (100 kDa MWCO, for ssRNA) for pre-capture buffer exchange and desalting
Details and 2 alternatives

This step coversConditioning before capture (TFF/buffer exchange or precipitation) → the reaction liquid after DNase treatment is concentrated and buffer-exchanged by ultrafiltration/diafiltration or LiCl/ammonium acetate precipitation, preliminarily removing NTPs, short fragments, salts and enzymes and adjusting the loading conditions for oligo-dT/IEX capture (pre-capture conditioning/buffer exchange)

Prep work before the main capture purification. The reaction liquid after DNase treatment is rich in NTPs, short fragments, salts and enzymes, and as it is, binding to the resin does not go well.

So the liquid is concentrated and its buffer replaced by TFF (ultrafiltration/diafiltration) or salt precipitation. The idea is to lower conductivity and drop interfering small molecules in advance.

The aim is to stabilize the 'binding efficiency and reproducibility' of the next capture step. It is a preparatory step for standardization, making the conditions of the loaded liquid the same every time.

The IVT reaction liquid after DNase treatment has high conductivity and contains large amounts of free NTPs and small molecules, which impair the binding efficiency and reproducibility of capture by oligo-dT or ion exchange. Standardizing the loading conditions by buffer exchange with TFF (or concentration and desalting by precipitation) before the capture step is a standard process step.

MakersCytiva, Merck, Sartorius, Repligen, Beckman Coulter

Why it is the standard. Hollow fiber TFF concentrates and desalts at low shear while preserving the integrity of large ssRNA, and allows the conductivity of the capture load to be adjusted reproducibly; this is the actual equipment and MWCO configuration.

Alternatives in practice

  • Concentration and desalting by LiCl/ammonium acetate precipitation · Refrigerated centrifuge (Beckman Coulter Avanti J-15R) + LiCl/ammonium acetate reagents — For small scale and lab use. Large-scale salt removal requires redissolution and washing.
  • Concentration and diafiltration by flat-sheet cassette TFF · Sartorius Sartoflow Smart/Cytiva ÄKTA readyflux + Sartocon Hydrosart 100 kDa/Merck Pellicon 3 Ultracel 100 kDa — High flux with flat sheet instead of hollow fiber. Watch for shear sensitivity.

Step guide

What you actually do

  1. Put the reaction mixture into the TFF tank and circulate it along a membrane of around 100 kDa to reduce the liquid volume
  2. Keep circulating while adding loading buffer for the capture step to replace what has passed through, washing out NTPs, short fragments, salts and enzymes
  3. Confirm that the conductivity and pH of the retentate have reached the conditions for the next capture step
  4. Push the liquid left in the membrane and the piping out with buffer and recover it

Decision points

Exchange by TFF, or concentrate by salt precipitation?
Precipitation with LiCl or ammonium acetate is easy at small scale, but removing a large amount of salt needs redissolution and washing. As the scale rises, TFF tends to be chosen, taking the centrifugation equipment involved into account as well.
Hollow fiber, or a flat sheet cassette?
Hollow fiber has wide flow channels and gentle shear, while a flat sheet cassette gives a higher flux (flow per membrane area). For a large molecule such as mRNA, which shear can break, choose by confirming in a test with the real liquid, from the before-and-after difference in the proportion of full-length material, whether the shear in a flat sheet is acceptable.
What decides the molecular weight cut-off (MWCO) of the membrane?
mRNA is a large molecule of the several-hundred kDa class, so a membrane of around 100 kDa (for example) retains the mRNA and lets NTPs and short fragments through. However, membranes in the tens of kDa class retain dsRNA along with it, so this step is not treated as a step that removes dsRNA. Adsorption of mRNA to the membrane material and the recovery are compared as well.
How do you set the conductivity after the exchange?
Oligo-dT wants high salt and ion exchange wants low conductivity: the next capture method sets opposite conditions. Start from the supplier's recommended conditions and set the range in which binding is stable at the chain length of your own mRNA by a small-scale study of the capture step.

Common mistakes

  • Membranes in the tens of kDa class retain dsRNA as well, yet it is taken as having been removed
  • In a small batch, much liquid is left in the membrane and the piping, so the recovery cannot be read
  • Redissolution after salt precipitation is incomplete and the yield shrinks

Check at your own site

  • Is it written in your design documents that your TFF is not a step that removes dsRNA?
  • Is the end point of your buffer exchange set by actual measurement of conductivity and pH?

Step 6: Capture purification

The core of purification, capturing only full-length mRNA on dT resin with the poly(A) tail as a fishhook and washing away incomplete chains and enzymes.The core of purification, capturing only full-length mRNA on dT resin with the poly(A) tail as a fishhook and washing away incomplete chains and enzymes.

Standard methodOligo(dT) ligand affinity chromatography (poly(A) capture)Oligo(dT) ligand affinity chromatography (poly(A) capture)
Details and 3 alternatives

This step coversCapture purification of mRNA by oligo-dT affinity chromatography (capture purification)

The core of mRNA purification. The poly(A) tail at the end (3' end) of the mRNA is likened to a 'fishhook,' and a resin that grabs it picks out only full-length mRNA.

Resin lined with dT (thymidine) sticks only to poly(A), so NTPs, incomplete chains cut partway and enzymes can be washed away together. Only the target molecule remains.

Because it uses poly(A), the 'proof of the real thing,' the selection is smart and highly reproducible. It is the separation step that plays the lead role in downstream processing in many mRNA manufacturing processes.

The poly(A) tail at the 3' end of mRNA is used as a 'fishhook.' A resin lined with dT grabs only poly(A), so NTPs, short incomplete chains and enzymes are washed away together and only full-length mRNA is recovered. It is the core step of mRNA purification.

MakersCytiva, Sartorius, Agilent

Why it is the standard. Oligo-dT affinity is a robust drug substance purification method that selectively captures full-length mRNA via the poly(A) tail, and QbD by continuous operation (continuous oligo-dT) has also been reported (Mol Ther Nucleic Acids 2024, PMC11458983). It is also used for approved mRNA vaccines.

Alternatives in practice

  • Preparative purification by ion-pair reverse-phase HPLC (IP-RP) · Agilent 1290 Infinity II Bio LC/preparative LC + DNAPac RP, TEA/HFIP mobile phase — Separates full-length product from impurities independent of poly(A). Scale-up and solvent load are the challenges.
  • Core/multimodal chromatography · Cytiva ÄKTA process+Capto Core 700/Capto adhere, or BIA Separations (Sartorius) CIMmultus QA/oligo-dT monolith — Monoliths are advantageous for mass transfer of large mRNA molecules. They also allow simultaneous clearance of dsRNA/fragments.
  • mRNA enrichment and purification by cellulose/ethanol treatment · Cellulose column (Sigma cellulose powder) + ethanol buffer — Simple and low-cost, but selectivity is inferior to affinity resin. Reduction of dsRNA is also expected.

Step guide

What you actually do

  1. Equilibrate the support with immobilized oligo-dT (resin column, membrane or monolith) with a high-salt buffer
  2. Pass the buffer-exchanged mRNA solution through at the set flow rate (residence time) so that mRNA with a poly(A) tail binds
  3. Wash with buffer to flush away enzymes, NTPs, template fragments and chains that stopped part way without a tail
  4. Elute with water or a low-salt buffer (warmed in some practice), watching the UV absorbance peak, and collect
  5. Wash and regenerate the support, ready for the next cycle

Decision points

A resin, or a membrane or monolith, as the support?
Resin has high binding capacity and more of an operating track record, but a large molecule such as mRNA diffuses slowly inside the pores and needs sufficient residence time (a low linear flow rate). In membranes and monoliths the material is carried by convection, so they can be run at high flow rates in a short time and are easier to make single-use, while the absolute capacity figure changes with the conditions.
Can you use the binding capacity from the catalogue in your design as it stands?
Dynamic binding capacity changes with the chain length of the mRNA and the residence time, and comparisons have been reported in which the ranking of supports swaps around with the conditions. Ask the supplier "at how many nt of mRNA, and at what residence time, was the value measured", and set the amount of support and the number of cycles from a value measured again at your own chain length.
Elute with low salt alone, or combine it with warming?
Base pairs come apart as the temperature rises, so warming raises recovery even with a low-salt buffer. On the other hand, RNA degrades faster at high temperature, so set the temperature and the time and measure the amount of degradation under those conditions first.
Regenerate the support and use it repeatedly, or make it single-use?
The ligand is DNA, so washing with the strong alkali used in protein purification can lower the ligand density cycle by cycle. If you use it repeatedly, confirm in advance the capacity after the number of wash cycles you assume, and that there is no carry-over. A single-use membrane can lighten the burden of cleaning and packing verification.

Common mistakes

  • Using a catalogue binding capacity measured at around 2,000 nt as it stands for a construct of more than 4,000 nt
  • Washing with strong alkali, so the ligand density falls cycle by cycle
  • Taking oligo-dT as having removed dsRNA as well, and not putting a separate removal step into the design

Check at your own site

  • Is your capture step designed on a binding capacity measured again at your own chain length and residence time?
  • Does your process write down which step is responsible for removing which impurity?

Step 7: DsRNA removal and quantitation

dsRNA triggers strong innate immunity and lowers efficacy, yet is hard to remove, so dedicated removal and quantitation with J2 antibody are needed.dsRNA triggers strong innate immunity and lowers efficacy, yet is hard to remove, so dedicated removal and quantitation with J2 antibody are needed.

Standard methoddsRNA removal by cellulose/IP-RP-HPLC + J2 anti-dsRNA antibody immunoassaydsRNA removal by cellulose/IP-RP-HPLC + J2 anti-dsRNA antibody immunoassay
Details and 3 alternatives

This step coversRemoval and quantitation of double-stranded RNA (dsRNA) impurities (dsRNA removal and quantitation)

A step that targets and removes the double-stranded RNA (dsRNA) born as a by-product of IVT. dsRNA triggers a strong innate immune response, lowering efficacy and causing adverse reactions.

The troublesome part is that its size is close to that of the mRNA itself. It passes through together with the mRNA in TFF or ordinary purification, so a dedicated removal step using cellulose or the like is needed.

After removal, it is quantified accurately with the anti-dsRNA antibody called J2. 'Managing a hard-to-remove impurity with a dedicated hand and a dedicated yardstick' is a step directly tied to safety.

dsRNA formed as a by-product of IVT triggers a strong innate immune response, lowering efficacy and causing adverse reactions. Its size is close to that of the mRNA itself, so it is hard to remove by TFF or simple purification, and a dedicated removal step and precise quantitation are required.

MakersNordic-MUbio, Bio-Rad, Agilent, Molecular Devices, Sartorius, Waters

Why it is the standard. The J2 antibody is regarded as the reference method for dsRNA detection, and residual dsRNA is quantified by sandwich ELISA or immunoblot (PMC11359411). In the 2024 revision of the guideline for mRNA vaccine analytical methods, immunoblot was added for dsRNA detection. dsRNA is hard to remove by SEC/TFF/oligo-dT, and cellulose or IP-RP-HPLC is the standard removal method at manufacturing scale (Mol Ther Nucleic Acids 2025).

Check also. dsRNA is close in size to the mRNA itself and is not fully removed by simple TFF or general purification. Even after removal, the remaining amount is confirmed with a highly sensitive quantitation (such as an anti-dsRNA antibody immunoassay) to cross-check how well the removal worked.

Alternatives in practice

  • dsRNA quantitation by RP-UHPLC/UV · Thermo Vanquish/Waters ACQUITY Premier + Oligonucleotide BEH C18, heated separation — Quantifies the dsRNA peak from retention time differences without antibody. High throughput.
  • dsRNA detection by biolayer interferometry (BLI) · Sartorius Octet R8/BLI + J2 antibody-immobilized biosensor — Label-free and real-time. For in-process monitoring.
  • Qualitative confirmation by dot blot/immunoblot · Bio-Rad ChemiDoc MP+J2 antibody (Nordic-MUbio) — Simple pass/fail screening. Semi-quantitative.

Step guide

What you actually do

  1. Adjust the captured mRNA into a buffer containing ethanol (around 16% in reported examples), pass it over a cellulose support and leave only the double-stranded material on the support. Fractionation by ion-pair reverse-phase HPLC is also a method
  2. Collect the fraction that passes through (single-stranded mRNA), remove the ethanol and so on, and hand it on to the next step
  3. Measure samples from before and after the treatment by ELISA or dot blot using the J2 antibody, and compare the amount of dsRNA
  4. Where needed, use a system that puts the material into cells and looks at the interferon response, to confirm even the stimulation the antibody cannot pick up

Decision points

Cellulose, or ion-pair reverse-phase?
Cellulose is cheap and easy to obtain and is widely used, but reports put the recovery of mRNA at a little over 65%, which means letting the rest go. Ion-pair reverse-phase wins on resolution, but procuring and disposing of organic solvents and removing the ion-pairing reagent are burdens. Choose by which burden weighs more at your own scale.
How far do you reduce dsRNA on the IVT side?
The amount formed moves with the temperature, the Mg2+ concentration, the NTP ratio and how the template is cut. What you reduce here is less material to discard in the removal step, and recovery rises as well. It does not reach zero, so plan the split between suppression and removal on the premise that the removal step stays.
On which measurement do you set the specification?
The J2 antibody recognizes double-stranded material of roughly 40 base pairs or more regardless of sequence, but does not pick up shorter duplexes. The reaction of the antibody can also change with modified nucleotides, so confirm dilution linearity and spike recovery with your own mRNA. Combining it with a biological response (interferon) is also an option.

Common mistakes

  • Taking dsRNA as having been removed by oligo-dT, TFF or precipitation, and not putting a removal step into the design
  • Reading a fall in the J2 figure as a reduction that includes short duplexes
  • Reading the use of modified nucleosides as meaning immunogenicity is held down, and not confirming the response after purification

Check at your own site

  • Is it written in the definition of your specification how many base pairs of double-stranded material your dsRNA measurement sees?
  • Is the drop in recovery from your dsRNA removal built into the cost estimate?

Step 8: UF/DF, buffer exchange

A finishing step that removes salts, small molecules and ethanol and adjusts to the target concentration and composition, deciding LNP formation and storage stability.A finishing step that removes salts, small molecules and ethanol and adjusts to the target concentration and composition, deciding LNP formation and storage stability.

Standard methodBuffer exchange and concentration by ultrafiltration/diafiltration (UF/DF, TFF)Buffer exchange and concentration by ultrafiltration/diafiltration (UF/DF, TFF)
Details and 3 alternatives

This step coversBuffer exchange and concentration by ultrafiltration/diafiltration (TFF) (UF/DF, buffer exchange)

A finishing step that readies the purified mRNA to be handed to the next step. Salts, small molecules and leftover ethanol are removed and the material is brought to the target concentration and composition.

The method is TFF (ultrafiltration/diafiltration). While the mRNA is retained by the membrane, unneeded small molecules are washed away, and the material is concentrated while being exchanged into a buffer suited to LNP formation and storage.

This is the 'bridge' from purification to formulation. If buffer and concentration are not aligned, the later LNP formation is not stable, so it is a key point that affects both yield and stability.

Salts, small molecules and ethanol are removed from the purified mRNA, and it is adjusted to the target concentration and composition. This finishing step replaces the buffer with one suitable for LNP formation and storage, and affects yield and stability.

MakersCytiva, Merck, Sartorius, Pall, Repligen

Why it is the standard. TFF is the standard operation for concentrating biomolecules and exchanging buffer, and has a track record of application in the manufacture of approved COVID-19 mRNA vaccines (Creative Enzymes; US patent literature). Widely adopted as the final adjustment step for mRNA drug substance.

Alternatives in practice

  • Single-pass TFF (SPTFF) · Repligen KrosFlo SPTFF/Pall Cadence Inline Concentrator — High-fold concentration without recirculation, with reduced shear and shorter residence time.
  • Size exclusion/desalting chromatography · Cytiva ÄKTA pure+HiPrep 26/10 Desalting/HiTrap Desalting — Buffer exchange at small to medium scale. Dilution is involved, so concentration is needed afterward.
  • Dialysis · Thermo Slide-A-Lyzer G2/Spectrum Spectra/Por dialysis membranes — Lab scale only. Time-consuming and cannot be scaled up.

Formulation and fill-finish(4 steps)

Step 9: Lipid phase

Because the aqueous phase and the lipid ethanol phase must be precisely prepared according to the N/P ratio before microfluidic mixingBecause the aqueous phase and the lipid ethanol phase must be precisely prepared according to the N/P ratio before microfluidic mixing

Standard methodTwo-phase preparation in which mRNA drug substance is diluted/prepared in low-pH acetate/citrate buffer (pH4–5) and a lipid phase dissolved separately in ethanol is made ready (the stage before rapid microfluidic mixing)Two-phase preparation in which mRNA drug substance is diluted/prepared in low-pH acetate/citrate buffer (pH4–5) and a lipid phase dissolved separately in ethanol is made ready (the stage before rapid microfluidic mixing)
Details and 2 alternatives

This step coversPreparation of two-phase solutions before formulation → an aqueous phase in which mRNA drug substance is diluted/prepared in low-pH acetate/citrate buffer (pH4–5), and an organic phase in which ionizable lipid, cholesterol, DSPC and PEG lipid are dissolved in ethanol at a defined molar ratio, are each prepared and filtered to match the defined N/P ratio (lipid phase/aqueous phase preparation)

The preparation of two phases immediately before making LNP (lipid nanoparticles). On one side, the 'aqueous phase' in which the drug substance mRNA is dissolved in low-pH (pH4–5) acetate/citrate buffer is prepared.

On the other side, the 'organic phase' is made by dissolving the 4 components, ionizable lipid, cholesterol, DSPC and PEG lipid, in ethanol at a defined ratio. Both are prepared separately.

Particle formation occurs by instantaneously mixing the two phases, so strictly matching the N/P ratio and lipid molar ratio and pre-filtering before mixing is the starting point of quality.

LNP formation is carried out by instantaneously mixing mRNA in a low-pH aqueous phase and 4-component lipids in ethanol at a fixed N/P ratio and flow rate ratio, so it is essential to strictly prepare and pre-filter both phases for concentration, pH and lipid molar ratio (e.g., ionizable lipid:DSPC:Chol:PEG = 50:10:38.5:1.5, N/P ≈ 6) immediately before mixing.

MakersMerck, Sartorius, Pall, Watson-Marlow, Avanti

Why it is the standard. A standard configuration using pharmaceutical-grade lipid suppliers and pre-filtration filters, allowing phase preparation immediately before mixing and N/P ratio setting to be done with high reproducibility

Alternatives in practice

  • Aqueous phase preparation in a citrate buffer system · The preparation system above + 50 mM sodium citrate pH4.0 — An alternative to the acetate system. Chosen by optimizing buffering capacity and encapsulation efficiency.
  • Premix the lipids and supply them as a single lipid phase · Stainless steel preparation vessel + stirring + 0.2 µm PTFE/PES filter — The 4 lipids are dissolved in ethanol and premixed. Ensures consistency of the mixing ratio.

Step guide

What you actually do

  1. Dilute the mRNA drug substance to the target concentration with a sodium acetate or citrate buffer (pH 4 to 5 in examples) to make the aqueous phase
  2. Dissolve the ionizable lipid, the helper lipid DSPC, cholesterol and the PEG lipid in ethanol at the specified molar ratio to make the organic phase
  3. From the volumes and concentrations of the two phases, confirm from the weighed values and the concentration measurements that the ratio of lipid amine to mRNA phosphate (the N/P ratio) will come out as intended
  4. Pass both phases through a 0.2 µm filter to reduce particles and bioburden, and put them into the vessels that feed the mixing unit

Decision points

An acetate-based or a citrate-based buffer for the aqueous phase?
Both are at low pH and are used to give the ionizable lipid a positive charge so that it attracts the mRNA. Choose on buffering capacity, encapsulation efficiency and stability in subsequent storage, deciding from data compared during formulation development.
Do you hold the lipids as separate solutions, or as 1 premixed solution?
Premixing makes it easier to secure consistency of the mixing ratio, but you have to settle the stability of the premixed solution during storage and the control of its hold period. Separate solutions let you adjust the composition freely, but add more liquid feed lines.
Do you measure out by volume or by weight?
Ethanol differs from water in density and is also volatile, so configurations that measure and control by weight are used. Linking this to a balance or a load cell lets the weighed values be kept as electronic records. Where volume is converted to weight, write it into the procedure so that the density is not mistaken.
Which membrane do you filter the ethanol solution through?
The aqueous phase uses a hydrophilic membrane such as PES, and the ethanol solution a membrane whose solvent compatibility has been confirmed (PTFE, PES and the like). Confirm compatibility between membrane material and solvent against supplier data, and measure the loss from lipid adsorption with your own solutions.

Common mistakes

  • Not setting a hold period for the premixed lipid solution, and carrying on using it with the concentration shifted by precipitation or degradation
  • Lipid or mRNA adsorbing to the membrane during filtration of the aqueous or organic phase, so the ratio shifts
  • Mistaking the density of ethanol and getting the conversion between volume and weight wrong

Check at your own site

  • From which records (weighed values, concentration measurements) does your two-phase preparation confirm the N/P ratio?
  • Do your components in contact with ethanol have extractables data for that solvent and those conditions?

Step 10: LNP encapsulation, mixing

The core step that determines particle size, uniformity and encapsulation efficiency: aqueous-phase mRNA and lipid phase are mixed rapidly and the particles self-assemble.The core step that determines particle size, uniformity and encapsulation efficiency: aqueous-phase mRNA and lipid phase are mixed rapidly and the particles self-assemble.

Standard methodLNP self-assembly by rapid microfluidic (or turbulent jet) mixingLNP self-assembly by rapid microfluidic (or turbulent jet) mixing
Details and 3 alternatives

This step coversEncapsulation of mRNA into lipid nanoparticles (LNP) (microfluidic/turbulent mixing) (LNP encapsulation, mixing)

An encapsulation step that wraps mRNA, which hardly enters cells and is fragile when naked, in particles made of 4 lipids. It can be called the core unique to LNP manufacturing.

When the aqueous-phase mRNA and the ethanol-phase lipids are mixed rapidly by microfluidics or turbulence, the lipids gather spontaneously to form particles and trap the mRNA inside.

Conditions such as mixing flow rate and ratio determine particle size, uniformity and encapsulation efficiency. How well this is engineered greatly affects the delivery performance of the product, and it is the decisive point.

mRNA in its naked form hardly enters cells and is easily broken, so it is wrapped in particles made of 4 lipids. When the mRNA in the aqueous phase and the lipids in the ethanol phase are mixed rapidly, the particles self-assemble and trap the mRNA inside. It is the core step unique to this product, where the way of mixing determines particle size, uniformity and encapsulation efficiency.

MakersCytiva, Merck, Avanti, MedChemExpress, Microfluidics

Why it is the standard. The self-assembly of RNA-LNP by solvent exchange through rapid mixing of the aqueous-phase mRNA and the ethanol lipid phase is the standard for mRNA-LNP formulations (Pharmaceutics 2025, PMC12473834). NanoAssemblr's microfluidics (herringbone/toroidal mixing) is widely used from research to GMP.

Check also. Encapsulation is the core step in which the mixing conditions determine particle size, uniformity and encapsulation efficiency, but when moving from lab microfluidics to commercial scale, throughput tends to become the limiting factor. At scale-up, flow rate ratio and total flow rate are reoptimized, and as a rule small-scale conditions are not carried over as they are.

Alternatives in practice

  • T/impinging jet mixer (turbulent mixing) · Knauer Impingement Jets Mixer (IJM NanoScaler)/in-house stainless steel T mixer + HPLC pump — Advantageous for high flow rates and large scale. The equipment is simpler than microfluidics.
  • Preformed vesicle (PFV) method · Lipid thin film hydration + extruder (Avanti Mini-Extruder/LIPEX), followed by mRNA mixing — mRNA is taken up into preformed liposomes. High degree of freedom in process control.
  • High-pressure microfluidization · Microfluidics LM20/M-110P Microfluidizer — High shear gives uniform fine particles. Suited to mass production, but the equipment burden is high.

Step guide

What you actually do

  1. Feed the aqueous and the organic phase into the mixer at the same time, at the set flow rate ratio (around 3 to 1 as an example that makes the aqueous phase the thicker stream) and total flow rate
  2. The two liquids mix rapidly inside the mixer, the ethanol is diluted, the lipids self-assemble, and particles that hold the mRNA inside them form
  3. Receive the emerging liquid straight into the dilution and neutralization vessel to stop the self-assembly and stabilize the particles
  4. Handle the liquid from the start-up and the end of the run (whether it goes into the product or is kept apart) as the procedure says, and draw samples

Decision points

Microfluidic, or impingement jet (or T-junction)?
Microfluidics is reproducible at small volumes and suits development, but there is a limit to how much a single channel can process. Impingement jet mixing readily takes throughput and has been used at commercial scale. The T-junction sits between the two. There are reports that particle size, encapsulation efficiency and internal structure changed with the mixing method even for the same formulation, so confirm in advance the relationship between the method used in development and the method of the production machine.
Which operating conditions do you move?
The main conditions that move particle size are the flow rate ratio (the proportion of the aqueous phase), the total flow rate (how fast the mixing is), the lipid concentration and the N/P ratio (around 6 in examples). The total flow rate has a region where going faster no longer changes the particle size, and putting the operating point there makes the process robust against fluctuations in flow. Choose not only for the target particle size but for an operating point robust against fluctuation.
How do you scale up?
There is a method that keeps the conditions of one mixing unit and puts units side by side to gain throughput, and a method that redesigns the jet geometry and flow velocity to make it larger. For either, measure particle size, distribution and encapsulation efficiency again on the scaled-up equipment. Changing the mixing method needs equivalence to be explained, so put that into the plan.
Is the mixing part a single-use cartridge, or washed and reused?
A single-use cartridge needs no cleaning verification, but confirm the documentation for sterilization and for the absence of animal-derived components, and the lot-to-lot reproducibility of its dimensions. A mixing part that is washed and reused carries the burden of component lifetime and of confirming that there is no carry-over after cleaning.

Common mistakes

  • Carrying the conditions set on a laboratory microfluidic device straight over to a production machine, so the particle size and encapsulation efficiency shift
  • Putting the operating point short of where the total flow rate plateaus, so that a slight difference in flow becomes a difference in particle size
  • Comparing figures from manual mixing or a development mixer with figures from the production machine on the same scale

Check at your own site

  • Is the mixer you use for formulation studies the same type of mixing method as the production machine? If not, which attributes do you plan to compare at transfer?
  • Is the operating point of your LNP encapsulation in the region where particle size moves little even if the flow fluctuates?

Step 11: Ethanol removal, concentration

Because formulation is not possible unless residual ethanol and low pH are removed and the suspension is exchanged into a final buffer at physiological pHBecause formulation is not possible unless residual ethanol and low pH are removed and the suspension is exchanged into a final buffer at physiological pH

Standard methodFlat-sheet cassette TFF after dilution and neutralizationFlat-sheet cassette TFF after dilution and neutralization
Details and 2 alternatives

This step coversDilution and neutralization + ethanol removal, buffer exchange and concentration by TFF → the low-pH, ethanol-containing LNP suspension immediately after microfluidic mixing is diluted and neutralized with buffer to stabilize the particles, ethanol is removed by ultrafiltration/diafiltration and exchanged into the final formulation buffer, the suspension is concentrated to the target mRNA concentration (1–3 mg/mL), and bioburden-reduction filtration follows (ethanol removal, concentration)

Immediately after mixing, LNP is a thin suspension containing about 7% ethanol and low pH, and cannot yet be used as is. First it is diluted and neutralized with buffer to settle the particles.

The following TFF (ultrafiltration/diafiltration) removes the ethanol and concentrates to the target concentration (1–3 mg/mL) while exchanging into a final formulation buffer at physiological pH containing cryoprotectant.

In other words, this step 'shapes freshly made particles into the form of a formulation that can be injected.' It is a real formulation step that must be passed before proceeding to sterile filtration and filling.

Immediately after mixing, LNP is a dilute suspension of about 7% ethanol at low pH (mRNA about 0.03 mg/mL) and cannot be formulated and filled as is. Removing ethanol by TFF (UF/DF), exchanging into the final formulation buffer (physiological pH, containing cryoprotectant) and concentrating to the target concentration is a mandatory standard step before sterile filtration and filling.

MakersCytiva, Merck, Sartorius, Pall, Repligen

Why it is the standard. A real equipment configuration in which low-shear TFF can carry out ethanol removal, buffer exchange and concentration together while maintaining the particle size and encapsulation efficiency of LNP; the standard workflow of commercial mRNA-LNP manufacturing

Alternatives in practice

  • Ethanol removal, diafiltration and concentration by hollow fiber TFF · Cytiva ÄKTA flux+UFP-100-E-55 hollow fiber/Repligen KrosFlo mPES 100 kDa — Gentle shear on the LNP. Flux may be somewhat lower than with flat sheet.
  • Immediate dilution/neutralization into a high-pH buffer + dialysis (small scale) · Neutralization buffer tank + Thermo Slide-A-Lyzer dialysis — For early development/small-volume studies. Cannot be scaled up.

Step guide

What you actually do

  1. Dilute and neutralize the liquid just after mixing with buffer, either in line (through a static mixer) or in a tank, to stabilize the particles
  2. Circulate along a membrane of around 100 kDa in TFF, holding the LNP while ethanol and free mRNA pass through to the other side of the membrane
  3. Add final formulation buffer (in some examples containing sucrose as a cryoprotectant) to replace what has passed through, and concentrate to the target mRNA concentration (1 to 3 mg/mL in examples)
  4. Pass through a 0.2 µm bioburden reduction filter and collect in the hold vessel

Decision points

Dilution and neutralization in line, or in a tank?
The liquid just after mixing is acidic and contains ethanol, and the particles are unstable as it is. An in-line static mixer can shorten the time from mixing to neutralization and suits continuous operation. Doing it in a tank needs simpler equipment and is used at small scale.
A flat sheet cassette, hollow fiber, or single-pass TFF?
Hollow fiber gives gentler shear to the LNP, while a flat sheet cassette tends to give a higher flux. Single-pass TFF is presented by vendors as able to concentrate and remove ethanol without recirculation. Whichever the format, choose by confirming with the real liquid that particle size and encapsulation efficiency are preserved and that recovery is as expected.
What decides the amount of diafiltration and its end point?
Ethanol has to be brought below the residual solvent limit (ethanol is Class 3 in ICH Q3C). Set the number of diavolumes needed with the real liquid for each formulation, and judge the end point on actual measurement of residual ethanol.
What is this filtration aimed at?
This is filtration to lower bioburden, not to assure sterility. The final sterile filtration sits separately, before filling. Confirm that the LNP, being particles, are not taken up by the membrane or plugging it, by comparing recovery before and after filtration and comparing particle size and encapsulation efficiency.

Common mistakes

  • Concentrating too far, or shearing too strongly, so the particles aggregate and the distribution broadens
  • Comparing the encapsulation efficiency after free mRNA has passed through with the ethanol in TFF against the value just after mixing, on the same scale
  • Going on without confirming whether particles are taken up by the membrane during filtration and the encapsulation efficiency or concentration shifts across it

Check at your own site

  • Which time point, before or after TFF, gives the encapsulation efficiency value you put in the specification?
  • Is the end point of your ethanol removal set by the diavolume calculation or by actual measurement?

Step 12: Fill-finish

0.2 µm sterile filtration and aseptic filling, stabilizing fragile LNP by optimizing freezing/lyophilization conditions.0.2 µm sterile filtration and aseptic filling, stabilizing fragile LNP by optimizing freezing/lyophilization conditions.

Standard method0.2 µm sterile filtration + aseptic filling (with lyophilization/freezing as needed)0.2 µm sterile filtration + aseptic filling (with lyophilization/freezing as needed)
Details and 3 alternatives

This step coversSterile filtration, aseptic filling and lyophilization/freezing (fill-finish)

A finishing step that decides the final form of the product. The finished LNP dispersion is sterilized with a 0.2 µm filter and aseptically filled into vials.

LNP is fragile to freezing and drying. So conditions for frozen storage and lyophilization are optimized, so that particle size and encapsulation are preserved even after thawing and reconstitution.

Only when these 2, 'how to keep sterility' and 'how to bring it to a storage form without breaking it,' are both achieved does it become a product that reaches patients. It is the final gate of manufacturing.

The finished LNP dispersion is sterilized with a 0.2 µm filter and aseptically filled into vials. LNP is easily damaged by freezing and drying, so freezing or lyophilization conditions are optimized so that particles and encapsulation are preserved even after thawing.

MakersMerck, Sartorius, Pall, Eppendorf, Getinge

Why it is the standard. LNP-mRNA is sterile-filtered and aseptically filled as a sterile final product, and stabilized by freezing/lyophilization. Optimizing the cryoprotectant is essential for maintaining particle size and encapsulation (mRNA-LNP formulation review, PubMed 40069324).

Check also. 0.2 µm sterile filtration can leave part of the LNP on the membrane and shift encapsulation efficiency and concentration slightly, so quality is confirmed before and after filtration. Freezing and lyophilization easily damage the particles, and verification that particle size and encapsulation are preserved after thawing and redissolution is needed separately.

Alternatives in practice

  • Liquid supply by frozen (−20 to −70 °C) storage · Ultra-low temperature freezers (Thermo TSX/Eppendorf CryoCube) + single-use bio-containers (Sartorius/Thermo) — Avoids lyophilization. Presupposes cold-chain logistics (COVID mRNA vaccine type).
  • Spray drying/thin-film drying · Büchi Mini Spray Dryer B-290/GEA spray dryers — Aims for room-temperature stabilization. Evaluation of the effect on LNP properties is essential.
  • Aseptic processing manufacture for large-particle-size systems that are difficult to filter · Class A isolator/RABS (Getinge/SKAN) + aseptic preparation — The last resort for systems that cannot be 0.2 µm filtered. Validation burden is heavy.

Step guide

What you actually do

  1. Confirm the integrity of the sterilized sterilizing filter before use (PUPSIT, pre-use integrity testing)
  2. Pass the LNP solution through a 0.2 µm sterilizing-grade filter and send it to the filling machine inside the isolator or RABS
  3. Fill the set volume into sterilized and depyrogenated vials, insert the rubber stopper, and seal with a cap
  4. For storage frozen, move to an ultra-low temperature freezer or controlled freezing; for lyophilization, load the shelves and run freezing, primary drying and secondary drying
  5. Record the integrity of the filter after use, the fill volumes and the environmental monitoring results

Decision points

Freeze as a liquid, or lyophilize?
Frozen storage as a liquid presumes a cold chain that holds a low temperature in the -20 to -70 °C range (as an example). Lyophilization can raise the temperature range in which the product can be stored, but it needs cycle development and equipment, and the particle size, encapsulation efficiency and integrity of the mRNA after reconstitution have to be confirmed. Whether near-ambient temperature can be aimed at with mRNA-LNP depends on the formulation and the process and cannot be generalized.
What do you look at to choose a freeze dryer?
Shelf area is a measure of how many vials fit in 1 batch; it says nothing about how far ice crystals can be made uniform, or how far product temperature can be measured during drying. Compare on whether controlled nucleation and PAT that measures product temperature can be carried over from the development pilot unit to the production machine.
On what do you compare sterilizing filters, and what do you do about testing before use?
The 0.22 µm and 0.2 µm ratings are much the same across suppliers and do not decide a comparison. Compare on adsorption and throughput as set by membrane material and construction (single layer or double layer), and on how the acceptance value of the integrity test is derived. EU GMP Annex 1 asks as a rule for confirmation of integrity after sterilization and before use, and allows an alternative based on a risk assessment where that cannot be done. For a liquid containing particles, include in the assessment whether plugging during filtration can move the test value after use.
Is part of the LNP taken up during filtration?
In 0.2 µm filtration, LNP can be taken up by the membrane and the encapsulation efficiency or the concentration can shift slightly. Measure particle size, encapsulation efficiency and mRNA concentration before and after filtration, and compare the difference when membrane material and construction are changed, using your own solution.

Common mistakes

  • Judging how the lyophilization turned out from the appearance of the cake alone, and not confirming the particle size and encapsulation efficiency after reconstitution
  • Carrying on using a buffer in which the pH can swing on freeze concentration, such as PBS, as the standard composition
  • Not looking at the loss of particles in filtration, and treating the post-filtration concentration as a calculated value

Check at your own site

  • Is there a record confirming the particle passage and recovery of your sterile filtration with the LNP solution?
  • By which attributes after reconstitution does your freezing or lyophilization show validity?

Quality attributes and analytics(8 steps)

Step 13: Integrity evaluation

Fragmented mRNA is not translated and efficacy drops, so the full-length percentage (% full-length) is measured by CGE as a quality indicator.Fragmented mRNA is not translated and efficacy drops, so the full-length percentage (% full-length) is measured by CGE as a quality indicator.

Standard methodFull-length percentage and smear analysis by capillary gel electrophoresis (CGE)Full-length percentage and smear analysis by capillary gel electrophoresis (CGE)
Details and 3 alternatives

This step coversEvaluation of mRNA integrity (full-length percentage) (integrity evaluation)

A quality test that measures 'how much of the mRNA remains full-length (% full-length).' mRNA is a long molecule and tends to break during manufacturing and storage.

Fragments broken partway are not translated, and efficacy drops. Making the degree of breakage visible is therefore indispensable to assuring potency.

Capillary gel electrophoresis analyzes the full-length percentage and the smear (the spread of degraded fragments). It is not a manufacturing step itself, but a step that provides the yardstick for judging quality.

mRNA is a long molecule and is easily broken during manufacturing and storage. Fragments broken partway are not translated and efficacy drops, so 'how much remains full-length (% full-length)' is measured as a quality indicator.

MakersBio-Rad, Agilent

Why it is the standard. Fragment Analyzer is an array-type CGE that quantifies the %area of the full-length transcript (integrity) by ProSize smear analysis, and is the standard method (Agilent technical documentation). Widely adopted as the main QC method for mRNA integrity.

Check also. Even if the full-length percentage (% full-length) is high, uncapped or poly(A)-shortened molecules slip through this test. If integrity is not viewed together with capping efficiency (#2) and poly(A) tail length (#14), it is easy to jump to the conclusion that quality is assured.

Alternatives in practice

  • Integrity profiling by IP-RP-HPLC/UV · Thermo Vanquish/Agilent 1290 Bio LC+DNAPac RP, TEA/HFIP — High-resolution separation of full-length and degraded fragments. Sequence confirmation is also possible with MS coupling.
  • Denaturing agarose/glyoxal gel electrophoresis · Bio-Rad Sub-Cell GT+ChemiDoc, NorthernMax-Gly (Thermo) reagents — Low-cost orthogonal method. Throughput and quantitation are low.
  • Automated scoring such as RIN · Agilent 2100 Bioanalyzer (automatic RIN/RINe calculation) — Reports integrity as a single index. Detailed smear needs separate analysis.

Step guide

What you actually do

  1. Dilute the sample to the specified concentration and heat to denature it as the kit specifies, to unfold the secondary structure
  2. Load it, together with the RNA kit and a size standard (ladder), onto a capillary electrophoresis or gel tape instrument and run it
  3. From the trace (electropherogram), read the position and the area proportion of the main peak, and the proportion of smear (tailing) on the short-chain side
  4. Record the integrity score the instrument calculates together with the image of the trace

Decision points

Capillary type, or gel tape type?
Models with many capillaries suit running many samples together, for process control or lot screening. The gel tape type suits bench work that puts a few samples through quickly. There are models where 16 samples are loaded on 1 tape and the run finishes in about 30 minutes.
Can you decide pass or fail on the integrity score?
RQN and RINe are measures originally created for total RNA. For a single mRNA species, use them for relative comparison between lots or between conditions under the same conditions, rather than deciding pass or fail on the absolute value. Define pass or fail by reading the trace itself, for example the area proportion of the main peak.
Can a long construct be measured with the same kit?
For a construct near the top of the size range the kit can measure, fragments close to full length cannot be fully separated and the separation becomes difficult. For chains reaching 9,000 to 12,000 nt, such as self-amplifying ones, look first at the change in the shape of the trace rather than at the movement of the score.
How do you divide the work with other analytical methods?
Automated electrophoresis plays the part of a net that watches broadly in a short time. Identification of trace by-products, and separation of impurities close in size, are passed to ion-pair reverse-phase HPLC. How the cap is added and the length of the poly(A) tail are measured separately by dedicated methods.

Common mistakes

  • Recording the score alone and keeping no trace (samples with the same score but different shapes cannot then be told apart)
  • Reading a high proportion of full-length material as grounds for there being no problem even with uncapped material or short poly(A) tails
  • Carrying a score guide meant for total RNA (such as 7 or above) straight over to a single mRNA species

Check at your own site

  • Is your integrity specification defined by the score or by the proportion of the main peak?
  • Does the length of your mRNA sit comfortably within the range the kit you use can measure?

Step 14: Identity, sequence confirmation

An identity test, directly tied to safety and efficacy, that prevents mix-ups and mutations and confirms the intended sequence.An identity test, directly tied to safety and efficacy, that prevents mix-ups and mutations and confirms the intended sequence.

Standard methodConfirmation of sequence identity by reverse transcription sequencing (next-generation sequencing/RT-PCR)Confirmation of sequence identity by reverse transcription sequencing (next-generation sequencing/RT-PCR)
Details and 3 alternatives

This step coversIdentity and sequence confirmation (identity, sequence confirmation)

An identity test that confirms the mRNA made is 'the intended molecule' and that 'there is no error in the sequence.' It is a check to prevent mix-ups and mutations.

In particular, whether the sequence of the protein produced by translation is correct is directly tied to safety and efficacy. A one-letter difference can change the effect.

After reverse transcription, the sequence is read by next-generation sequencing or RT-PCR to verify that it is as designed. It is an analytical step that assures not manufacturing but 'whether the contents are the real thing.'

Confirms that it is 'the intended mRNA' and that 'there is no error in the sequence.' It is a check to prevent mix-ups and mutations, and in particular whether the sequence of the translated protein is correct is directly tied to safety and efficacy.

MakersBio-Rad, Waters, Illumina, NEB, Oxford Nanopore

Why it is the standard. Sequencing after RT is used to confirm identity and mutations across the whole sequence, and is the core of regulatory identity testing. RT-qPCR is also used for rapid identity/content confirmation.

Alternatives in practice

  • Oligo mapping/sequence confirmation by LC-MS · Thermo Vanquish+Orbitrap Exploris 240/Waters BioAccord+DNAPac RP — Directly confirms sequence and modifications from the mass of RNase-digested fragments. Independent of NGS.
  • Identity confirmation by RT-qPCR · QuantStudio 5 + sequence-specific TaqMan probes — A rapid, low-cost ID test. Cannot judge the full-length sequence.
  • Restriction/probe hybridization · Hybridization system (Bio-Rad blotting) + sequence-specific probes — An orthogonal method to confirm the presence of the target sequence.

Step guide

What you actually do

  1. Convert the mRNA to cDNA with reverse transcriptase
  2. Prepare a sequencing library under conditions that can read the full length and run it on a next-generation sequencer (looking at a specific region by RT-PCR is also a method)
  3. Overlay the sequence read on the designed sequence and analyse it for substitutions, deletions and insertions
  4. Record, with the results, the version of the analysis pipeline used and the reference sequence

Decision points

NGS, RT-PCR, or oligo mapping by LC-MS?
If you want to confirm the full-length sequence, NGS. RT-PCR is rapid and low in cost and can be used to confirm identity, but cannot judge the full-length sequence. Oligo mapping by LC-MS can confirm sequence and modifications directly from the masses of RNase-digested fragments, without depending on NGS.
Can misreading occur in the reverse transcription?
Misreads and stops can arise at the reverse transcription stage, so run RNA of known sequence alongside and get to know how errors of the method appear. For mRNA using modified nucleotides, confirm as well that the modification does not affect the reading.
How far do you require it to go as an identity test?
Whether the full length is read for every lot, or the full length is confirmed at the development stage and identity confirmed thereafter by a specific method, is set by the approved specification and test method, agreement with the regulatory authorities, and the policy of the quality unit.
How do you show the reliability of the analysis?
Treat everything from library preparation to analysis as one test method, and include in the scope of validation not only standardization of the procedure, prevention of contamination and record control, but also control of false positives and false negatives.

Common mistakes

  • Reporting a misread as a sequence mutation, without confirming the errors of reverse transcription or the conditions of the analysis
  • Not fixing the version of the analysis pipeline, so the results cannot be reproduced
  • Treating an RT-qPCR identity test as having confirmed the whole of the sequence

Check at your own site

  • Which range of the sequence is your identity test approved as confirming?
  • Does your analysis pipeline have records of version control and verification?

Step 15: Poly(A) tail analysis

Poly(A) tail length affects stability and translation efficiency, so the distribution is confirmed by HPLC/MS as intended.Poly(A) tail length affects stability and translation efficiency, so the distribution is confirmed by HPLC/MS as intended.

Standard methodPoly(A) tail length analysis by enzymatic digestion + IP-RP-HPLC/LC-MSPoly(A) tail length analysis by enzymatic digestion + IP-RP-HPLC/LC-MS
Details and 3 alternatives

This step coversEvaluation of poly(A) tail length (poly(A) tail analysis)

A test that measures whether the poly(A) tail at the end (3' end) of mRNA falls within the designed length. Tail length is a hidden determinant of quality.

The length of the poly(A) tail affects the stability and translation efficiency of mRNA. Efficacy drops whether it is too short or the lengths are scattered.

The mRNA is cut out with an enzyme, and the tail length distribution is analyzed by IP-RP-HPLC or LC-MS. An analytical step that confirms 'whether the lengths are aligned as targeted.'

The length of the 3' poly(A) tail affects the stability and translation efficiency of mRNA. Efficacy drops whether it is too short or too heterogeneous, so confirm that the length distribution is as intended.

MakersAgilent, Waters, SCIEX, NEB, Honeywell

Why it is the standard. The 3' poly(A) tail is generally evaluated by IP-RP-HPLC and LC-MS, which requires upstream enzymatic digestion/fragmentation (LCGC; chromatographyonline). A one-step method that evaluates the cap and poly(A) chain length simultaneously with DNAzyme has also been reported (J Pharm Biomed Anal 2025, S0731708525000366).

Alternatives in practice

  • Simultaneous cap + poly(A) analysis by the DNAzyme method · The LC-MS above (Waters BioAccord) + site-specific DNAzyme — Evaluates the 5' cap and the poly(A) tail together. Makes sample preparation more efficient.
  • High-resolution capillary electrophoresis · SCIEX PA800 Plus/Agilent Fragment Analyzer (CGE) — Evaluates the poly(A) length distribution by size separation. No MS needed.
  • Oligo-dT excision + gel analysis · After RNase H + oligo-dT guide digestion, Agilent Bioanalyzer/TapeStation — Simple tail length screening. Resolution is inferior to LC-MS.

Step 16: Particle size, zeta potential

Particle size, PDI and zeta potential affect distribution in the body, uptake and stability, so they are measured by DLS/ELS for every release.Particle size, PDI and zeta potential affect distribution in the body, uptake and stability, so they are measured by DLS/ELS for every release.

Standard methodParticle size and PDI by dynamic light scattering (DLS), zeta potential by electrophoretic light scatteringParticle size and PDI by dynamic light scattering (DLS), zeta potential by electrophoretic light scattering
Details and 3 alternatives

This step coversLNP particle size, polydispersity index (PDI) and zeta potential (particle size, zeta potential)

A test that measures the 'size, uniformity (PDI) and surface charge (zeta potential)' of LNP particles. These 3 are the basic indicators reflecting how the particles behave in the body.

Particle size, uniformity and charge affect distribution in the body, uptake and stability. If particles are too large, heterogeneous or have unsuitable charge, delivery efficiency and safety are affected.

Particle size and PDI are measured by dynamic light scattering and zeta potential by electrophoretic light scattering at every release. An analytical step that assures particle quality numerically.

Particle size, uniformity (PDI) and surface charge affect distribution in the body, uptake and stability. Particles that are too large, heterogeneous or have unsuitable charge affect delivery efficiency and safety, so they are measured for every release.

MakersWaters, Malvern Panalytical, Wyatt, Anton Paar

Why it is the standard. Measuring particle size, PDI and zeta potential by DLS/ELS on a Zetasizer Ultra is the standard for LNP characterization (Pharmaceutics; barnett-technical). Low dilution and speed suit release testing, and it is widely adopted in the industry.

Alternatives in practice

  • Nanoparticle tracking analysis (NTA) · Malvern Panalytical NanoSight Pro/NS300 — Visualizes particle size and concentration particle by particle. Advantageous for low-concentration, polydisperse samples.
  • Cryo-electron microscopy (cryo-TEM) · Thermo Glacios/Talos Arctica+Vitrobot — Direct observation of internal structure and morphology. For confirming the encapsulation state. Low throughput.
  • SEC-MALS/FFF-MALS · Waters/Wyatt DAWN MALS + ÄKTA/HPLC SEC, or Postnova AF2000 FFF — Obtains absolute molecular weight and size distribution after separation. Strong for evaluating aggregates.

Step guide

What you actually do

  1. Dilute the sample with the set buffer (Tris and the like) and put it into the measurement cell without introducing bubbles or dust
  2. Derive the mean particle size and the PDI by dynamic light scattering (DLS, a method that derives size from the speed of Brownian motion of the particles)
  3. For zeta potential, dilute in a liquid of low conductivity and measure the surface charge by electrophoretic light scattering
  4. Record the DLS results alongside complementary methods such as nanoparticle tracking analysis (NTA)

Decision points

Is DLS alone enough?
The DLS signal is pulled strongly by large particles, and the mean and the PDI jump with a very small number of aggregates. If the values worsen, separate whether the whole population has grown or aggregates have come in, using NTA or SEC-MALS. For direct observation of morphology cryo-TEM is used, but it is a method with low throughput.
How do you set the dilution conditions?
The values of particle size and zeta potential move with the dilution buffer and the dilution factor. Zeta potential is hard to measure at high conductivity, so the sample is diluted in a low-conductivity liquid, but diluting changes the pH and the ionic strength. The value depends on pH, so record the conditions along with it.
How do you set the specification?
There are examples of practice with a PDI of roughly 0.2 or below, and around 0.1 when good. It is not a single universal value, however: set it together with particle size and encapsulation efficiency, from the track record of lots whose efficacy has been confirmed and from the measurement scatter.

Common mistakes

  • Misreading a jump in PDI caused by a small number of aggregates as a drop in quality of the whole
  • Comparing values measured under changed dilution conditions with past lots as they stand
  • Looking only at the mean particle size and not checking the shape of the distribution

Check at your own site

  • Under which dilution conditions is your PDI specification written as having been measured?
  • When your DLS figures move, by which method do you separate aggregates coming in from a change in the whole population?

Step 17: Encapsulation efficiency, content

Naked mRNA is unstable and stimulates immunity, so content and the in-particle encapsulation efficiency (EE%) are checked against specification with RiboGreen.Naked mRNA is unstable and stimulates immunity, so content and the in-particle encapsulation efficiency (EE%) are checked against specification with RiboGreen.

Standard methodQuant-iT RiboGreen fluorescence assayQuant-iT RiboGreen fluorescence assay
Details and 3 alternatives

This step coversmRNA content and encapsulation efficiency (EE%) (encapsulation efficiency, content)

A test that simultaneously measures the amount of mRNA in 1 vial (content) and what percent of it is protected inside the particles (encapsulation efficiency/EE%).

Unencapsulated naked mRNA is fragile and stimulates the immune system. So it is important that both the 'amount' and the 'protected fraction' are within specification.

A fluorescent reagent called RiboGreen is used, and both values are obtained from the fluorescence difference with and without detergent. An analytical step directly tied to delivery and safety.

Measure how much mRNA is in 1 vial (content) and what percent of it is protected inside the particles (encapsulation efficiency). Naked, unencapsulated mRNA is fragile and stimulates the immune system, so it is important that both are within specification.

MakersAgilent, Tecan, SCIEX, Molecular Devices

Why it is the standard. The modified RiboGreen method is a well-established method widely used to determine the concentration and encapsulation efficiency of LNP-encapsulated mRNA (mRNA-LNP formulation review). Free and total mRNA are obtained from the fluorescence difference with and without detergent, and EE% is calculated.

Check also. Content and encapsulation efficiency are normally measured by a fluorescent dye assay, but lipids and contaminants can interfere with fluorescence and shift the values. For critical lots, it is reassuring to confirm content by another method (such as quantitation after separation) and cross-check.

Alternatives in practice

  • CGE-LIF (capillary gel electrophoresis with laser-induced fluorescence detection) · SCIEX PA800 Plus+LIF detection/Agilent Fragment Analyzer — Evaluates the size and amount of encapsulated mRNA at the same time. Integrity information is also obtained.
  • Free/total RNA quantitation using UV (A260) · Thermo NanoDrop One/plate UV (SpectraMax) — Calculates encapsulation efficiency from the A260 difference with and without detergent. No dye needed; simple.
  • RiboGreen improved method using sustainable surfactants · The fluorescence reader above + alternative solubilizer (e.g., deoxycholic acid type) — Replaces Triton X-100 (subject to REACH regulation). Improves the robustness of encapsulation efficiency measurement.

Step guide

What you actually do

  1. Split the sample in two, leaving one half as it is and breaking the LNP in the other with a surfactant (Triton X-100 and the like)
  2. Add a dye that fluoresces strongly when it binds to RNA, such as RiboGreen, and read the fluorescence on a plate reader
  3. Convert the fluorescence to a concentration using the calibration curve for each (RNA standards of known concentration)
  4. From the difference between the unencapsulated RNA on the outside and the total amount, derive the encapsulation efficiency, and from the total value the mRNA content

Decision points

Which surfactant do you use, and at what concentration?
If the LNP is not fully broken, the total RNA is undercounted and the encapsulation efficiency looks lower than it is. Triton X-100 has long been used, but its uses were restricted under REACH in the EU from 2021. In work on alternatives, a study has reported that Tween 20 extracted the RNA stably over the concentration range tested. How easily the particles break also changes with the formulation, so confirm the conditions for each formulation, and when you change them, handle it as a change of test method.
How do you build the calibration curve?
A procedure often seen is to build a calibration curve in the same liquid as the sample, for both the liquid without surfactant and the liquid with it. If the fluorescence of the solubilized side goes above the top of the calibration curve the total comes out low, so dilute to stay within the range. Choose RNA standards close to the mRNA being measured.
Besides the fluorescence method, what else is there?
New methods are appearing, such as instruments that measure total RNA by ultraviolet-visible absorbance without breaking the particles, and absorption spectroscopy that removes the effect of scattering. There is no guarantee that the values agree with RiboGreen, so measure the same lots side by side with your own formulation before deciding how to use each, and decide as well which value the specification is set on.
How do you set the specification for encapsulation efficiency?
Published information includes statements taking 85 to 90% or above as a guide, but it differs with the formulation and there is no universal threshold. Set it from the track record of lots whose efficacy has been confirmed and from the measurement scatter, and assess it together with content, particle size and PDI.

Common mistakes

  • Comparing an encapsulation efficiency figure whose conditions are not stated with other figures
  • Building the calibration curve in a different liquid from the sample, so the difference in liquid goes straight into a shift in the numbers
  • Confusing encapsulation efficiency (a proportion) with content (an amount) and using it as the basis of the dose

Check at your own site

  • Is your encapsulation efficiency written into the specification together with the type and concentration of surfactant and the liquid of the calibration curve?
  • Do your solubilization conditions come with a procedure for confirming them again when the formulation changes?

Step 18: Lipid quantitation, impurities

Lipid ratio and degradation products are directly tied to delivery efficiency and safety, so this step follows quantity and impurities by HPLC-CAD/LC-MS.Lipid ratio and degradation products are directly tied to delivery efficiency and safety, so this step follows quantity and impurities by HPLC-CAD/LC-MS.

Standard methodLipid quantitation and impurity profile by HPLC-charged aerosol detection (CAD)/LC-MSLipid quantitation and impurity profile by HPLC-charged aerosol detection (CAD)/LC-MS
Details and 3 alternatives

This step coversIdentification and quantitation of lipid components and lipid-related impurities (lipid quantitation, impurities)

A test that confirms the 4 lipids composing the particle are present in the designed ratio. It also checks whether degradation products have increased during storage.

The quality of the ionizable lipid in particular is directly tied to delivery efficiency and safety of the mRNA. If hydrolysis products and oxidized forms increase, efficacy drops and the risk of adverse reactions rises.

Each lipid is quantified by HPLC-CAD or LC-MS and an impurity profile is drawn. An analytical step that assures the contents of the formulation at the component level.

Confirm that the 4 lipids are present in the designed ratio and that degradation products (hydrolysis products and oxidized forms) have not increased during storage. The quality of the ionizable lipid is directly tied to the delivery efficiency and safety of the mRNA.

MakersMerck, Agilent, Waters, Bruker, Avanti

Why it is the standard. Chemical stability and content of lipids are normally evaluated by HPLC-CAD and LC-MS (nanosphere.ch; LNP characterization review). CAD can detect lipids with weak UV absorption universally and can also follow the degradation product profile.

Alternatives in practice

  • RP-HPLC-ELSD (evaporative light scattering detection) · Agilent 1260 Infinity II+ELSD/Sedere SEDEX — General-purpose detection when CAD is not available. Linearity and sensitivity are somewhat inferior to CAD.
  • Thin-layer chromatography (TLC) · Merck TLC Silica gel 60 plates + densitometer (CAMAG) — Simple, low-cost lipid screening. Limited quantitative accuracy.
  • Nuclear magnetic resonance (NMR) · Bruker Avance NEO 400/500 MHz — An orthogonal method for lipid structure identification and ratio confirmation. Not good at detecting trace impurities.

Step guide

What you actually do

  1. Break the LNP with an organic solvent, dissolving the lipids out into a test solution
  2. Separate the 4 lipids and their degradation products on a reverse-phase column with a gradient
  3. Detect each lipid with a charged aerosol detector (CAD) or by mass spectrometry, and quantify against a calibration curve from reference standards
  4. Compare the molar ratio of each lipid, and the amount of lipid-derived impurities, with the specification and with the previous lot

Equipment and materials

Decision points

CAD, ELSD, or mass spectrometry as the detector?
Many lipids absorb ultraviolet light weakly, so CAD and the evaporative light scattering detector (ELSD) are used. ELSD serves as general-purpose detection where there is no CAD, but it is said to be a little behind CAD in linearity and sensitivity. LC-MS is used to identify degradation products, and NMR is used as an orthogonal method for structure and ratio, but it is not suited to detecting trace impurities.
How do you obtain the reference standards?
Quantitation needs a reference standard for each lipid. Confirm standards of known purity and provenance in incoming testing, and if you make the analysis used for incoming testing of raw materials and the lipid testing of the product run on the same standards, figures become easier to compare between lots.
How far do you watch the degradation products?
Hydrolysis products and oxidation products of the ionizable lipid, and loss of the PEG lipid, are attributes that increase during storage. Confirm that the method is stability-indicating (one in which forced degradation separates the degradation products as peaks) before using it in stability studies.

Common mistakes

  • Measuring lipids with ultraviolet detection alone, and missing the lipids that absorb weakly
  • Quantifying without confirming the lot and the purity of the reference standard, so the basis of the figures cannot be explained
  • Using a method that cannot separate the degradation products that increase during storage in a stability study

Check at your own site

  • Does your lipid analysis look at all 4 lipids and the degradation products by the same method?
  • Do your incoming testing of lipids and the lipid testing of the product run on the same reference standards?

Step 19: Potency test

A potency test that measures whether cells can make the target protein, assuring the true efficacy of the product and lot-to-lot consistency.A potency test that measures whether cells can make the target protein, assuring the true efficacy of the product and lot-to-lot consistency.

Standard methodCell-based in vitro expression assayCell-based in vitro expression assay
Details and 3 alternatives

This step coversPotency/in vitro protein expression (potency test)

A potency test that puts the product into cells and confirms whether the target protein can be produced properly. It is the indicator reflecting the 'real efficacy' of the product.

Even when the full-length percentage and encapsulation efficiency are in order, in the end expression in cells is the proof of effect. Only here can whether the product functions be confirmed directly.

A cell-based in vitro expression assay measures expression to assure that potency is constant between lots. An analytical step that supports release decisions.

Ultimately, confirm 'whether, when put into cells, the target protein can be produced properly.' This is the indicator of the real efficacy of the product. Dosed to cells, expression is measured to assure that potency is constant between lots.

MakersBio-Rad, Promega, Molecular Devices, BD, Applied Biosystems

Why it is the standard. The potency of mRNA-LNP is generally evaluated by flow cytometry or qRT-PCR for transfection efficiency and target protein production (npj Vaccines 2022, s41541-022-00470-4). A sandwich ELISA for quantifying the target protein has also been developed and characterized (PMC10383996). HEK293 gives high detectability with 24-hour expression and is widely used.

Check also. Cell-based expression is an indicator of potency, but values move with cell type and measurement system. Potency as a specification must be designed in line with the mechanism of action of the product and aligned to an assay that can assure lot-to-lot consistency.

Alternatives in practice

  • Reporter (luciferase/eGFP) expression assay · Promega GloMax+ONE-Glo/Bright-Glo, or eGFP by flow cytometry (CytoFLEX) — Evaluates expression with high sensitivity and quantitatively. Requires reporter-construct mRNA.
  • Evaluation of cellular uptake/expression by qRT-PCR · QuantStudio 5+TaqMan — Measures uptake before translation/transcript amount. A different indicator from protein expression.
  • Confirmation of expressed protein by Western blot · Bio-Rad Trans-Blot Turbo+ChemiDoc MP — An orthogonal method to confirm the molecular weight and identity of the expressed protein. Semi-quantitative.

Step guide

What you actually do

  1. Seed the assay cells (HEK293, HepG2 and the like) onto plates at the set density
  2. Dilute the reference lot and the test lot in a dilution series, add them to the cells, and culture for the set time
  3. Measure the expressed protein by ELISA, flow cytometry, or the luminescence of a reporter such as luciferase
  4. Fit the dose-response curves and derive the relative potency against the reference lot

Equipment and materials

Materials and reagentsReporter gene assay (luciferase and the like)ELISA kit (target protein)Reference lot (reference standard for potency)

Decision points

What do you use as the read-out?
Design it in line with the mechanism of action of the product. ELISA and flow cytometry, which measure the target protein directly, are close to the mechanism; a reporter (luciferase or eGFP) is sensitive and quantitative but needs an mRNA carrying the reporter. RT-qPCR is a different measure, looking at uptake and the amount of transcript, which is not the same thing as protein expression.
How do you choose and control the cells?
The values move with the cell type and the measurement system, so choose cells in line with the mechanism of action. Settle the control of the cell bank, the upper limit on passage number and the checks on their condition, so that the system can be maintained for a long time.
How do you handle the reference lot?
Relative potency is derived as a ratio against the reference lot. Write how the reference lot is made, how it is stored, and the bridging study for when it is replaced, into the procedure in advance.
How do you set the specification?
Set the acceptable range for relative potency from the scatter of the assay and from the track record of the lots used in the clinic. Parallelism of the curves, precision and robustness are confirmed in validation.

Common mistakes

  • Not controlling the passage number or the change in condition of the cells, and reading variation in potency as a difference between lots
  • Not running a bridging study when the reference lot is renewed
  • Comparing results from a reporter mRNA with the potency of the product on the same scale

Check at your own site

  • Is your potency assay a read-out in line with the mechanism of action of the product?
  • How is your reference lot made, and how is its replacement controlled?

Step 20: General and safety tests

pH, osmolality, appearance, sterility and endotoxin are release specification items essential for administration of an injectable to patients.pH, osmolality, appearance, sterility and endotoxin are release specification items essential for administration of an injectable to patients.

Standard methodPharmacopoeia-compliant general tests (pH/osmolality/appearance/sterility test/endotoxin test (LAL))Pharmacopoeia-compliant general tests (pH/osmolality/appearance/sterility test/endotoxin test (LAL))
Details and 3 alternatives

This step coversGeneral properties and safety tests (pH, osmolality, appearance, sterility, endotoxin) (general and safety tests)

A general test that checks the basic safety items of an injectable together. Confirmation of sterility and endotoxin is added to pH, osmolality and appearance (foreign matter and aggregation).

That the product is sterile and that endotoxin, a pyrogen, is at or below the specification are conditions that absolutely cannot be omitted for administration to patients.

Tested by pharmacopoeia-compliant methods (sterility test, LAL test, etc.) and judged on whether the release specifications are met. An analytical step that ultimately protects patient safety.

Confirm the basic safety items for an injectable. In addition to pH, osmolality and appearance (foreign matter, aggregation), sterility and endotoxin (pyrogen) at or below the specification are mandatory release specifications for administration to patients.

MakersAdvanced Instruments, Getinge, Yokogawa, Lonza, bioMérieux

Why it is the standard. pH, osmolality, appearance, sterility (pharmacopoeial sterility test) and bacterial endotoxin (LAL; ICH/each pharmacopoeia) are mandatory as release specifications for injectable formulations. Recombinant Factor C (rFC) is increasingly listed in the pharmacopoeias as an alternative to LAL.

Alternatives in practice

  • Endotoxin quantitation by recombinant Factor C (rFC) method · Lonza PyroGene rFC Assay + fluorescence reader (BMG/SpectraMax)/bioMérieux ENDONEXT — A sustainable method that does not use horseshoe crab-derived LAL. Listing in the Ph.Eur. and others is progressing.
  • Rapid sterility test (nucleic acid amplification/respiration-metabolism method) · bioMérieux BACT/ALERT/Rapid Micro Biosystems Growth Direct — Results in a short turnaround. Presupposes validation against the culture method.
  • Subvisible particle test (light obscuration/flow imaging) · Beckman Coulter HIAC 9703+/Yokogawa FlowCam 8000 — Particle test such as USP<788>. Also applied to evaluating LNP aggregates.

Step guide

What you actually do

  1. Measure with a pH meter and an osmometer under the set conditions (for a lyophilized product, measure on the reconstituted liquid)
  2. Check for foreign matter, aggregation and colour by eye or with an inspection machine
  3. Run the sterility test in a sterile environment by membrane filtration or the direct method
  4. Measure endotoxin with LAL or recombinant factor C (rFC) reagent, and count subvisible particles by light obscuration or flow imaging

Equipment and materials

Materials and reagentsStandard solutions for pH and osmolalityEndotoxin standard and LAL reagent water

Decision points

LAL or rFC for the endotoxin test?
rFC is a reagent that does not use horseshoe-crab-derived LAL, and its inclusion in the pharmacopoeias is advancing. When you switch, measure the same lots by both methods to confirm how they correspond, and handle it as a change of test method. LAL also comes in gel-clot, turbidimetric and chromogenic methods; choose by the quantitation needed and the number of samples.
Do the components of the formulation interfere with the endotoxin test?
Components such as the buffer and the lipids can weaken or strengthen the reaction. Confirm with the formulation that the recovery of a positive control spiked with a known amount of endotoxin falls within the specified range (50 to 200%, for example). The phenomenon of low endotoxin recovery (LER), in which spiked endotoxin becomes undetectable over time, is also known, so study it including the hold time.
Culture method or rapid method for the sterility test?
Against the culture method, rapid methods using nucleic acid amplification or respiratory metabolism give results on a short turnaround. A rapid method presumes validation against the culture method, and which test is approved by which method follows the regulatory documents and the policy of the quality unit.
What do you count subvisible particles with?
Light obscuration is the basis, and where you also want to see the shape, flow imaging is used alongside. It is also applied to assessment of LNP aggregates, but confirm in method validation whether what is being counted is particles or bubbles.

Common mistakes

  • Adopting endotoxin figures without confirming interference from the formulation
  • Overlooking endotoxin contamination from equipment or from water
  • Changing the test method but treating it as mere reagent procurement and not going through the change procedure for a test method

Check at your own site

  • Does your endotoxin test confirm the recovery of the positive control in a liquid containing the buffer and lipids of the formulation?
  • Are the environmental monitoring and analyst records of your sterility testing linked to the test records of the lot?

Guides that go deeper

The reference (“gold standard”) method is the one broadly recognized in published practice and guidelines; your process, product and regulator may justify another. Makers and products are examples, not endorsements. This map is translated from the Japanese edition, where it is updated first (Japanese version).